Evaluation of MLC leaf transmission on IMRT treatment plan quality of patients with advanced lung cancer

被引:7
|
作者
Chen, Jiayun
Fu, Guishan
Li, Minghui
Song, Yixin
Dai, Jianrong [1 ,2 ]
Miao, Junjie
Liu, Zhiqiang
Li, Yexiong
机构
[1] Chinese Acad Med Sci, Dept Radiat Oncol, Natl Canc Ctr, Canc Hosp, Beijing 100021, Peoples R China
[2] Peking Union Med Coll, Beijing 100021, Peoples R China
关键词
MLC transmission; IMRT; plan quality; lung sparing; 3-DIMENSIONAL CONFORMAL RADIOTHERAPY; TREATMENT-RELATED PNEUMONITIS; CONCURRENT CHEMOTHERAPY; MULTILEAF COLLIMATOR; RADIATION; VALIDATION;
D O I
10.1016/j.meddos.2017.10.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The purpose of this paper was to evaluate the impact of leaf treatment of multileaf collimator (MLC) in plan quality of intensity-modulated radiotherapy (IMRT) of patients with advanced lung cancer. Five MLCs with different leaf transmissions (0.01%, 0.5%, 1.2%, 1.8%, and 3%) were configured for an accelerator in a treatment planning system. Correspondingly, 5 treatment plans with the same optimization setting were created and evaluated quantitatively for each patient (11 patients total) who was diagnosed with advanced lung cancer. All of the 5 plans for each patient met the dose requirement for the planning treatment volumes (PTVs) and had similar target dose homogeneity and conformity. On average, the doses to selected organs were as follows: (1) V-5, V-20, and the mean dose of total lung; (2) the maximum and mean dose to spinal cord planning organ-at-risk volume (PRV); and (3) V-30 and V-40 of heart, decreased slightly when MLC transmission was decreased, but with no statistical differences. There is a clear grouping of plans having total quality score (S-D) value, which is used to evaluate plan quality: (1) more than 1 (patient nos. 1 to 3, 5, and 8), and more than 2.5 (patient no. 6); (2) less than 1 (patient nos. 7 and 10); (3) around 1 (patient nos. 4, 9, and 11). As MLC transmission increased, overall S-D values increased as well and plan dose requirement was harder to meet. The clinical requirements were violated increasingly as MLC transmission became large. Total S-D with and without normal tissue (NT) showed similar results, with no statistically significant differences. Therefore, decrease of MLC transmission did have minimum impact on plan, and it improved target coverage and reduced normal tissue radiation slightly, with no statistical significance. Plan quality could not be significantly improved by MLC transmission reduction. However, lower MLC transmission may have advantages on lung sparing to low- and intermediate-dose exposure. Besides conventional fraction, hyperfraction, or sfeivotactic body radiotherapy (SBRT), the reduction on lung sparing is still essential because it is highly relevant to radiation pneumonitis (RP). It has potential to diminish incidence of RP and improve patient's quality of life after irradiation with lowered MLC transmission. (C) 2017 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:313 / 318
页数:6
相关论文
共 50 条
  • [1] Impact of MLC leaf size on the quality of IMRT plan
    Cheng, C.
    Das, I.
    MEDICAL PHYSICS, 2006, 33 (06) : 2108 - 2109
  • [2] Improving IMRT-plan quality with MLC leaf position refinement post plan optimization
    Niu, Ying
    Zhang, Guowei
    Berman, Barry L.
    Parke, William C.
    Yi, Byongyong
    Yu, Cedric X.
    MEDICAL PHYSICS, 2012, 39 (08) : 5118 - 5126
  • [3] Improving IMRT-Plan Quality with MLC Leaf Position Refinement Post Plan Optimization
    Niu, Y.
    Zhang, G.
    Berman, B.
    Yu, C.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [4] Planning quality Variations : for IMRT lung cancer based on treatment plan database
    Koo, J. H.
    Yoon, M. G.
    Chung, W. K.
    Kim, D. W.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S838 - S838
  • [5] Plan quality & deliverability of IMRT plans for lung cancer using various leaf motion calculators
    Kong, E.
    Fung, N. T.
    Lee, M. C.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1813 - S1814
  • [6] Evaluation of Treatment Plan Quality for Single Arc IMRT
    Esthappan, J.
    Michaletz-Lorenz, M.
    Rangaraj, D.
    Goddu, S. M.
    Santanam, L.
    Low, D. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S724 - S724
  • [7] Leaf transmission reduction using moving jaws for dynamic MLC IMRT
    Schmidhalter, D.
    Fix, M. K.
    Niederer, P.
    Mini, R.
    Manser, P.
    MEDICAL PHYSICS, 2007, 34 (09) : 3674 - 3687
  • [8] Reduction of leaf transmission by using moving jaws for dynamic MLC IMRT
    Schmidhalter, D.
    Fix, M.
    Manser, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 : 183 - 183
  • [9] Does MLC With Finer Leaf Width Improve the Treatment Plan Quality in Stereotactic Body Radiation Therapy of Lung Cancers
    Ye, J.
    Wu, J.
    Chen, F.
    Hill, G.
    Spiegel, J.
    Mehta, V.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S899 - S899
  • [10] The effect of extremely narrow MLC leaf width on the plan quality of VMAT for prostate cancer
    Park, Jong Min
    Park, So-Yeon
    Kim, Jin Ho
    Carlson, Joel
    Kim, Jung-In
    RADIATION ONCOLOGY, 2016, 11