Real-Time Quantification of Amino Acids in the Exhalome by Secondary Electrospray Ionization-Mass Spectrometry: A Proof-of-Principle Study

被引:33
|
作者
Garcia-Gomez, Diego [1 ]
Gaisl, Thomas [2 ,3 ]
Bregy, Lukas [1 ]
Cremonesi, Alessio [4 ]
Sinues, Pablo Martinez-Lozano [1 ]
Kohler, Malcolm [2 ,5 ]
Zenobi, Renato [1 ]
机构
[1] Swiss Fed Inst Technol, Dept Chem & Appl Biosci, Zurich, Switzerland
[2] Univ Zurich Hosp, Dept Pulmonol, Zurich, Switzerland
[3] Hosp Zollikerberg, Zollikerberg, Switzerland
[4] Univ Childrens Hosp Zurich, Div Clin Chem & Biochem, Zurich, Switzerland
[5] Univ Zurich, Ctr Integrat Human Physiol, Zurich, Switzerland
基金
瑞士国家科学基金会;
关键词
INSULIN-RESISTANCE; LIQUID-CHROMATOGRAPHY; BREATH; PROFILES; CHILDREN; FUTURE; RISK;
D O I
10.1373/clinchem.2016.256909
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Amino acids are frequently determined in clinical chemistry. However, current analysis methods are time-consuming, invasive, and suffer from artifacts during sampling, sample handling, and sample preparation. We hypothesized in this proof-of-principle study that plasma concentrations of amino acids can be estimated by measuring their concentrations in exhaled breath. A novel breath analysis technique described here allows such measurements to be carried out in real-time and noninvasively, which should facilitate efficient diagnostics and give insights into human physiology. METHODS: The amino acid profiles in 37 individuals were determined by ion-exchange HPLC in blood plasma and simultaneously in breath by secondary electrospray ionization coupled to high-resolution mass spectrometry. Participants were split into training and test sets to validate the analytical accuracy. Longitudinal profiles in 3 individuals were additionally obtained over a 12-h period. RESULTS: Concentrations of 8 slightly volatile amino acids (A, V, I, G, P, K, F, Orn) could be determined in exhaled breath with a CV of <10%. Exhalome validation studies yielded high accuracies for each of these amino acids, on average only 3% less compared to plasma concentrations (95% CI +/- 13%). Higher variations were found only for amino acids with a low plasma concentration. CONCLUSIONS: This study demonstrates for the first time that amino acids can be quantified in the human breath and that their concentrations correlate with plasma concentrations. Although this noninvasive technique needs further investigation, exhalome analysis may provide significant benefits over traditional, offline analytical methods. (C) 2016 American Association for Clinical Chemistry
引用
收藏
页码:1230 / 1237
页数:8
相关论文
共 50 条
  • [41] Preliminary study of the analysis of oligogalacturonic acids by electrospray ionization mass spectrometry
    Zhu, L
    Lee, HK
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (12) : 975 - 978
  • [42] Conceptual design and proof-of-principle testing of the real-time multispectral imaging system MANTIS
    Vijvers, W. A. J.
    Mumgaard, R. T.
    Andrebe, Y.
    Classen, I. G. J.
    Duval, B. P.
    Lipschultz, B.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [43] Real-Time Hydrogen/Deuterium Exchange Kinetics via Supercharged Electrospray Ionization Tandem Mass Spectrometry
    Sterling, Harry J.
    Williams, Evan R.
    ANALYTICAL CHEMISTRY, 2010, 82 (21) : 9050 - 9057
  • [44] Real-time hydrogen/deuterium exchange kinetics via supercharged electrospray ionization tandem mass spectrometry
    Department of Chemistry, University of California, Berkeley, CA 94720-1460, United States
    Anal. Chem., 21 (9050-9057):
  • [45] Real-time toxicity prediction of Aconitum stewing system using extractive electrospray ionization mass spectrometry
    Zi-Dong Qiu
    Jin-Long Chen
    Wen Zeng
    Ying Ma
    Tong Chen
    Jin-Fu Tang
    Chang-Jiang-Sheng Lai
    Lu-Qi Huang
    ActaPharmaceuticaSinicaB, 2020, 10 (05) : 903 - 912
  • [46] Real-time analysis of metabolites in vivo by online extraction electrospray ionization mass spectrometry coupled to microdialysis
    Li, Tuo
    Li, Shumu
    Shi, Junwen
    Li, Xing
    Liu, Jianan
    Yang, Hui
    Wu, Wenjing
    Zhao, Lingyu
    Zhao, Zhenwen
    ANALYTICA CHIMICA ACTA, 2022, 1205
  • [47] Real-time toxicity prediction of Aconitum stewing system using extractive electrospray ionization mass spectrometry
    Qiu, Zi-Dong
    Chen, Jin-Long
    Zeng, Wen
    Ma, Ying
    Chen, Tong
    Tang, Jin-Fu
    Lai, Chang-Jiang-Sheng
    Huang, Lu-Qi
    ACTA PHARMACEUTICA SINICA B, 2020, 10 (05) : 903 - 912
  • [48] Advanced oxidation of caffeine in water:: On-line and real-time monitoring by electrospray ionization mass spectrometry
    Dalmázio, I
    Santos, LS
    Lopes, RP
    Eberlin, MN
    Augusti, R
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2005, 39 (16) : 5982 - 5988
  • [49] Direct Detection of Amino Acids Using Extractive Electrospray Ionization Tandem Mass Spectrometry
    Xu Ning
    Zhu Zhi-Qiang
    Yang Shui-Ping
    Wang Jiang
    Gu Hai-Wei
    Zhou Zhen
    Chen Huan-Wen
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2013, 41 (04) : 523 - 528
  • [50] Electrospray ionization tandem mass spectrometry differentiation of N-phosphoryl-α-, β- and γ-amino acids
    Qiang, Liming
    Cao, Shuxia
    Zhao, Xiaoyang
    Mao, Xiangju
    Guo, Yanchun
    Liao, Xincheng
    Zhao, Yufen
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 266 (1-3) : 42 - 49