First Clinical Photon-counting Detector CT System: Technical Evaluation

被引:267
|
作者
Rajendran, Kishore [1 ]
Petersilka, Martin [2 ]
Henning, Andre [2 ]
Shanblatt, Elisabeth R. [3 ]
Schmidt, Bernhard [2 ]
Flohr, Thomas G. [2 ]
Ferrero, Andrea [1 ]
Baffour, Francis [1 ]
Diehn, Felix E. [1 ]
Yu, Lifeng [1 ]
Rajiah, Prabhakar [1 ]
Fletcher, Joel G. [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Siemens Healthineers, Forchheim, Germany
[3] Siemens Med Solut, Malvern, PA USA
基金
美国国家卫生研究院;
关键词
COMPUTED-TOMOGRAPHY; NOISE; ACCURACY;
D O I
10.1148/radiol.212579
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The first clinical CT system to use photon-counting detector (PCD) technology has become available for patient care. Purpose: To assess the technical performance of the PCD CT system with use of phantoms and representative participant examinations. Materials and Methods: Institutional review board approval and written informed consent from four participants were obtained. Technical performance of a dual-source PCD CT system was measured for standard and high-spatial-resolution (HR) collimations. Noise power spectrum, modulation transfer function, section sensitivity profile, iodine CT number accuracy in virtual monoenergetic images (VMIs), and iodine concentration accuracy were measured. Four participants were enrolled (between May 2021 and August 2021) in this prospective study and scanned using similar or lower radiation doses as their respective clinical examinations performed on the same day using energy-integrating detector (EID) CT. Image quality and findings from the participants' PCD CT and EID CT examinations were compared. Results: All standard technical performance measures met accreditation and regulatory requirements. Relative to filtered back-projection reconstructions, images from iterative reconstruction had lower noise magnitude but preserved noise power spectrum shape and peak frequency. Maximum in-plane spatial resolutions of 125 and 208 mu m were measured for HR and standard PCD CT scans, respectively. Minimum values for section sensitivity profile full width at half maximum measurements were 0.34 mm (0.2-mm nominal section thickness) and 0.64 mm (0.4-mm nominal section thickness) for HR and standard PCD CT scans, respectively. In a 120-kV standard PCD CT scan of a 40-cm phantom, VMI iodine CT numbers had a mean percentage error of 5.7%, and iodine concentration had root mean squared error of 0.5 mg/cm(3), similar to previously reported values for EID CT. VMIs, iodine maps, and virtual noncontrast images were created for a coronary CT angiogram acquired with 66-msec temporal resolution. Participant PCD CT images showed up to 47% lower noise and/or improved spatial resolution compared with EID CT. Conclusion: Technical performance of clinical photon-counting detector (PCD) CT is improved relative to that of a current stateof-the-art CT system. The dual-source PCD geometry facilitated 66-msec temporal resolution multienergy cardiac imaging. Study participant images illustrated the effect of the improved technical performance. (C) RSNA, 2022
引用
收藏
页码:130 / 138
页数:9
相关论文
共 50 条
  • [31] Clinical CT Applications with Photon-Counting Detectors
    Levinson, R.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [32] Adaptive Spectral Inconsistency Modeling for Photon-Counting Detector CT
    Qi, Binxiang
    Gao, Hewei
    MEDICAL PHYSICS, 2020, 47 (06) : E377 - E377
  • [33] Photon-Counting Detector CT Angiography in Cervical Artery Dissection
    Keser, Zafer
    Diehn, Felix E.
    Lanzino, Giuseppe
    STROKE, 2024, 55 (03) : E48 - E49
  • [34] A methodology for incorporating a photon-counting CT system into routine clinical use
    Bache, Steven T.
    Samei, Ehsan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (08):
  • [35] Photon-counting detector CT for abdominal imaging: opportunities and challenges
    Schwartz, Fides Regina
    Alkadhi, Hatem
    EUROPEAN RADIOLOGY, 2023, 33 (11) : 7805 - 7806
  • [36] Investigation of abdominal artery delineation by photon-counting detector CT
    Ota, Takashi
    Onishi, Hiromitsu
    Itoh, Toshihide
    Fukui, Hideyuki
    Tsuboyama, Takahiro
    Nakamoto, Atsushi
    Enchi, Yukihiro
    Tatsumi, Mitsuaki
    Tomiyama, Noriyuki
    RADIOLOGIA MEDICA, 2024, : 1265 - 1274
  • [37] Virtual Noncontrast Abdominal Imaging with Photon-counting Detector CT
    Mergen, Victor
    Racine, Damien
    Jungblut, Lisa
    Sartoretti, Thomas
    Bickel, Steven
    Monnin, Pascal
    Higashigaito, Kai
    Martini, Katharina
    Alkadhi, Hatem
    Euler, Andre
    RADIOLOGY, 2022, 305 (01) : 106 - 114
  • [38] Photon-counting detector CT for abdominal imaging: opportunities and challenges
    Fides Regina Schwartz
    Hatem Alkadhi
    European Radiology, 2023, 33 : 7805 - 7806
  • [39] Investigation of the LabPET™ detector and electronics for photon-counting CT imaging
    Berard, Philippe
    Riendeau, Joel
    Pepin, Catherine M.
    Rouleau, Daniel
    Cadorette, Jules
    Fontaine, Rejean
    Lecomte, Roger
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 114 - 117
  • [40] Editorial Comment: Unleashing Precision of Photon-Counting Detector CT
    Hsu, Yin-Chen
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (03)