Development of a dedicated phantom for multi-target single-isocentre stereotactic radiosurgery end to end testing

被引:22
|
作者
Poder, Joel [1 ]
Brown, Ryan [1 ]
Porter, Harry [1 ]
Gupta, Rashmi [1 ]
Ralston, Anna [1 ]
机构
[1] St George Hosp, Canc Care Ctr, Kogarah, NSW, Australia
来源
关键词
dosimetric validation; end-to-end testing; MRI deformation; multi-target; single isocentre; stereotactic radiosurgery; RADIOCHROMIC FILM DOSIMETRY; BRAIN RADIATION-THERAPY; MODULATED ARC THERAPY; IMAGE REGISTRATION; LINEAR-ACCELERATOR; TARGET DELINEATION; CT SCANS; SRS QA; RADIOTHERAPY; METASTASES;
D O I
10.1002/acm2.12452
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this project was to design and manufacture a cost-effective end-to-end (E2E) phantom for quantifying the geometric and dosimetric accuracy of a linear accelerator based, multi-target single-isocenter (MTSI) frameless stereotactic radiosurgery (SRS) technique. Method A perspex Multi-Plug device from a Sun Nuclear ArcCheck phantom (Sun Nuclear, Melbourne, FL) was enhanced to make it more applicable for MTSI SRS E2E testing. The following steps in the SRS chain were then analysed using the phantom: magnetic resonance imaging (MRI) distortion, planning computed tomography (CT) scan and MRI image registration accuracy, phantom setup accuracy using CBCT, dosimetric accuracy using ion chamber, planar film dose measurements and coincidence of linear accelerator mega-voltage (MV), and kilo-voltage (kV) isocenters using Winston-Lutz testing (WLT). Results The dedicated E2E phantom was able to successfully quantify the geometric and dosimetric accuracy of the MTSI SRS technique. MRI distortions were less than 0.5 mm, or half a voxel size. The average MRI-CT registration accuracy was 0.15 mm (+/- 0.31 mm), 0.20 mm (+/- 0.16 mm), and 0.39 mm (+/- 0.11 mm) in the superior/inferior, left/right and, anterior/posterior directions, respectively. The phantom setup accuracy using CBCT was better than 0.2 mm and 0.1 degrees. Point dose measurements were within 5% of the treatment planning system predicted dose. The comparison of planar film doses to the planning system dose distributions, performed using gamma analysis, resulted in pass rates greater than 97% for 3%/1 mm gamma criteria. Finally, off-axis WLT showed MV/kV coincidence to be within 1 mm for off-axis distances up to 60 mm. Conclusion A novel, versatile and cost-effective phantom for comprehensive E2E testing of MTSI SRS treatments was developed, incorporating multiple detector types and fiducial markers. The phantom is capable of quantifying the accuracy of each step in the MTSI SRS planning and treatment process.
引用
收藏
页码:99 / 108
页数:10
相关论文
共 50 条
  • [31] A New Approach to the Characterization of a Radiosurgery System: End-To-End Testing Using Phantom-Based 3D Dosimetry System
    Wang, Y.
    Fan, P.
    Li, F.
    Xu, A.
    Wang, T.
    Choi, J.
    Adamovics, J.
    Wuu, C.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E384 - E384
  • [32] Single Fraction Spine SBRT End to End Testing On TomoTherapy and Vero Treatment Platforms Using a Novel Anthropomorphic Phantom
    Gallo, J.
    Kaufman, I.
    Bossenberger, T.
    Powell, R.
    Ramirez, E.
    Reynolds, R.
    Solberg, T.
    Burmeister, J.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [33] Experimental determination of set-up displacements in anthropomorphic phantom in single-isocentre radiosurgery for multiple brain metastases by off-axis Winston-Lutz test: ExacTrac™ v.6 versus Dynamic™
    Rojas-Lopez, J. A.
    Chesta, M. A.
    Venencia, C. D.
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2023, 22
  • [34] Search-Based Testing of Procedural Programs: Iterative Single-Target or Multi-target Approach?
    Scalabrino, Simone
    Grano, Giovanni
    Di Nucci, Dario
    Oliveto, Rocco
    De Lucia, Andrea
    [J]. SEARCH BASED SOFTWARE ENGINEERING, SSBSE 2016, 2016, 9962 : 64 - 79
  • [35] Evaluation of multi-institutional end-to-end testing for post-operative spine stereotactic body radiation therapy
    Furuya, Tomohisa
    Lee, Young K.
    Archibald-Heeren, Ben R.
    Byrne, Mikel
    Bosco, Bruno
    Phua, Jun H.
    Shimizu, Hidetoshi
    Hashimoto, Shimpei
    Tanaka, Hiroshi
    Sahgal, Arjun
    Karasawa, Katsuyuki
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2020, 16 : 61 - 68
  • [36] Can Optical Surface Imaging Replace Non-coplanar Cone-beam Computed Tomography for Non-coplanar Set-up Verification in Single-isocentre Non-coplanar Stereotactic Radiosurgery and Hypofractionated Stereotactic Radiotherapy for Single and Multiple Brain Metastases?
    Lai, J. L.
    Liu, S. P.
    Jiang, X. X.
    Liu, J.
    Li, A.
    Li, B.
    Li, X. K.
    Ye, X. J.
    Lei, K. J.
    Zhou, L.
    [J]. CLINICAL ONCOLOGY, 2023, 35 (12) : E657 - E665
  • [37] Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc
    Hyunuk Jung
    Jihyung Yoon
    Olga Dona Lemus
    Sean Tanny
    Yuwei Zhou
    Michael Milano
    Kenneth Usuki
    Sara Hardy
    Dandan Zheng
    [J]. Radiation Oncology, 19
  • [38] An Integrated Quality Assurance Phantom for Frameless Single-Isocenter Multitarget Stereotactic Radiosurgery: A Multi-Institutional Trial
    Capaldi, D.
    Nano, T.
    Skinner, L.
    Chuang, C.
    Dubrowski, P.
    Xing, L.
    Yu, A.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E386 - E387
  • [39] Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc
    Jung, Hyunuk
    Yoon, Jihyung
    Lemus, Olga Dona
    Tanny, Sean
    Zhou, Yuwei
    Milano, Michael
    Usuki, Kenneth
    Hardy, Sara
    Zheng, Dandan
    [J]. RADIATION ONCOLOGY, 2024, 19 (01)
  • [40] The effect of MLC leaf width in single-isocenter multi-target radiosurgery with volumetric modulated arc therapy
    Abisheva, Zhanerke
    Floyd, Scott R.
    Salama, Joseph K.
    Kirkpatrick, John
    Yin, Fang-Fang
    Moravan, Michael J.
    Giles, William
    Adamson, Justus
    [J]. JOURNAL OF RADIOSURGERY AND SBRT, 2019, 6 (02): : 131 - 138