Automated direct comparison of two cryocooled 10 volt programmable Josephson voltage standards

被引:27
|
作者
Rufenacht, Alain [1 ]
Tang, Yi-hua [2 ]
Solve, Stephane [3 ]
Fox, Anna E. [1 ]
Dresselhaus, Paul D. [1 ]
Burroughs, Charles J. [1 ]
Schwall, Robert E. [1 ]
Chayramy, Regis [3 ]
Benz, Samuel P. [1 ]
机构
[1] NIST, Boulder, CO 80305 USA
[2] NIST, Gaithersburg, MD 20899 USA
[3] BIPM, Pavillon Breteuil, F-92312 Sevres, France
关键词
Digital-analog conversion; Josephson arrays; standards; superconducting integrated circuits; voltage measurement; AC POWER STANDARD; PLANCK CONSTANT; JUNCTIONS; BALANCE;
D O I
10.1088/1681-7575/aacbeb
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have performed direct dc comparisons between two cryocooled 10 V programmable Josephson voltage standards. Utilizing an automated synchronization scheme for the voltage reversals enables use of a high-sensitivity analog null detector on its 10 mu V range. No switches or manual operations are necessary to protect the null detector from overload signals. The agreement measured between the two voltage standards at 10 V is 9 parts in 10(12) with a relative combined uncertainty of 29 pV V-1 (k = 2 coverage factor). Since both systems can be operated floating from ground, various grounding configurations of the measurement circuit were investigated to evaluate possible leakage current paths to ground. Comparing the two systems under various grounding conditions enables minimization of leakage-current errors, optimization of system performance by revealing leakage-limiting components, and verification of key elements in the uncertainty budget for the measurement method(4).
引用
收藏
页码:585 / 596
页数:12
相关论文
共 50 条
  • [21] Direct Comparison between the NIST 10 V Conventional Josephson Voltage Standard and 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2011, 26 (04): : 339 - 348
  • [22] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    Tang, Y.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2009, 24 (04): : 207 - 214
  • [23] 10 VSINIS Josephson junction series arrays for programmable voltage standards
    Kohlmann, J
    Schulze, H
    Behr, R
    Müller, F
    Niemeyer, J
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (02) : 192 - 194
  • [24] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Shiv Kumar Jaiswal
    Y. Tang
    MAPAN, 2009, 24 : 207 - 214
  • [25] Direct comparison between the NIST 10 V Conventional Josephson Voltage Standard and 2.5 V Programmable Josephson Voltage Standard
    Shiv Kumar Jaiswal
    MAPAN, 2011, 26 : 339 - 348
  • [26] Direct Comparison of DC Josephson Voltage Standard and Programmable Josephson Voltage Standard of TUBITAK UME
    Turhan, Saliha
    Kanatoglu, Naylan
    Orhan, Recep
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [27] Direct comparison between a programmable and a conventional Josephson voltage standard at the level of 10V
    Djordjevic, S.
    Seron, O.
    Solve, S.
    Chayramy, R.
    METROLOGIA, 2008, 45 (04) : 429 - 435
  • [28] 10 Volt Programmable Josephson Voltage Standard Circuits Using NbSi-Barrier Junctions
    Dresselhaus, Paul D.
    Elsbury, Michael M.
    Olaya, David
    Burroughs, Charles J.
    Benz, Samuel P.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2011, 21 (03) : 693 - 696
  • [29] A Josephson Impedance Bridge Based on Programmable Josephson Voltage Standards
    Hagen, Thomas
    Palafox, Luis
    Behr, Ralf
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (06) : 1539 - 1545
  • [30] Cryogen-Free Operation of 10 V Programmable Josephson Voltage Standards
    Howe, L.
    Burroughs, C. J.
    Dresselhaus, P. D.
    Benz, S. P.
    Schwall, R. E.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2013, 23 (03)