To study the shielding effect of radiologists' eye lens with lead glasses of different equivalent thicknesses and sizes in interventional radiology procedures. Using the human voxel phantom with a more accurate model of the eye and MCNPX software, eye lens doses of the radiologists who wearing different kinds of lead glasses were simulated, different beam projections were taken into consideration during the simulation. Measurements were also performed with the physical model to verify simulation results. Simulation results showed that the eye lens doses were reduced by a factor from 3 to 9 when wearing a 20 cm(2)-sized lead glasses with the equivalent thickness ranging from 0.1 to 1.0 mm Pb. The increase of dose reduction factor (DRF) was not significant whenever increase the lead equivalent of glasses of which larger than 0.35 mm. Furthermore, the DRF was proportional to the size of glass lens from 6 to 30 cm(2) with the same lead equivalent. The simulation results were in well agreements with the measured ones. For more reasonable and effective protection of the eye lens of interventional radiologists, a pair of glasses with a lead equivalent of 0.5 mm Pb and large-sized (at least 27 cm(2) per glass) lens are recommended.
引用
收藏
页码:136 / 140
页数:5
相关论文
共 50 条
  • [1] Eye Lens Dose Reduction of Radiologists in Interventional Procedures with Lead Glasses
    Hu, P.
    Zhuo, W.
    Liu, H.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3719 - 3719
  • [2] EYE LENS RADIATION EXPOSURE OF WORKERS DURING MEDICAL INTERVENTIONAL PROCEDURES AND SURGERY
    Bera, Geraldine
    Gellie, Gabriel
    Jamet, Emilie
    Entine, Fabrice
    Michel, Xavier
    [J]. RADIATION PROTECTION DOSIMETRY, 2018, 182 (03) : 323 - 328
  • [3] RADIATION EXPOSURE TO PATIENTS AND RADIOLOGISTS DURING INTERVENTIONAL PROCEDURES
    Molyvda-Athanasopoulou, E.
    Karlatira, M.
    Gotzamani-Psarrakou, A.
    Koulouris, Ch.
    Siountas, A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) : 86 - 89
  • [4] Occupational dose constraints for the lens of the eye for interventional radiologists and interventional cardiologists in the UK
    Mairs, William D. A.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1062):
  • [5] A study of the underestimation of eye lens dose with current eye dosemeters for interventional clinicians wearing lead glasses
    da Silva, Edilaine Honorio
    Martin, Colin J.
    Vanhavere, Filip
    Buls, Nico
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (01) : 215 - 224
  • [6] Occupational Exposure of the Eye Lens in Interventional Procedures: How to Assess and Manage Radiation Dose
    Ciraj-Bjelac, Olivera
    Carinou, Eleftheria
    Ferrari, Paolo
    Gingaume, Merce
    Merce, Marta Sans
    O'Connor, Una
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2016, 13 (11) : 1347 - 1353
  • [7] Lateral Head Shielding for Eye Lens Dose Reduction for Neuro-CBCT and Interventional Fluoroscopic Procedures
    Xiong, Z.
    Vijayan, S.
    Guo, C.
    Choudhary, P.
    Rudin, S.
    Bednarek, D.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [8] Dose estimation of eye lens for interventional procedures in diagnosis
    Liu, Yu-Rong
    Huang, Chia-Yu
    Hsu, Ching-Han
    Hsu, Fang-Yuh
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 247 - 251
  • [9] EXAMINATION OF LENS EXPOSURE DOSES OF CARDIOLOGISTS, NEUROSURGEONS AND RADIOLOGISTS IN INTERVENTIONAL RADIOLOGY
    Osawa, Mitsuharu
    Tawada, Yuka
    Kaneda, Naoki
    Fujita, Naotoshi
    Koyama, Shuji
    [J]. RADIATION PROTECTION DOSIMETRY, 2022, 198 (20) : 1585 - 1597
  • [10] Exposure of the eye lens and brain for interventional cardiology staff
    Kidon, Joanna
    Polaczek-Grelik, Kinga
    Zurek, Przemyslaw
    Wojakowski, Wojciech
    Ochala, Andrzej
    [J]. POSTEPY W KARDIOLOGII INTERWENCYJNEJ, 2021, 17 (03): : 298 - 304