Log file based Monte Carlo calculations for proton pencil beam scanning therapy

被引:23
|
作者
Winterhalter, Carla [1 ,2 ]
Meier, Gabriel [1 ]
Oxley, David [1 ]
Weber, Damien C. [1 ,3 ,4 ]
Lomax, Antony J. [1 ,2 ]
Safai, Sairos [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
[2] Swiss Fed Inst Technol, Dept Phys, Zurich, Switzerland
[3] Univ Hosp Bern, Radiat Oncol Dept, Bern, Switzerland
[4] Univ Hosp Zurich, Radiat Oncol Dept, Zurich, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 03期
关键词
proton therapy; pencil beam scanning; Monte Carlo simulation; log file analysis; quality assurance; QUALITY-ASSURANCE; DOSE CALCULATIONS; SYSTEM; IMPACT;
D O I
10.1088/1361-6560/aaf82d
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Patient specific quality assurance is crucial to guarantee safety in proton pencil beam scanning. In current clinical practice, this requires extensive, time consuming measurements. Additionally, these measurements do not consider the influence of density heterogeneities in the patient and are insensitive to delivery errors. In this work, we investigate the use of log file based Monte Carlo calculations for dose reconstructions in the patient CT, which takes the combined influence of calculational and delivery errors into account. For one example field, 87%190% of the voxels agree within +/- 3% when taking either calculational or delivery uncertainties into account (analytical versus Monte Carlo calculation/Monte Carlo from planned versus Monte Carlo from log file). 78% agree when considering both uncertainties simultaneously (nominal field versus Monte Carlo from log files). We then show the application of the log file based Monte Carlo calculations as a patient specific quality assurance tool for a set of five patients (16 fields) treated for different indications. For all fields, absolute dose scaling factors based on the log file Monte Carlo agree within +/- 3% to the measurement based absolute dose scaling. Relative comparison shows that more than 90% of the voxels agree within +/- 5% between the analytical calculated plan and the Monte Carlo based on log files. The log file based Monte Carlo approach is an end-to-end test incorporating all requirements of patient specific quality assurance. It has the potential to reduce the workload and therefore to increase the patient throughput, while simultaneously enabling more accurate dose verification directly in the patient geometry.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Log-file based Monte Carlo simulations for proton pencil beam scanning (PBS)
    Winterhalter, C.
    Meier, G.
    Safai, S.
    Oxley, D.
    Bolsi, A.
    Fredh, A.
    Weber, D. C.
    Lomax, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1018 - S1019
  • [2] Pitfalls in the beam modelling process of Monte Carlo calculations for proton pencil beam scanning
    Winterhalter, Carla
    Aitkenhead, Adam
    Oxley, David
    Richardson, Jenny
    Weber, Damien C.
    MacKay, Ranald, I
    Lomax, Antony J.
    Safai, Sairos
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1107):
  • [3] Monte Carlo Dose Verification of Pencil Beam Scanning Proton Therapy
    Giantsoudi, D.
    Clasie, B.
    Grassberger, C.
    Dowdell, S.
    Depauw, N.
    Madden, T.
    Kooy, H.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [4] A Monte Carlo study on the collimation of pencil beam scanning proton therapy beams
    Charlwood, Frances C.
    Aitkenhead, Adam H.
    Mackay, Ranald I.
    [J]. MEDICAL PHYSICS, 2016, 43 (03) : 1462 - 1472
  • [5] Monte Carlo model of a gantry for proton pencil beam scanning
    Winterhalter, Carla
    Safai, Sairos
    Grassberger, Clemens
    Oxley, David
    Weber, Damien C.
    Lomax, Tony
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 837 - 837
  • [6] The Application of Cloud Computing in Pencil Beam Scanning Proton Therapy Monte Carlo Simulation
    Wang, Z.
    Gao, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 296 - 296
  • [7] Clinical validation of Monte Carlo dose calculation for pencil beam scanning proton therapy.
    Widesott, L.
    Lorentini, S.
    Fracchiolla, F.
    Farace, P.
    Schwarz, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S45 - S46
  • [8] Validation and clinical implementation of an accurate Monte Carlo code for pencil beam scanning proton therapy
    Huang, Sheng
    Kang, Minglei
    Souris, Kevin
    Ainsley, Christopher
    Solberg, Timothy D.
    McDonough, James E.
    Simone, Charles B.
    Lin, Liyong
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 558 - 572
  • [9] Machine Learning Based Log File Data Analysis for Quality Assurance of Proton Pencil Beam Scanning
    Dou, T.
    Clasie, B.
    Depauw, N.
    Shen, T.
    Flanz, J.
    Lu, H.
    Jee, K.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E603 - E603
  • [10] Machine Log File Analysis for Quality Assurance of Proton Pencil Beam Scanning System
    Zhang, R.
    Clasie, B.
    Depauw, N.
    Lin, Y.
    Shen, T.
    Flanz, J.
    Lu, H.
    Jee, K.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E264 - E264