First Clinical Photon-counting Detector CT System: Technical Evaluation

被引:267
|
作者
Rajendran, Kishore [1 ]
Petersilka, Martin [2 ]
Henning, Andre [2 ]
Shanblatt, Elisabeth R. [3 ]
Schmidt, Bernhard [2 ]
Flohr, Thomas G. [2 ]
Ferrero, Andrea [1 ]
Baffour, Francis [1 ]
Diehn, Felix E. [1 ]
Yu, Lifeng [1 ]
Rajiah, Prabhakar [1 ]
Fletcher, Joel G. [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Siemens Healthineers, Forchheim, Germany
[3] Siemens Med Solut, Malvern, PA USA
基金
美国国家卫生研究院;
关键词
COMPUTED-TOMOGRAPHY; NOISE; ACCURACY;
D O I
10.1148/radiol.212579
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The first clinical CT system to use photon-counting detector (PCD) technology has become available for patient care. Purpose: To assess the technical performance of the PCD CT system with use of phantoms and representative participant examinations. Materials and Methods: Institutional review board approval and written informed consent from four participants were obtained. Technical performance of a dual-source PCD CT system was measured for standard and high-spatial-resolution (HR) collimations. Noise power spectrum, modulation transfer function, section sensitivity profile, iodine CT number accuracy in virtual monoenergetic images (VMIs), and iodine concentration accuracy were measured. Four participants were enrolled (between May 2021 and August 2021) in this prospective study and scanned using similar or lower radiation doses as their respective clinical examinations performed on the same day using energy-integrating detector (EID) CT. Image quality and findings from the participants' PCD CT and EID CT examinations were compared. Results: All standard technical performance measures met accreditation and regulatory requirements. Relative to filtered back-projection reconstructions, images from iterative reconstruction had lower noise magnitude but preserved noise power spectrum shape and peak frequency. Maximum in-plane spatial resolutions of 125 and 208 mu m were measured for HR and standard PCD CT scans, respectively. Minimum values for section sensitivity profile full width at half maximum measurements were 0.34 mm (0.2-mm nominal section thickness) and 0.64 mm (0.4-mm nominal section thickness) for HR and standard PCD CT scans, respectively. In a 120-kV standard PCD CT scan of a 40-cm phantom, VMI iodine CT numbers had a mean percentage error of 5.7%, and iodine concentration had root mean squared error of 0.5 mg/cm(3), similar to previously reported values for EID CT. VMIs, iodine maps, and virtual noncontrast images were created for a coronary CT angiogram acquired with 66-msec temporal resolution. Participant PCD CT images showed up to 47% lower noise and/or improved spatial resolution compared with EID CT. Conclusion: Technical performance of clinical photon-counting detector (PCD) CT is improved relative to that of a current stateof-the-art CT system. The dual-source PCD geometry facilitated 66-msec temporal resolution multienergy cardiac imaging. Study participant images illustrated the effect of the improved technical performance. (C) RSNA, 2022
引用
收藏
页码:130 / 138
页数:9
相关论文
共 50 条
  • [21] An introduction to photon-counting detector CT (PCD CT) for radiologists
    Nakamura, Yuko
    Higaki, Toru
    Kondo, Shota
    Kawashita, Ikuo
    Takahashi, Isao
    Awai, Kazuo
    JAPANESE JOURNAL OF RADIOLOGY, 2023, 41 (03) : 266 - 282
  • [22] Clinical evaluation of a photon-counting tomosynthesis mammography system
    Maidment, Andrew D. A.
    Ullberg, Christer
    Francke, Tom
    Lindqvist, Lars
    Sokolov, Skiff
    Lindman, Karin
    Adelow, Leif
    Sunden, Per
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2006, 4046 : 144 - 151
  • [23] An introduction to photon-counting detector CT (PCD CT) for radiologists
    Yuko Nakamura
    Toru Higaki
    Shota Kondo
    Ikuo Kawashita
    Isao Takahashi
    Kazuo Awai
    Japanese Journal of Radiology, 2023, 41 : 266 - 282
  • [24] Advantages of Photon-Counting Detector CT in Aortic Imaging
    Zanon, Chiara
    Cademartiri, Filippo
    Toniolo, Alessandro
    Bini, Costanza
    Clemente, Alberto
    Colacchio, Elda Chiara
    Cabrelle, Giulio
    Mastro, Florinda
    Antonello, Michele
    Quaia, Emilio
    Pepe, Alessia
    TOMOGRAPHY, 2024, 10 (01) : 1 - 13
  • [25] CT technology: photon-counting detector computed tomography
    Layer, Yannik Christian
    Kravchenko, Dmitrij
    Dell, Tatjana
    Kuetting, Daniel
    RADIOLOGIE, 2023, 63 (07): : 497 - 506
  • [26] MSK - Myeloma Diagnostics with Photon-Counting Detector CT
    Graewert, Stephanie
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2023, 195 (11): : 976 - +
  • [27] AUTOMATIC SPECTROSCOPIC PHOTON-COUNTING DETECTOR SYSTEM
    KUREPA, JM
    PANIC, BM
    LEVI, BJ
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1977, 32 (9-10) : 413 - 417
  • [28] Standardization and Quantitative Imaging With Photon-Counting Detector CT
    McCollough, Cynthia H.
    Rajendran, Kishore
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 451 - 458
  • [29] Spectrum Optimization for Photon-Counting CT with a Silicon Detector
    Schmidt, T.
    Gronberg, F.
    Fan, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [30] Bone Mineral Density Assessment Using Spectral Topograms From a Clinical Photon-Counting Detector CT System: A Phantom Evaluation
    Moser, Lukas J.
    Pitteloud, Janique
    Mergen, Victor
    Frey, Diana
    Nowak, Tristan
    Distler, Oliver
    Eberhard, Matthias
    Alkadhi, Hatem
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (02)