Technical Note: Ion chamber angular dependence in a magnetic field

被引:14
|
作者
Reynolds, Michael [1 ]
Rathee, Satyapal [1 ,2 ]
Fallone, B. Gino [2 ,3 ,4 ]
机构
[1] Univ Alberta, Med Phys Div, Dept Oncol, 11560 Univ Ave, Edmonton, AB T6G 1Z2, Canada
[2] Cross Canc Inst, Dept Med Phys, 11560 Univ Ave, Edmonton, AB T6G 1Z2, Canada
[3] Univ Alberta, Dept Oncol, 11560 Univ Ave, Edmonton, AB T6G 1Z2, Canada
[4] Univ Alberta, Dept Phys, 11560 Univ Ave, Edmonton, AB T6G 1Z2, Canada
关键词
angular dependence; ion chamber; Linac-MR; magnetic field; Monte Carlo; setup inaccuracies; MV RADIOTHERAPY ACCELERATOR; MONTE-CARLO SIMULATIONS; REFERENCE DOSIMETRY; IONIZATION-CHAMBER; MRI SCANNER; DOSE DEPOSITION; TRANSPORT; PENELOPE; ALGORITHM; ELECTRONS;
D O I
10.1002/mp.12405
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: There have been several studies investigating dose deposition effects within radiation detectors in the presence of a magnetic field. However, to date there has only been a passing investigation which explicitly investigates detector dose-response as a function of detector orientation. Herein we will investigate the dose-response as a function angular orientation of a PR06C ionization chamber. We will also benchmark the Monte Carlo code PENELOPE with the newly developed magnetic field Fano test. Methods: The PENELOPE Monte Carlo package was used to simulate a PR06C ionization chamber in 0.35 T through 1.5 T magnetic fields oriented either parallel or orthogonal to an incident 6 MV radiation beam. The ionization chamber was rotated through a number of polar and azimuthal angles. The dose deposited within the chamber at each angular position and magnetic field strength was scored then normalized to that deposited in the same orientation with no magnetic field. The simulation was also benchmarked via a Fano test in magnetic field. Results: The Fano test yielded a 0.4% difference between simulation and expected result, which is similar to previous findings and sufficient for the purposes of this study. The angular dose-response map in all cases where the magnetic field is oriented orthogonal to the radiation beam is quite varied and can range from 0.89 to 1.08. Angular deviations as small as 3 degrees can lead to dose-response changes in excess of 1%. When the magnetic field is parallel to the photon beam, the angular dose-response map is homogeneous and less than 1% below 1.0 T. Conclusions: Within a magnetic field-oriented orthogonal to the radiation beam, the ionization chamber dose-response fluctuates greatly as a function of polar and azimuthal angle, where a parallel field yields a more homogeneous dose-response.(C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4322 / 4328
页数:7
相关论文
共 50 条
  • [1] Ion chamber quality and magnetic field conversion factors and angular dependence in magnetic fields
    Malkov, Victor
    Rogers, David W. O.
    MEDICAL PHYSICS, 2017, 44 (08) : 4371 - 4371
  • [2] Technical Note: Sensitive volume effects on ion chamber responses in longitudinal magnetic fields
    Reynolds, Michael
    Rathee, Satyapal
    Fallone, B. Gino
    MEDICAL PHYSICS, 2019, 46 (07) : 3306 - 3310
  • [3] Determination of the angular dependence of a Pin Point ion chamber
    Prado, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S966 - S967
  • [4] Technical Note: Consistency of PTW30013 and FC65-G ion chamber magnetic field correction factors
    Woodings, S. J.
    van Asselen, B.
    van Soest, T. L.
    de Prez, L. A.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    Wolthaus, J. W. H.
    MEDICAL PHYSICS, 2019, 46 (08) : 3739 - 3745
  • [5] Angular Dependence of Dosimeters in the Magnetic Field
    Funke, L.
    Hering, D. A.
    Glashoerster, M.
    Channaoui, M.
    Khoukaz, A.
    Haverkamp, U.
    Eich, H. T.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2020, 196 (SUPPL 1) : S66 - S66
  • [6] Determination of the angular dependence of a CC04 ion chamber
    Prado, A.
    Gaitan, A.
    Manzano, M.
    Leonor, M.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S910 - S911
  • [7] Resonant angular dependence in a weak magnetic field
    Granot, E
    Azbel, MY
    JOURNAL OF PHYSICS-CONDENSED MATTER, 1999, 11 (20) : 4031 - 4038
  • [8] Resonant angular dependence in a weak magnetic field
    J Phys Condens Matter, 20 (4031-4038):
  • [9] Technical Note: Angular dependence of a 2D monolithic silicon diode array for small field dosimetry
    Stansook, Nauljun
    Utitsarn, Kananan
    Petasecca, Marco
    Newall, Matthew K.
    Duncan, Mitchell
    Nitschke, Kym
    Carolan, Martin
    Metcalfe, Peter
    Lerch, Michael L. F.
    Perevertaylo, Vladimir L.
    Tome, Wolfgang A.
    Rosenfeld, Anatoly B.
    MEDICAL PHYSICS, 2017, 44 (08) : 4313 - 4321
  • [10] Measurement of Ionization Chamber Angular Response to Magnetic Field in An MRLinac
    Iakovenko, V.
    Keller, B.
    Sahgal, A.
    Sarfehnia, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E107 - E107