Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc

被引:1
|
作者
Jung, Hyunuk [1 ]
Yoon, Jihyung [1 ]
Lemus, Olga Dona [1 ]
Tanny, Sean [1 ]
Zhou, Yuwei [1 ]
Milano, Michael [1 ]
Usuki, Kenneth [1 ]
Hardy, Sara [1 ]
Zheng, Dandan [1 ]
机构
[1] Univ Rochester, Dept Radiat Oncol, Rochester, NY 14627 USA
关键词
Brain metastases; Stereotactic radiosurgery; LINAC-based SRS; HyperArc; RapidArc; MME; BRAIN METASTASES; RADIATION-THERAPY; CONFORMITY;
D O I
10.1186/s13014-024-02416-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
BackgroundTo compare the dosimetric quality of three widely used techniques for LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery (fSRS) with more than 20 targets: dynamic conformal arc (DCA) in BrainLAB Multiple Metastases Elements (MME) module and volumetric modulated arc therapy (VMAT) using RapidArc (RA) and HyperArc (HA) in Varian Eclipse.MethodsTen patients who received single-isocenter fSRS with 20-37 targets were retrospectively replanned using MME, RA, and HA. Various dosimetric parameters, such as conformity index (CI), Paddick CI, gradient index (GI), normal brain dose exposures, maximum organ-at-risk (OAR) doses, and beam-on times were extracted and compared among the three techniques. Wilcoxon signed-rank test was used for statistical analysis.ResultsAll plans achieved the prescribed dose coverage goal of at least 95% of the planning target volume (PTV). HA plans showed superior conformity compared to RA and MME plans. MME plans showed superior GI compared to RA and HA plans. RA plans resulted in significantly higher low and intermediate dose exposure to normal brain compared to HA and MME plans, especially for lower doses of >= 8Gy and >= 5Gy. No significant differences were observed in the maximum dose to OARs among the three techniques. The beam-on time of MME plans was about two times longer than RA and HA plans.ConclusionsHA plans achieved the best conformity, while MME plans achieved the best dose fall-off for LINAC-based single-isocenter multi-target multi-fraction SRS with more than 20 targets. The choice of the optimal technique should consider the trade-offs between dosimetric quality, beam-on time, and planning effort.
引用
收藏
页数:9
相关论文
共 10 条
  • [1] Dosimetric evaluation of LINAC-based single-isocenter multi-target multi-fraction stereotactic radiosurgery with more than 20 targets: comparing MME, HyperArc, and RapidArc
    Hyunuk Jung
    Jihyung Yoon
    Olga Dona Lemus
    Sean Tanny
    Yuwei Zhou
    Michael Milano
    Kenneth Usuki
    Sara Hardy
    Dandan Zheng
    Radiation Oncology, 19
  • [2] Modeling Dosimetric Distributions for Brain Stereotactic Radiosurgery (SRS) for Single-Isocenter Multi-Target (SIMT) Treatments
    Zoller, W.
    Addington, M.
    Taylor, W.
    Sebastian, N.
    Palmer, J. D.
    Raval, R.
    DiCostanzo, D. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E91 - E91
  • [3] Biological Plan Evaluation for Linac-Based Stereotactic Radiosurgery: Comparison Between Single-Isocenter and Multiple-Isocenter Techniques
    Lee, M.
    Wang, H.
    Tsai, H.
    MEDICAL PHYSICS, 2020, 47 (06) : E711 - E711
  • [4] Innovative margin design and optimized isocenter to minimize the normal tissue in target volumes for single-isocenter multi-target stereotactic radiosurgery
    Yock, Adam D.
    Grees, Beshoi
    Luo, Guozhen
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (19):
  • [5] Predicting Brain V12Gy for Single-Isocenter Multi-Target Stereotactic Radiosurgery (SRS)
    Liu, H.
    Li, T.
    Zhang, H.
    Shi, W.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S692 - S692
  • [6] Conformality Indices and Brain Dose Objective for Multi Lesion Brain Single Isocenter Linac-Based Stereotactic Radiosurgery
    Wadi-Ramahi, S.
    Tavakoli, M.
    Elgohari, B.
    Ashmeg, S.
    Lalonde, R. J.
    Siddiqui, Z. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E154 - E154
  • [7] Determining Institutional Plan Quality Metrics for Intracranial Stereotactic Radiosurgery (SRS) for Single-Isocenter Multi-Target (SIMT) Treatments
    Taylor, W.
    Zoller, W.
    Addington, M.
    Palmer, J.
    Raval, R.
    Sebastian, N.
    DiCostanzo, D.
    MEDICAL PHYSICS, 2019, 46 (06) : E266 - E266
  • [8] Dosimetric Impact of Patient Rotational Setup Errors with Frameless Single-Isocenter, Multi-Target Volumetric Modulated Arc Radiosurgery for Multiple Brain Metastases
    Clark, G. M.
    Fiveash, J. B.
    Prendergast, B. M.
    Willey, C. W.
    Spencer, S. A.
    Thomas, E. M.
    Guthrie, B. L.
    Popple, R. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S888 - S888
  • [9] Dosimetric Performance and Planning/Delivery Efficiency of a Dual-Layer Stacked and Staggered MLC on Treating Multiple Small Targets: A Planning Study Based on Single-Isocenter Multi-Target Stereotactic Radiosurgery (SRS) to Brain Metastases
    Li, Taoran
    Irmen, Peyton
    Liu, Haisong
    Shi, Wenyin
    Alonso-Basanta, Michelle
    Zou, Wei
    Teo, Boon-Keng Kevin
    Metz, James M.
    Dong, Lei
    FRONTIERS IN ONCOLOGY, 2019, 9
  • [10] Evaluation of Margin for Intra-Fractional Patient Motion during Single-Isocenter Multi Targets Volumetric Modulated Arc Therapy Stereotactic Radiation Therapy for Brain Metastases Using Actual Target Coordinates
    Yamada, R.
    Yoshimura, T.
    Mori, T.
    Nishioka, K.
    Koizumi, F.
    Nishikawa, N.
    Fujita, Y.
    Takahashi, S.
    Kanehira, T.
    Yokokawa, K.
    Yamazaki, R.
    Horita, K.
    Tamura, H.
    Wakabayashi, Y.
    Ichiu, Y.
    Aoyama, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : E64 - E64