Dosimetric Performance and Planning/Delivery Efficiency of a Dual-Layer Stacked and Staggered MLC on Treating Multiple Small Targets: A Planning Study Based on Single-Isocenter Multi-Target Stereotactic Radiosurgery (SRS) to Brain Metastases

被引:31
|
作者
Li, Taoran [1 ]
Irmen, Peyton [1 ]
Liu, Haisong [2 ]
Shi, Wenyin [2 ]
Alonso-Basanta, Michelle [1 ]
Zou, Wei [1 ]
Teo, Boon-Keng Kevin [1 ]
Metz, James M. [1 ]
Dong, Lei [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[2] Thomas Jefferson Univ, Dept Radiat Oncol, Philadelphia, PA 19107 USA
来源
FRONTIERS IN ONCOLOGY | 2019年 / 9卷
关键词
halcyon; stereotactic radiosurgery (SRS); multi-leaf collimator (MLC); single-isocenter multi-target; truebeam; brain metastase; RADIATION-THERAPY; QUALITY;
D O I
10.3389/fonc.2019.00007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the dosimetric performance and planning/delivery efficiency of a dual-layer MLC system for treating multiple brain metastases with a single isocenter. Materials and Methods: 10 patients each with 6-10 targets with volumes from 0.11 to 8.57 cc, and prescription doses from 15 to 24Gy, were retrospectively studied. Halcyon has only coplanar delivery mode. Halcyon V1 MLC modulates only with the lower layer at 1 cm resolution, whereas V2 MLC modulates with both layers at an effective resolution of 0.5 cm. For each patient five plans were compared varying MLC and beam arrangements: the clinical plan using multi-aperture dynamic conformal arc (DCA) and non-coplanar arcs, Halcyon-V1 using coplanar-VMAT, Halcyon-V2 using coplanar-VMAT, HDMLC-0.25 cm using coplanar-VMAT, and HDMLC-0.25 cm using non-coplanar-VMAT. All same-case plans were generated following the same planning protocol and normalization. Conformity index (CI), gradient index (GI), V12Gy, V6Gy, V3Gy, and brain mean dose were compared. Results: All VMAT plans met clinical constraints for critical structures. For targets with diameter <1 cm, Halcyon plans showed inferior CI among all techniques. For targets with diameter >1 cm, Halcyon VMAT plans had CI similar to non-coplanar VMAT plans, and better than non-coplanar clinical DCA plans. For GI, Halcyon MLC plans performed similarly to coplanar HDMLC plans and inferiorly compared to non-coplanar HDMLC plans. All coplanar VMAT plans (Halcyon MLC and HDMLC) and clinical DCA plans had similar V12Gy, but were inferior compared to non-coplanar VMAT plans. Halcyon plans had slightly reduced V3Gy and mean brain dose compared to HDMLC plans. The difference between Halcyon V1 and V2 is only significant in CI of tumors less than 1cm in diameter. Halcyon plans required longer optimization than Truebeam VMAT plans, but had similar delivery efficiency. Conclusion: For targets with diameter >1 cm, Halcyon's dual-layer stacked and staggered MLC is capable of producing similar dose conformity compared to HDMLC while reducing low dose spill to normal brain tissue. GI and V12Gy of Halcyon MLC plans were, in general, inferior to non-coplanar DCA or VMAT plans using HDMLC, likely due to coplanar geometry and wider MLC leaves. HDMLC maintained its advantage in CI for smaller targets with diameter <1 cm.
引用
收藏
页数:11
相关论文
共 16 条
  • [1] Modeling Dosimetric Distributions for Brain Stereotactic Radiosurgery (SRS) for Single-Isocenter Multi-Target (SIMT) Treatments
    Zoller, W.
    Addington, M.
    Taylor, W.
    Sebastian, N.
    Palmer, J. D.
    Raval, R.
    DiCostanzo, D. J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E91 - E91
  • [2] Performance of a Dual-Layer Staggered MLC System in Stereotactic Treatment for Multiple Brain Metastases with a Single Isocenter: A Dosimetric Comparison Across Three Delivery Techniques
    Li, T.
    Liu, H.
    Shi, W.
    Zou, J.
    Teo, B.
    Dong, L.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E508 - E508
  • [3] Knowledge-Based Planning for Single-Isocenter Stereotactic Radiosurgery to Multiple Brain Metastases
    Ziemer, B.
    Shiraishi, S.
    Hattangadi-Gluth, J.
    Sanghvi, P.
    Moore, K.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3724 - 3725
  • [4] Single-isocenter multiple-target stereotactic radiosurgery for multiple brain metastases: dosimetric evaluation of two automated treatment planning systems
    Giorgio Hamid Raza
    Luca Capone
    Paolo Tini
    Martina Giraffa
    Piercarlo Gentile
    Giuseppe Minniti
    [J]. Radiation Oncology, 17
  • [5] Single-isocenter multiple-target stereotactic radiosurgery for multiple brain metastases: dosimetric evaluation of two automated treatment planning systems
    Raza, Giorgio Hamid
    Capone, Luca
    Tini, Paolo
    Giraffa, Martina
    Gentile, Piercarlo
    Minniti, Giuseppe
    [J]. RADIATION ONCOLOGY, 2022, 17 (01)
  • [6] Heuristic knowledge-based planning for single-isocenter stereotactic radiosurgery to multiple brain metastases
    Ziemer, Benjamin P.
    Sanghvi, Parag
    Hattangadi-Gluth, Jona
    Moore, Kevin L.
    [J]. MEDICAL PHYSICS, 2017, 44 (10) : 5001 - 5009
  • [7] Single-Isocenter Multi-Target SRS VMAT Treatment Planning Strategy: When Targets Compete for Limited MLC Beam Apertures
    Duan, J.
    Cui, G.
    Walls, N.
    Wang, I.
    Yin, F.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E964 - E965
  • [8] Dosimetric Evaluation of a Newly Available Planning Treatment Module That Is Capable of Planning and Treating Multiple Targets With a Single-Isocenter Setup for Brain Radiosurgery
    Luo, G.
    Morales-Paliza, M.
    Ding, G. X.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E663 - E664
  • [9] Single-Isocenter VMAT-Based Stereotactic Radiosurgery (SRS) for Multiple Brain Metastases: Potential Loss of Target Coverage Due to Isocenter Misalignment
    Palmiero, A.
    Molloy, J.
    Sanders, M.
    Randall, M.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E119 - E119
  • [10] Dosimetric Evaluation of a Cone-Based and a MLC-Based Treatment Planning Technique for Multiple Brain Metastases Stereotactic Radiosurgery (SRS)
    Vallone, N.
    Brindle, J.
    Saleh, Z.
    Schwer, M.
    Dipetrillo, T.
    Hepel, J.
    Kinsella, T.
    Leonard, K.
    Koffer, P.
    Yu, E.
    Yang, Y.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E755 - E755