Reduced susceptibility effects in perfusion fMRI with single-shot spin-echo EPI acquisitions at 1.5 Tesla

被引:62
|
作者
Wang, JJ [1 ]
Li, L
Roc, AC
Alsop, DC
Tang, K
Butler, NS
Schnall, MD
Detre, JA
机构
[1] Univ Penn, Dept Radiol, Metab Magnet Resonance Res & Comp Ctr, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Neurol, Philadelphia, PA 19104 USA
[3] Harvard Univ, Sch Med, Beth Israel Deaconess Med Ctr, Dept Radiol, Boston, MA 02115 USA
关键词
arterial spin label; susceptibility artifacts; spin-echo EPI; perfusion fMRI;
D O I
10.1016/S0730-725X(03)00210-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Arterial spin labeling (ASL) perfusion contrast is not based on susceptibility effects and can therefore be used to study brain function in regions of high static inhornogeneity. As a proof of concept, single-shot spin-echo echo-planar imaging (EPI) acquisition was carried out with a multislice continuous ASL (CASL) method at 1.5T. A bilateral finger tapping paradigm was used in the presence of an exogenously induced susceptibility artifact over left motor cortex. The spin-echo CASL technique was compared with a regular gradient-echo EPI sequence with the same slice thickness, as well as other imaging methods using thin slices and spin-echo acquisitions. The results demonstrate improved functional sensitivity and efficiency of the spin-echo CASL approach as compared with gradient-echo EPI techniques, and a trend of improved sensitivity as compared with spin-echo EPI approach in the brain regions affected by the susceptibility artifact. ASL images, either with or without subtraction of the control, provide a robust alternative to blood oxygenation level dependant (BOLD) methods for activation imaging in regions of high static field inhomogeneity. (C) 2004 Elsevier Inc. All rights reserved.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 50 条
  • [41] T2-weighted MRI of rectosigmoid carcinoma: Comparison of respiratory-triggered fast spin-echo, breathhold fast-recovery fast spin-echo, and breathhold single-shot fast spin-echo sequences
    Yamashita, Shuhei
    Masui, Takayuki
    Katayama, Motoyuki
    Sato, Kimihiko
    Yoshizawa, Nobuko
    Seo, Hidekazu
    Sakahara, Harumi
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (03) : 511 - 516
  • [42] MR cholangiopancreatography using HASTE (half-Fourier acquisition single-shot turbo spin-echo) sequences
    Miyazaki, T
    Yamashita, Y
    Tsuchigame, T
    Yamamoto, H
    Urata, J
    Takahashi, M
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (06) : 1297 - 1303
  • [43] MR sialography using half-Fourier acquisition single-shot turbo spin-echo (HASTE) sequences
    Murakami, R
    Baba, Y
    Nishimura, R
    Baba, T
    Matsumoto, N
    Yamashita, Y
    Ishikawa, T
    Takahashi, M
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1998, 19 (05) : 959 - 961
  • [44] MR imaging of pulmonary parenchyma with a half-Fourier single-shot turbo spin-echo (HASTE) sequence
    Hatabu, H
    Gaa, J
    Tadamura, E
    Edinburgh, KJ
    Stock, KW
    Garpestad, E
    Edelman, RR
    EUROPEAN JOURNAL OF RADIOLOGY, 1999, 29 (02) : 152 - 159
  • [45] Volumetric assessment of normal and hypoplastic fetal lungs by antenatal single-shot fast spin-echo MR imaging
    Coakley, FV
    Lopoo, J
    Hricak, H
    Albanese, CT
    Harrison, MR
    Filly, RA
    RADIOLOGY, 1999, 213P : 446 - 446
  • [46] Comparison of ultrafast half-Fourier single-shot turbo spin-echo sequence with turbo spin-echo sequences for T2-weighted imaging of the female pelvis
    Yamashita, Y
    Tang, Y
    Abe, Y
    Mitsuzaki, K
    Takahashi, M
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (06): : 1207 - 1212
  • [47] Single shot partial dual echo (SPADE) EPI-an efficient acquisition scheme for reducing susceptibility artefacts in fMRI
    Schwarzbauer, Christian
    Porter, David A.
    NEUROIMAGE, 2010, 49 (03) : 2234 - 2237
  • [48] Diffusion-Weighted MR Imaging in Head and Neck Cancer: Comparison between Half-Fourier Acquired Single-Shot Turbo Spin-Echo and EPI Techniques
    Verhappen, M. H.
    Pouwels, P. J. W.
    Ljumanovic, R.
    van der Putten, L.
    Knol, D. L.
    De Bree, R.
    Castelijns, J. A.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (07) : 1239 - 1246
  • [49] The value of single-shot turbo spin-echo diffusion-weighted MR imaging in the detection of middle ear cholesteatoma
    Bert De Foer
    Jean-Philippe Vercruysse
    Anja Bernaerts
    Joachim Maes
    Filip Deckers
    Johan Michiels
    Thomas Somers
    Marc Pouillon
    Erwin Offeciers
    Jan W. Casselman
    Neuroradiology, 2007, 49 : 841 - 848
  • [50] Half Fourier Single-shot Turbo Spin-echo Magnetic Resonance Urography for the Evaluation of Suspected Renal Colic in Pregnancy
    Mullins, Jeffrey K.
    Semins, Michelle J.
    Hyams, Elias S.
    Bohlman, Mark E.
    Matlaga, Brian R.
    UROLOGY, 2012, 79 (06) : 1252 - 1255