Variability of three activated clotting time point-of-care systems in cardiac surgery: reinforcing available evidence

被引:1
|
作者
Solis Clavijo, Diego [1 ]
Ortega Cotano, Abel [1 ]
Alonso Pena, Nuria [2 ]
Caballero Galvez, Sergio [1 ]
Arellano Nunez, Fernando [1 ]
Carballo Caro, Juan Manuel [1 ]
Munoz-Cacho, Pedro [3 ]
Tocon Ale, Cristina [1 ]
机构
[1] Virgen Rocio Univ Hosp, Ave Marie Curie 3,Portal 6 3 B, Seville 41927, Spain
[2] Marques Valdecilla Univ Hosp, Cardiovasc Surg Area, Santander, Cantabria, Spain
[3] Cantabrian Hlth Serv, Cantabria, Spain
来源
PERFUSION-UK | 2022年 / 37卷 / 07期
关键词
activated clotting time; anticoagulant drugs; blood coagulation test; point-of-care systems;
D O I
10.1177/02676591211023687
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Cardiac surgery with extracorporeal circulation (ECC) requires the administration of anticoagulant drugs to maintain ACT ranges 400-600 seconds, which requires exhaustive coagulation monitoring for which various point-of-care devices are available. However, there is variability between them, so we aimed to compare the values in ACT measurement. Methods: Simultaneous ACT measurements were performed with the Hemochron Response(R), Hemostasis Management System Plus(R) (HMS Plus(R)) and Hemochron Signature(R) systems. Results: A total of 255 simultaneous measurements were taken, the mean and standard deviation (SD) of each device were: Hemochron Signature(R) 361.1 seconds (SD: 156.9), HMS Plus(R) 412.8 seconds (SD: 180.9) and Hemochron Response(R) 422.8 seconds (SD: 187.9), being these differences statistically significant (Fridman's test p < 0.01). For comparisons the Bland-Altman method was used, resulting the Hemochron Response(R) has 61.7 seconds higher mean values than the Hemochron Signature(R), the Hemochron Response(R) 10 seconds higher than the HMS Plus(R) and the HMS Plus(R) 51.7 seconds higher than the Hemochron Signature(R). Conclusion: The differences found in comparisons are considered to be clinically relevant, which is why it is considered important to make the variability of the different monitoring systems known and to take them into account for optimal control of this parameter and its clinical repercussions.
引用
收藏
页码:711 / 714
页数:4
相关论文
共 50 条
  • [41] Assessment of a commercially available point-of-care assay for the measurement of bovine cardiac troponin I concentration
    Fraser, Brandon C.
    Anderson, David E.
    White, Brad J.
    Miesner, Matt D.
    Wheeler, Carrie
    Amrine, David
    Lakritz, Jeff
    Overbay, Tom
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2013, 74 (06) : 870 - 873
  • [42] Evaluation of the Ultegra®, point-of-care platelet function instrument during cardiac surgery
    Shore-Lesserson, L
    Saleem, R
    Stone, M
    Hillman, R
    Despotis, G
    ANESTHESIOLOGY, 2000, 93 (3A) : U146 - U146
  • [43] Retraction Note: Point-of-care (POC) measurement of coagulation after cardiac surgery
    J. Boldt
    G. Walz
    J. Triem
    S. Suttner
    B. Kumle
    Intensive Care Medicine, 2023, 49 : 489 - 489
  • [44] Level of agreement between laboratory and point-of-care prothrombin time before and after cardiopulmonary bypass in cardiac surgery
    Meesters, Michael I.
    Vonk, Alexander B. A.
    van de Weerdt, Emma K.
    Kamminga, Suzanne
    Boer, Christa
    THROMBOSIS RESEARCH, 2014, 133 (06) : 1141 - 1144
  • [45] RETRACTED ARTICLE: Point-of-care (POC) measurement of coagulation after cardiac surgery
    J. Boldt
    G. Walz
    J. Triem
    S. Suttner
    B. Kumle
    Intensive Care Medicine, 1998, 24 : 1187 - 1193
  • [46] Does tranexamic acid influence activated clotting time values in patients undergoing cardiac surgery?
    Bartlett, Elise
    Urman, Richard D.
    Urits, Ivan
    Kaye, Alan D.
    Viswanath, Omar
    JOURNAL OF CLINICAL ANESTHESIA, 2022, 79
  • [47] A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass
    Li, Han
    Serrick, Cyril
    Rao, Vivek
    Yip, Paul M.
    PERFUSION-UK, 2021, 36 (06): : 610 - 619
  • [48] Platelet-activated clotting time does not measure platelet reactivity during cardiac surgery
    Shore-Lesserson, L
    Ammar, T
    DePerio, M
    Vela-Cantos, F
    Fisher, C
    Sarier, K
    ANESTHESIOLOGY, 1999, 91 (02) : 362 - 368
  • [49] A Novel Point-of-Care Smartphone Based System for Monitoring the Cardiac and Respiratory Systems
    Kwanghyun Sohn
    Faisal M. Merchant
    Omid Sayadi
    Dheeraj Puppala
    Rajiv Doddamani
    Ashish Sahani
    Jagmeet P. Singh
    E. Kevin Heist
    Eric M. Isselbacher
    Antonis A. Armoundas
    Scientific Reports, 7
  • [50] A Novel Point-of-Care Smartphone Based System for Monitoring the Cardiac and Respiratory Systems
    Sohn, Kwanghyun
    Merchant, Faisal M.
    Sayadi, Omid
    Puppala, Dheeraj
    Doddamani, Rajiv
    Sahani, Ashish
    Singh, Jagmeet P.
    Heist, E. Kevin
    Isselbacher, Eric M.
    Armoundas, Antonis A.
    SCIENTIFIC REPORTS, 2017, 7