Induction ovens and electromagnetic interference: What is the risk for patients with implanted pacemakers?

被引:17
|
作者
Rickli, H
Facchini, M
Brunner, H
Ammann, P
Sagmeister, M
Klaus, G
Angehrn, W
Luechinger, R
Duru, F
机构
[1] Kantonsspital, Div Cardiol, CH-9007 St Gallen, Switzerland
[2] Univ Zurich Hosp, Div Cardiol, CH-8091 Zurich, Switzerland
[3] ETH, Swiss Fed Inst Technol, Inst Biomed Engn, Zurich, Switzerland
[4] Maxwave, Stafay, Switzerland
[5] Univ Zurich, CH-8006 Zurich, Switzerland
来源
关键词
electromagnetic interference; pacemaker; pacing; induction oven;
D O I
10.1046/j.1460-9592.2003.t01-1-00216.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Electromagnetic fields may interfere with normal pacemaker function. Despite the introduction of modern pacemakers and bipolar lead systems, electromagnetic interference (EMI) still remains to be a concern during daily lives when patients are exposed to cellular phones, electronic security systems, and several household appliances. The aim of this study was to evaluate potential EMI risk of induction ovens, which are increasingly used in private households. The study included 40 consecutive patients (22 men, 18 women; age 73 +/- 11 years) with implanted DDD, VVI, VDD, and AAI pacemaker systems. The pacemakers were programmed to unipolor sensing and pacing. Sensitivity remained unchanged, if the measured sensing threshold was more than twice the programmed value; otherwise, it was set at half of the measured sensing threshold. Patients were placed in a sitting position at the closest possible distance of about 20 cm between two cooking pots and pacemaker bending the upper part of the body slightly over the induction oven. The energy was increased stepwise to the maximum. One pot was removed and placed again at the highest oven level. Potential interference was monitored continuously. The study showed no incidence of pacemaker malfunction during the entire test while the patients with intrinsic cardiac rhythms were exposed to the induction oven at varying energy strengths. Likewise, there was no external interference when the patients were paced at heart rates of 10-15 beats/min above their heart rates. The programmed parameters remained unchanged after the study. In conclusion, this study shows no EMI risk of an induction oven in patients with bipolar or right-sided unipolar pacemakers.
引用
收藏
页码:1494 / 1497
页数:4
相关论文
共 50 条
  • [1] Induction ovens and electromagnetic interference: What is the risk for patients with implantable cardioverter defibrillators?
    Binggeli, C
    Rickli, H
    Ammann, P
    Brunckhorst, C
    Hufschmid, U
    Luechinger, R
    Duru, F
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (04) : 399 - 401
  • [2] Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators
    Niehaus, M
    Tebbenjohanns, J
    [J]. HEART, 2001, 86 (03) : 246 - 248
  • [3] MICROWAVE OVENS - POTENTIAL RISK TO PATIENTS WITH CARDIAC PACEMAKERS
    REFSHAUGE, WD
    [J]. MEDICAL JOURNAL OF AUSTRALIA, 1971, 1 (09) : 498 - +
  • [4] A model for predicting electromagnetic interference of implanted cardiac pacemakers by mobile telephones
    Wang, JQ
    Fujiwara, O
    Nojima, T
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2000, 48 (11) : 2121 - 2125
  • [5] INTERFERENCE OF IMPLANTED CARDIAC PACEMAKERS
    SCHULTEN, KH
    SMEKAL, PV
    ROHRIG, FR
    BALDUS, O
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1972, 97 (41) : 1539 - &
  • [6] Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven
    Hirose, M
    Hida, M
    Sato, E
    Kokubo, K
    Nie, M
    Kobayashi, H
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (06): : 540 - 548
  • [7] Concerns about sources of electromagnetic interference in patients with pacemakers
    Sakakibara, Y
    Mitsui, T
    [J]. JAPANESE HEART JOURNAL, 1999, 40 (06): : 737 - 743
  • [8] ELECTROSURGICAL DEVICE INTERFERENCE WITH IMPLANTED PACEMAKERS
    FURMAN, S
    PARKER, B
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1978, 239 (18): : 1910 - 1910
  • [9] INTERFERENCE OF IMPLANTED PACEMAKERS IN EVERYDAY LIFE
    BISPING, HJ
    EFFERT, S
    IRNICH, W
    MEYER, J
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1972, 97 (46) : 1773 - &
  • [10] INTERFERENCE OF MOBILE PHONES WITH FUNCTION OF IMPLANTED PACEMAKERS - HOW SIGNIFICANT IS THE RISK
    MEISEL, E
    KOPSCEK, H
    KLINGHAMMER, L
    DANIEL, WG
    [J]. CIRCULATION, 1995, 92 (08) : 3546 - 3546