Test-retest reliability of a portable monitor to assess energy expenditure

被引:0
|
作者
Brazeau, Anne-Sophie [1 ,2 ,3 ]
Karelis, Antony D. [2 ,4 ]
Mignault, Diane [2 ,5 ]
Lacroix, Marie-Josee [2 ]
Prud'homme, Denis [6 ]
Rabasa-Lhoret, Remi [1 ,2 ,3 ]
机构
[1] Montreal Clin Res Inst IRCM, Montreal, PQ, Canada
[2] Univ Montreal, Dept Nutr, Montreal, PQ H3T 1A8, Canada
[3] Univ Montreal Hosp Res Ctr CR CHUM, Montreal Diabet Res Ctr MDRC, Montreal, PQ H1W 4A4, Canada
[4] Univ Quebec, Dept Kinanthropol, Montreal, PQ H2X 3R9, Canada
[5] Montreal Clin Res Inst IRCM, Montreal, PQ H2W 1R7, Canada
[6] Univ Ottawa, Fac Hlth Sci, Sch Human Kinet, Ottawa, ON K1N 6N5, Canada
基金
加拿大健康研究院;
关键词
test-retest; activity monitoring; motion sensors; energy expenditure; accelerometer; physical activity; OBSTRUCTIVE PULMONARY-DISEASE; PHYSICAL-ACTIVITY; MULTISENSOR ARMBAND; WEIGHT-LOSS; EXERCISE; ADULTS; TECHNOLOGY; ACCURACY; SYSTEM; DEVICE;
D O I
10.1139/H11-016
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
The ability of the SenseWear Armband (SWA) to estimate energy expenditure (EE) in adults is established. However, except for resting metabolic rate, the test-retest reliability of the SWA for the estimation of EE in adults is unknown. To explore reliability, 34 healthy adults (50% women; age, 26.7 +/- 6.1 years; body mass index, 23.6 +/- 3.3 kg(.)m(-2)) completed a 2-day study protocol. During 13 h of direct supervision, subjects completed, each day, 60 min of lying awake rest, 30 min of structured sedentary activity, 45 min of walking circuit, and 45 min of ergocycling at 50% of their peak oxygen uptake. For the remaining 10 h, subjects remained seated. The SWA showed significant intraclass correlations between the 2 days of testing for the overall 13 h of direct supervision (r = 0.97; p < 0.001), the lying awake rest (r = 0.98; p < 0.001), the structured sedentary activity (r = 0.62; p < 0.05), the walking circuit (r = 0.95; p < 0.001), and the standardized physical activity (r = 0.85; p < 0.001). The results of this study indicate that the SWA is a reliable tool to estimate EE during multiple activities. These data provide additional evidence of the usefulness of the SWA in monitoring EE in healthy adults.
引用
收藏
页码:339 / 343
页数:5
相关论文
共 50 条
  • [21] TEST-RETEST RELIABILITY IN POPULATION RESEARCH
    BRACKBILL, Y
    STUDIES IN FAMILY PLANNING, 1974, 5 (08) : 261 - 266
  • [22] Cluster analysis for test-retest Reliability
    Roy, Debdulal Dutta
    INTERNATIONAL JOURNAL OF PSYCHOLOGICAL RESEARCH, 2010, 3 (01): : 131 - 139
  • [23] THE TEST-RETEST RELIABILITY OF QUALITATIVE DATA
    Guttman, Louis
    PSYCHOMETRIKA, 1946, 11 (02) : 81 - 95
  • [24] Test-retest reliability of functional MRI
    Tallent, KA
    Ramsey, NF
    Frank, JA
    Moonen, CT
    Weinberger, DR
    INTERNATIONAL JOURNAL OF PSYCHOLOGY, 1996, 31 (3-4) : 12451 - 12451
  • [25] An Investigation of the Test-Retest Reliability of the miniPXI
    Haider, Aqeel
    Wallner, Günter
    Gerling, Kathrin
    Vanden Abeele, Vero
    Proceedings of the ACM on Human-Computer Interaction, 2024, 8 (CHI PLAY)
  • [26] SKEW REDUCES TEST-RETEST RELIABILITY
    DUNLAP, WP
    CHEN, RS
    GREER, T
    JOURNAL OF APPLIED PSYCHOLOGY, 1994, 79 (02) : 310 - 313
  • [27] A BASIS FOR ANALYZING TEST-RETEST RELIABILITY
    Guttman, Louis
    PSYCHOMETRIKA, 1945, 10 (04) : 255 - 282
  • [28] Test-retest reliability in Electroencephalographic Recordings
    Ximena Suarez-Revelo, Jazmin
    Fredy Ochoa-Gomez, John
    Duque-Grajales, Jon
    Montoya-Betancur, Alejandro
    Sanchez-Lopez, Sebastian
    2015 20TH SYMPOSIUM ON SIGNAL PROCESSING, IMAGES AND COMPUTER VISION (STSIVA), 2015,
  • [29] Test-retest reliability of loudness scaling
    Robinson, K
    Gatehouse, S
    EAR AND HEARING, 1996, 17 (02): : 120 - 123
  • [30] TEST-RETEST RELIABILITY OF THE IDEA INVENTORY
    VESTRE, ND
    PSYCHOLOGICAL REPORTS, 1984, 54 (03) : 873 - 874