Computer navigation for total knee arthroplasty achieves better postoperative alignment compared to conventional and patient-specific instrumentation in a low-volume setting

被引:19
|
作者
Suero, Eduardo M. [1 ]
Lueke, Ulrich [1 ]
Stuebig, Timo [1 ]
Hawi, Nael [1 ]
Krettek, Christian [1 ]
Liodakis, Emmanouil [1 ]
机构
[1] Hannover Med Sch, Trauma Dept, Carl Neuberg Str 1, D-30625 Hannover, Germany
关键词
Knee; Total knee Arthroplasty; Patient-specific TKA; Computer navigation; Low-volume; Alignment; SURGEON PROCEDURE VOLUME; SHORT-TERM OUTCOMES; IMPROVED ACCURACY; CUTTING GUIDES; TKA; REPLACEMENT; ASSOCIATION; DIFFERENCE; BENEFIT; BLOCKS;
D O I
10.1016/j.otsr.2018.04.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Procedure volume is an important determinant of total knee arthroplasty (TKA) outcomes. We aimed to determine whether computer navigation or patient-specific instrumentation (PSI) would improve postoperative alignment in a low-volume setting. Hypothesis: PSI for TKA achieves better limb and implant alignment compared to conventional TKA and to computer navigated TKA. Materials and methods: This is a retrospective cohort study of 385 primary TI(As (Women = 59%. Mean age = 67 years. Mean BMI =30.1 kg/m(2)), which were performed using conventional instrumentation (n= 117; 30%), computer navigation (n= 209; 54%), or patient-specific instrumentation (n= 59; 15%) in a low-volume center (< 50 TKAs/year). The risk of postoperative leg and implant mechanical alignment outliers in the coronal plane (>3 degrees from neutral), average alignment and operation time were assessed. Results: The risk of postoperative mechanical alignment outliers (>3 degrees) was reduced by 89% in the navigated group (4% outliers) compared to the conventional group (35%) (RR = 0.11; p < 0.0001). No significant improvement was observed in the PSI group (27%) (RR = 0.91; p = 0.772). The risk of postoperative femoral component coronal alignment outliers was reduced by 63% in the navigated group (11%) compared to the conventional group (31%) (RR= 0.37; p = 0.018). No significant reduction in outliers was observed in the PSI group (32%) (RR= 1.08; p = 0.816). There was a reduction in the risk of tibial component coronal malalignment of 66% in the navigated group (5%) compared to the conventional group (13%) (RR= 0.33; p = 0.070). There was a two-fold increase in the risk of tibial component alignment outliers in the PSI group (29%) (RR= 1.94; p = 0.110). Discussion: Computer navigation improved postoperative alignment in TKA. No evidence of improved alignment was seen with patient-specific instrumentation. The routine use of patient-specific instrumentation in low-volume centers is not supported by the currently available data. Type of study: Retrospective cohort study. (C) 2018 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:971 / 975
页数:5
相关论文
共 50 条
  • [21] Patient-Specific Instrumentation in Total Knee Arthroplasty: A Review
    Nam, Denis
    McArthur, Benjamin A.
    Cross, Michael B.
    Pearle, Andrew D.
    Mayman, David J.
    Haas, Steven B.
    JOURNAL OF KNEE SURGERY, 2012, 25 (03) : 213 - 219
  • [22] Patient-specific instrumentation and custom total knee implants in total knee arthroplasty
    Isaacson, Mark
    Bunn, Kevin
    Incavo, Stephen
    CURRENT ORTHOPAEDIC PRACTICE, 2015, 26 (03): : 224 - 227
  • [23] Patient-specific instrumentation for total knee replacement verified by computer navigation: a case report
    Chan, Kok-Yu
    Teo, Yee-Hong
    JOURNAL OF ORTHOPAEDIC SURGERY, 2012, 20 (01) : 111 - 114
  • [24] Patient-specific instrumentation for total knee arthroplasty: Evaluation with computer-tomography tools
    Nabavi, Arash
    Olwill, Caroline M.
    Do, Mike
    Wanasawage, Tanya
    Harris, Ian A.
    JOURNAL OF ORTHOPAEDIC SURGERY, 2017, 25 (01)
  • [25] Patient-specific instrumentation for total knee arthroplasty: a literature review
    Camarda L.
    D’Arienzo A.
    Morello S.
    Peri G.
    Valentino B.
    D’Arienzo M.
    MUSCULOSKELETAL SURGERY, 2015, 99 (1) : 11 - 18
  • [26] Patient-Specific Instrumentation in Total Knee Arthroplasty: What Is the Evidence?
    Szczech, Bartlomiej
    McDermott, James D.
    Issa, Kimona
    Rifai, Aiman
    Festa, Anthony
    Matarese, William A.
    McInerney, Vincent K.
    JOURNAL OF KNEE SURGERY, 2016, 29 (04) : 341 - 345
  • [27] Reliability of Templating with Patient-Specific Instrumentation in Total Knee Arthroplasty
    Issa, Kimona
    Rifai, Aiman
    McGrath, Mike S.
    Callaghan, John J.
    Wright, Craig
    Malkani, Arthur L.
    Mont, Michael A.
    McInerney, Vincent K.
    JOURNAL OF KNEE SURGERY, 2013, 26 (06) : 429 - 433
  • [28] Patient-specific instrumentation makes sense in total knee arthroplasty
    Leon-Munoz, Vicente J.
    Lopez-Lopez, Mirian
    Santonja-Medina, Fernando
    EXPERT REVIEW OF MEDICAL DEVICES, 2022, 19 (06) : 489 - 497
  • [29] Total Knee Arthroplasty: Validating The Benefits Of Patient-Specific Instrumentation
    Moore, S.
    Linton, K. N.
    Martin, P.
    Headon, R. J.
    Bennett, D.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2018, 187 : S74 - S75
  • [30] In-Vivo Alignment Comparing Patient Specific Instrumentation with both Conventional and Computer Assisted Surgery (CAS) Instrumentation in Total Knee Arthroplasty
    Barrett, William
    Hoeffel, Daniel
    Dalury, David
    Mason, J. Bohannon
    Murphy, Jeff
    Himden, Sam
    JOURNAL OF ARTHROPLASTY, 2014, 29 (02): : 343 - 347