Retinal adaptive optics imaging with a pyramid wavefront sensor

被引:11
|
作者
Brunner, Elisabeth [1 ]
Shatokhina, Julia [2 ]
Shirazi, Muhammad Faizan [1 ]
Drexler, Wolfgang [1 ]
Leitgeb, Rainer [1 ]
Pollreisz, Andreas [3 ]
Hitzenberger, Christoph K.
Ramlau, Ronny [1 ,2 ,4 ]
Pircher, Michael [1 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, Waehringer Guertel 18-20, A-1090 Vienna, Austria
[2] Johann Radon Inst Computat & Appl Math, Altenbergerstr 69, A-4040 Linz, Austria
[3] Med Univ Vienna, Dept Ophthalmol & Optometry, Waehringer Guertel 18-20, A-1090 Vienna, Austria
[4] Johannes Kepler Univ Linz, Ind Math Inst, Altenbergerstr 69, A-4040 Linz, Austria
来源
BIOMEDICAL OPTICS EXPRESS | 2021年 / 12卷 / 10期
基金
奥地利科学基金会; 欧盟地平线“2020”;
关键词
PUPIL;
D O I
10.1364/BOE.438915
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The pyramid wavefront sensor (P-WFS) has replaced the Shack-Har tmann (SH-) WFS as the sensor of choice for high-performance adaptive optics (AO) systems in astronomy. Many advantages of the P-WFS, such as its adjustable pupil sampling and superior sensitivity, are potentially of great benefit for AO-supported imaging in ophthalmology as well. However, so far no high quality ophthalmic AO imaging was achieved using this novel sensor. Usually, a P-WFS requires modulation and high precision optics that lead to high complexity and costs of the sensor. These factors limit the competitiveness of the P-WFS with respect to other WFS devices for AO correction in visual science. Here, we present a cost-effective realization of AO correction with a non-modulated P-WFS based on standard components and apply this technique to human retinal in vivo imaging using optical coherence tomography (OCT). P-WFS based high quality AO imaging was successfully performed in 5 healthy subjects and smallest retinal cells such as central foveal cone photoreceptors are visualized. The robustness and versatility of the sensor is demonstrated in the model eye under various conditions and in vivo by high-resolution imaging of other structures in the retina using standard and extended fields of view. As a quality benchmark, the performance of conventional SH-WFS based AO was used and successfully met. This work may trigger a paradigm shift with respect to the wavefront sensor of choice for AO in ophthalmic imaging.
引用
收藏
页码:5969 / 5990
页数:22
相关论文
共 50 条
  • [21] Dimensioning the Gravity adaptive optics wavefront sensor
    Clenet, Y.
    Gendron, E.
    Rousset, G.
    Hippler, S.
    Eisenhauer, F.
    Gillessen, S.
    Perrin, G.
    Amorim, A.
    Brandner, W.
    Perraut, K.
    Straubmeier, C.
    ADAPTIVE OPTICS SYSTEMS II, 2010, 7736
  • [22] Testing the pyramid wavefront sensor without modulation used in the closed-loop adaptive optics system
    Wang, Shengqian
    Rao, Changhui
    Zhang, Ang
    Zhang, Xuejun
    Wei, Kai
    Tian, Yu
    Liao, Zhou
    Zhang, Cheng
    Xian, Hao
    Zhang, Xiaojun
    Wei, Ling
    ADAPTIVE OPTICS SYSTEMS III, 2012, 8447
  • [23] Laboratory demonstrations on a pyramid wavefront sensor without modulation for closed-loop adaptive optics system
    Wang, Shengqian
    Rao, Changhui
    Xian, Hao
    Zhang, Jianlin
    Wang, Jianxin
    Liu, Zheng
    OPTICS EXPRESS, 2011, 19 (09): : 8135 - 8150
  • [24] Three-sided pyramid wavefront sensor, part 1: simulations and analysis for astronomical adaptive optics
    Schatz, Lauren
    Males, Jared R.
    Correia, Carlos
    Neichel, Benoit
    Chambouleyron, Vincent
    Codona, Johanan
    Fauvarque, Olivier
    Sauvage, Jean-Francois
    Fusco, Thierry
    Hart, Michael
    Janin-Potiron, Pierre
    Johnson, Robert
    Long, Joseph D.
    Mateen, Mala
    JOURNAL OF ASTRONOMICAL TELESCOPES INSTRUMENTS AND SYSTEMS, 2021, 7 (04)
  • [25] Near-infrared pyramid wavefront sensor for Keck adaptive optics: opto-mechanical design
    Lilley, Scott J.
    Wizinowich, Peter
    Mawet, Dimitri
    Chun, Mark
    Bond, Charlotte Z.
    Wallace, James K.
    Jovanovic, Nemanja
    Delorme, Jacques-Robert
    Jacobson, Shane M.
    Taheri, Mojtaba
    Vandenberg, Adam
    ADAPTIVE OPTICS SYSTEMS VI, 2018, 10703
  • [26] Adaptive optics and retinal imaging
    Roorda, A
    Williams, DR
    VISION SCIENCE AND ITS APPLICATIONS, PROCEEDINGS, 2000, 35 : 151 - 162
  • [27] Retinal imaging with adaptive optics
    Le Gargasson, JF
    Glanc, M
    Léna, P
    COMPTES RENDUS DE L ACADEMIE DES SCIENCES SERIE IV PHYSIQUE ASTROPHYSIQUE, 2001, 2 (08): : 1131 - 1138
  • [28] Wavefront sensorless adaptive optics optical coherence tomography for in vivo retinal imaging in mice
    Jian, Yifan
    Xu, Jing
    Gradowski, Martin A.
    Bonora, Stefano
    Zawadzki, Robert J.
    Sarunic, Marinko V.
    BIOMEDICAL OPTICS EXPRESS, 2014, 5 (02): : 547 - 559
  • [29] High-accuracy wavefront control for retinal imaging with Adaptive-Influence-Matrix Adaptive Optics
    Zou, Weiyao
    Burns, Stephen A.
    OPTICS EXPRESS, 2009, 17 (22): : 20167 - 20177
  • [30] Semianalytical error budget for adaptive optics systems with pyramid wavefront sensors
    Agapito, Guido
    Pinna, Enrico
    JOURNAL OF ASTRONOMICAL TELESCOPES INSTRUMENTS AND SYSTEMS, 2019, 5 (04)