摘要
Patients who undergo interventional radiological examinations and the medical doctors who perform them receive a noticeable radiation dose. Twenty-six angiographies and six angioplasties were included in the present study. They were classified as examinations of lower limbs, abdominal aorta and aortic arch/carotid artery. Thermoluminescent dosemeters (TLDs) were placed underneath and over the lead apron of the doctors at all of the 32 examinations as well as next to the eyes and over the thyroid of 22 patients. Dose area product (DAP) values, time duration and other parameters were registered. Patients' effective dose (ED) and doses to the various organs were calculated with the aid of ODS-60 software. The EDs were normalised to DAP measured in each procedure. Based on TLD measurements, the Niklason method was applied for the calculation of doctors' ED. The calculated ED by ODS-60 were in the range of 0.00-1.46, 2.63-49.32 and 0.07-45.12 mSv for the three groups, respectively, while E/DAP indices were 0.023, 0.310 and 0.105 mSv Gy(-1) cm(-2). Very good correlation was found between TLD measurements of the eye and the thyroid of the patients and the relative values calculated by ODS-60. The ED for the radiologists ranged from 0.4 to 47.0 mu Sv for all the procedures. Taking into consideration the annual number of examinations performed in the department, the estimated dose to the radiologists' eyes is considerably high, so wearing leaded glasses is recommended for the optimisation of the procedure.
引用
收藏
页码:86 / 89
页数:4
相关论文
共 50 条
  • [21] Patient radiation exposure during diagnostic and therapeutic interventional neuroradiology procedures
    Alexander, M. D.
    Oliff, M. C.
    Olorunsola, O. G.
    Brus-Ramer, M.
    Nickoloff, E. L.
    Meyers, P. M.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2010, 2 (01) : 6 - 10
  • [22] Radiation doses to the legs of radiologists performing interventional procedures: are they a cause for concern?
    Whitby, M
    Martin, CJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (905): : 321 - 327
  • [23] Radiation exposure to patients undergoing diagnostic and interventional cardiac catheterization procedures
    Bakalyar, DM
    Castellani, MD
    Safian, RD
    [J]. CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1997, 42 (02): : 121 - 125
  • [24] Reduction of the radiation exposure of patients caused by selected interventional and angiographic procedures
    Seifert, H
    El-Jamal, A
    Roth, R
    Urbanczyk, K
    Kramann, B
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (12): : 1057 - 1064
  • [25] RADIATION EXPOSURE OF PATIENTS AND OPERATORS DURING INTERVENTIONAL RADIOLOGY
    KRAHE, T
    EWEN, K
    LACKNER, K
    KOSTER, O
    NICOLAS, V
    [J]. FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1986, 145 (02): : 217 - 220
  • [26] Radiation exposure of patients during endourological procedures
    De Coninck, Vincent
    Mortiers, Xavier
    Hendrickx, Laura
    De Wachter, Stefan
    Traxer, Olivier
    Keller, Etienne X.
    [J]. WORLD JOURNAL OF UROLOGY, 2024, 42 (01)
  • [27] EYE LENS RADIATION EXPOSURE OF WORKERS DURING MEDICAL INTERVENTIONAL PROCEDURES AND SURGERY
    Bera, Geraldine
    Gellie, Gabriel
    Jamet, Emilie
    Entine, Fabrice
    Michel, Xavier
    [J]. RADIATION PROTECTION DOSIMETRY, 2018, 182 (03) : 323 - 328
  • [28] Shielding Effect of Lead Glasses on Radiologists' Eye Lens Exposure in Interventional Procedures
    Hu, Panpan
    Kong, Yan
    Chen, Bo
    Liu, Qianqian
    Zhuo, Weihai
    Liu, Haikuan
    [J]. RADIATION PROTECTION DOSIMETRY, 2017, 174 (01) : 136 - 140
  • [29] CT fluoroscopy-guided interventional procedures: Techniques and radiation dose to radiologists
    Paulson, EK
    Sheafor, DH
    Enterline, DS
    McAdams, HP
    Yoshizumi, TT
    [J]. RADIOLOGY, 2001, 220 (01) : 161 - 167
  • [30] Radiation exposure to cardiologists performing interventional cardiology procedures
    Delichas, M
    Psarrakos, K
    Molyvda-Athanassopoulou, E
    Giannoglou, G
    Sioundas, A
    Hatziioannou, K
    Papanastassiou, E
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2003, 48 (03) : 268 - 273