Long-term comparison of Charnley and Stanmore design total hip replacements

被引:48
|
作者
Britton, AR [1 ]
Murray, DW [1 ]
Bulstrode, CJ [1 ]
McPherson, K [1 ]
Denham, RA [1 ]
机构
[1] NUFFIELD ORTHOPAED CTR,DEPT ORTHOPAED SURG,OXFORD OX3 7LD,ENGLAND
来源
关键词
D O I
10.1302/0301-620X.78B5.0780802
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We reviewed the records of the long-term outcome of 208 Charnley and 982 Stanmore total hip replacements (THR) performed by or under the supervision of one surgeon from 1973 to 1987, The Stanmore implant had a better survival rate before revision at 14 years (86% to 79%, p=0.004), but the difference only became apparent at ten years, The later Stanmore implants did better than the early ones (97% to 92% at ten years, p=0,005), the improvement coinciding with the introduction of a new cementing technique using a gun, Most of the Charnley implants were done before most of the Stanmore implants so that the difference between the results may in part be explained by improved methods, but this is not the complete explanation since a difference persisted for implants carried out during the same period of time, We conclude that improved techniques have reduced failure rates substantially. This improvement was much greater than that observed between these two designs of implant, Proof of the difference would require a very large randomised controlled trial over a ten-year period.
引用
收藏
页码:802 / 808
页数:7
相关论文
共 50 条
  • [12] 351 TOTAL HIP REPLACEMENTS ACCORDING TO CHARNLEY - A REVIEW OF COMPLICATIONS AND FUNCTION
    CARLSSON, AS
    ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (03): : 339 - 344
  • [13] A Matched Comparison of the Long-Term Outcomes of Cemented and Cementless Total Knee Replacements
    Mohammad, Hasan R.
    Judge, Andrew
    Murray, David W.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2021, 103 (24): : 2270 - 2280
  • [14] LONG-TERM RESULTS OF CHARNLEY LOW-FRICTION ARTHROPLASTY OF THE HIP
    DALL, DM
    LEARMONTH, ID
    DALL, G
    RECORD, C
    SOUTH AFRICAN MEDICAL JOURNAL, 1988, 73 (02): : 89 - 92
  • [15] CHARNLEY LOW-FRICTION ARTHROPLASTY OF THE HIP - LONG-TERM RESULTS
    WROBLEWSKI, BM
    SINEY, PD
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (292) : 191 - 201
  • [16] POST-OPERATIVE COMPLICATIONS AFTER 605 CHARNLEY TOTAL HIP REPLACEMENTS
    DYREBORG, E
    JORGENSEN, J
    KROGH, P
    PETERSEN, EK
    ROHR, R
    SCHLANBUSCH, C
    ACTA ORTHOPAEDICA SCANDINAVICA, 1979, 50 (03): : 347 - 347
  • [17] LONG-TERM RESULTS OF CHARNLEY LOW-FRICTION ARTHROPLASTY OF THE HIP
    DALL, DM
    LEARMONTH, I
    SOLOMON, M
    DAVENPORT, M
    SOUTH AFRICAN JOURNAL OF SURGERY, 1992, 30 (04) : 171 - 174
  • [18] Long-term followup of nonmodular total knee replacements
    Keating, EM
    Meding, JB
    Faris, PM
    Ritter, MA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (404) : 34 - 39
  • [19] THE REVIEW OF 4300 CHARNLEY TOTAL HIP REPLACEMENTS INSERTED BETWEEN 1968 AND 1979
    PICAULT, C
    MICHEL, CR
    VIDIL, R
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1980, 66 (02): : 57 - 67
  • [20] LONG-TERM RESULTS OF CHARNLEY TOTAL HIP-REPLACEMENT - REVIEW OF 92 PATIENTS AT 15 TO 20 YEARS
    NEUMANN, L
    FREUND, KG
    SORENSON, KH
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (02): : 245 - 251