Comparison among Ocular Response Analyzer, Corvis ST and Goldmann applanation tonometry in healthy children

被引:12
|
作者
Salouti, Ramin [1 ]
Alishiri, Ali Agha [2 ]
Gharebaghi, Reza [1 ]
Naderi, Mostafa [2 ]
Jadidi, Khosmw [2 ]
Shojaei-Baghini, Ahmad [2 ]
Talebnejad, Mohammadreza [1 ]
Nasiri, Zahra [3 ]
Hosseini, Seyedmorteza [2 ]
Heidary, Fatemeh [4 ,5 ]
机构
[1] Shiraz Univ Med Sci, Dept Ophthalmol, Poostchi Ophthalmol Res Ctr, Shiraz 7134814336, Iran
[2] Baqiyatallah Univ Med Sci, Dept Ophthalmol, Tehran 141551856, Iran
[3] Tarbiat Modares Univ, Fac Math Sci, Tehran 141551856, Iran
[4] Shahid Beheshti Univ Med Sci, Ophthalm Res Ctr, Tehran 141551856, Iran
[5] Shahed Univ, Inununoregulat Res Ctr, Tehran 141551856, Iran
关键词
Goldmann applanation tonometer; Ocular Response Analyzer; Corvis ST; intraocular pressure; children; CORNEAL BIOMECHANICAL PROPERTIES; INTRAOCULAR-PRESSURE MEASUREMENT; THICKNESS;
D O I
10.18240/ijo.2018.08.13
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
AIM: To explore the relationship between different parameters of Ocular Response Analyzer (ORA) and Corvis ST (CST) in a sample of healthy Iranian school-aged children and the relationship between parameters of these 2 instruments against intraocular pressure (IOP), measured by the Goldmann applanation tonometer (GAT-IOP), age and gender, and find possible correlation between ORA and CST with GAT. METHODS: This cross-sectional study included 90 healthy children. A general interview and complete eye examination were performed. Following successful GAT-IOP measurement, ORA and CST were conducted. The CST parameters were A 1/2 length (A1L, A2L), A 1/2 velocity (A1V, A2V), highest concavity deformation amplitude (HCDA), radius of curvature (RoC), peak distance (PD), central corneal thickness (CCT) and IOP. The ORA parameters were corneal hysteresis (CH), corneal resistance factor (CRF), Goldmann-correlated IOP (IOP-G) and corneal compensated IOP (IOP-CC). Extracted data was analyzed using the Statistical Package for Social Science software. RESULTS: Totally 39 males with age of 9.08 +/- 1.60 (6-12) y and 51 females with age of 8.96 +/- 1.55 (6-13) y were included. Many CST parameters were significantly correlated with CH, CRF, IOP-G and IOP-CC. Some CST parameters had a significant correlation with GAT-IOP, including IOP-CST in both eyes and HCDA, A2L, PD, and RoC in the left eye, but none with age, except A2L in the right eye. The CRF measurement showed a significant correlation with GAT-IOP in both eyes and CH in the right eye, yet, none with age. Among all CST and ORA parameters, CCT-CST in both eyes and A1L in right eye had a significant correlation with gender, although this was a negligible negative correlation. Comparison of mean IOP values by different devices showed a significantly highest IOP overestimation by CST and lowest by IOP-CC compared with GAT. Also, IOP-G versus IOP-CST significantly had the lowest IOP overestimation among others. Overall, either low positive correlation or negligible correlation was found between IOP measurements by 3 instruments. CONCLUSION: The study finds the highest IOP overestimation by CST and lowest by IOP-CC compared with GAT. Overall, either low positive correlation or negligible correlation is found between IOP measurements by the 3 instruments.
引用
收藏
页码:1330 / 1336
页数:7
相关论文
共 50 条
  • [31] Comparison of dynamic contour tonometry and Goldmann applanation tonometry in glaucoma patients and healthy subjects
    Barleon, Lorenz
    Hoffmann, Esther M.
    Berres, Manfred
    Pfeiffer, Norbert
    Grus, Franz H.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 142 (04) : 583 - 590
  • [32] Measurement of intraocular pressure on LASIK and LASEK patients using the Reichert ocular response analyzer and Goldmann applanation tonometry
    Kirwan, Caitriona
    O'Keefe, Michael
    JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (04) : 366 - 370
  • [33] IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry
    Yang Zhang
    Lin Zheng
    Ailing Bian
    Qi Zhou
    International Ophthalmology, 2018, 38 : 697 - 703
  • [34] Effectiveness of the Goldmann Applanation Tonometer, the Dynamic Contour Tonometer, the Ocular Response Analyzer and the Corvis ST in Measuring Intraocular Pressure following FS-LASIK
    Bao, Fangjun
    Huang, Wei
    Zhu, Rong
    Lu, Nanji
    Wang, Yuan
    Li, Hechen
    Wu, Songan
    Lin, Huini
    Wang, Junjie
    Zheng, Xiaobo
    Huang, JinHai
    Li, Yiyu
    Wang, Qinmei
    Elsheikh, Ahmed
    CURRENT EYE RESEARCH, 2020, 45 (02) : 144 - 152
  • [35] IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry
    Zhang, Yang
    Zheng, Lin
    Bian, Ailing
    Zhou, Qi
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (02) : 697 - 703
  • [36] Comparison of handheld rebound tonometry with Goldmann applanation tonometry in children with glaucoma: a cohort study
    Dahlmann-Noor, Annegret Hella
    Puertas, Renata
    Tabasa-Lim, Shenille
    El-Karmouty, Ahmed
    Kadhim, Mustafa
    Wride, Nicholas Kloster
    Lewis, Amanda
    Grosvenor, Dawn
    Rai, Poornima
    Papadopoulos, Maria
    Brookes, John
    Bunce, Catey
    Khaw, Peng Tee
    BMJ OPEN, 2013, 3 (04):
  • [37] Objective tonometry: A comparison of three methods with Goldmann Applanation tonometry
    White, ET
    Tonnu, PAA
    Sharma, K
    Garway-Heath, DF
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U54 - U54
  • [38] Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Keratoconus
    Unterlauft, Jan Darius
    Schaedle, Nina
    Kasper, Karsten
    Klink, Thomas
    Geerling, Gerd
    CORNEA, 2011, 30 (10) : 1078 - 1082
  • [39] A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry
    Schreiber, W.
    Vorwerk, C. K.
    Langenbucher, A.
    Behrens-Baumann, W.
    Viestenz, A.
    OPHTHALMOLOGE, 2007, 104 (04): : 299 - 304
  • [40] Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry
    Luebke, Jan
    Bryniok, L.
    Neuburger, M.
    Jordan, J. F.
    Boehringer, D.
    Reinhard, T.
    Wecker, T.
    Anton, A.
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (11) : 2517 - 2521