Note: Calibration of atomic force microscope cantilevers using only their resonant frequency and quality factor

被引:18
|
作者
Sader, John E. [1 ]
Friend, James R. [2 ,3 ]
机构
[1] Univ Melbourne, Dept Math & Stat, Melbourne, Vic 3010, Australia
[2] RMIT Univ, Sch Elect & Comp Engn, Melbourne, Vic 3001, Australia
[3] Univ Calif San Diego, Dept Mech & Aerosp Engn, La Jolla, CA 92122 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2014年 / 85卷 / 11期
基金
澳大利亚研究理事会;
关键词
Atomic force microscope cantilevers - Highly accurate - Hydrodynamic functions - Laser Doppler Vibrometry - Non-ideal conditions - Quality factors - Simplified method - Spring constants;
D O I
10.1063/1.4901227
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A simplified method for calibrating atomic force microscope cantilevers was recently proposed by Sader et al. [Rev. Sci. Instrum. 83, 103705 (2012); Sec. III D] that relies solely on the resonant frequency and quality factor of the cantilever in fluid (typically air). This method eliminates the need to measure the hydrodynamic function of the cantilever, which can be time consuming given the wide range of cantilevers now available. Using laser Doppler vibrometry, we rigorously assess the accuracy of this method for a series of commercially available cantilevers and explore its performance under non-ideal conditions. This shows that the simplified method is highly accurate and can be easily implemented to perform fast, robust, and non-invasive spring constant calibration. (c) 2014 AIP Publishing LLC.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Note: Improved calibration of atomic force microscope cantilevers using multiple reference cantilevers
    Sader, John E.
    Friend, James R.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (05):
  • [2] Calibration of atomic force microscope cantilevers using piezolevers
    Aksu, Saltuk B.
    Turner, Joseph A.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (04):
  • [3] Calibration of rectangular atomic force microscope cantilevers
    Sader, JE
    Chon, JWM
    Mulvaney, P
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (10): : 3967 - 3969
  • [4] Calibration of silicon atomic force microscope cantilevers
    Gibson, CT
    Smith, DA
    Roberts, CJ
    NANOTECHNOLOGY, 2005, 16 (02) : 234 - 238
  • [5] Lateral force calibration for atomic force microscope cantilevers using a suspended nanowire
    Zhang, Guangjie
    Li, Peng
    Wei, Dawei
    Hu, Kui
    Qiu, Xiaohui
    NANOTECHNOLOGY, 2020, 31 (47)
  • [6] METHOD FOR THE CALIBRATION OF ATOMIC-FORCE MICROSCOPE CANTILEVERS
    SADER, JE
    LARSON, I
    MULVANEY, P
    WHITE, LR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (07): : 3789 - 3798
  • [7] Temperature calibration of heated silicon atomic force microscope cantilevers
    Nelson, B. A.
    King, W. P.
    SENSORS AND ACTUATORS A-PHYSICAL, 2007, 140 (01) : 51 - 59
  • [8] Calibration of the lateral spring constant of atomic force microscope cantilevers
    Song, Yunpeng
    Wu, Sen
    Xu, Linyan
    Fu, Xing
    AOPC 2015: MICRO/NANO OPTICAL MANUFACTURING TECHNOLOGIES; AND LASER PROCESSING AND RAPID PROTOTYPING TECHNIQUES, 2015, 9673
  • [9] A virtual instrument to standardise the calibration of atomic force microscope cantilevers
    Sader, John E.
    Borgani, Riccardo
    Gibson, Christopher T.
    Haviland, David B.
    Higgins, Michael J.
    Kilpatrick, Jason I.
    Lu, Jianing
    Mulvaney, Paul
    Shearer, Cameron J.
    Slattery, Ashley D.
    Thoren, Per-Anders
    Tran, Jim
    Zhang, Heyou
    Zhang, Hongrui
    Zheng, Tian
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (09):
  • [10] Thermal calibration of heated silicon atomic force microscope cantilevers
    Nelson, Brent A.
    King, William P.
    TRANSDUCERS '07 & EUROSENSORS XXI, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2007,