4-D Intracardiac Ultrasound Vector Flow Imaging-Feasibility and Comparison to Phase-Contrast MRI

被引:55
|
作者
Wigen, Morten Smedsrud [1 ]
Fadnes, Solveig [1 ]
Rodriguez-Molares, Alfonso [1 ]
Bjastad, Tore [1 ,2 ]
Eriksen, Marius [3 ]
Stensaeth, Knut Haakon [1 ,3 ]
Stoylen, Asbjorn [1 ]
Lovstakken, Lasse [1 ]
机构
[1] NTNU, Dept Circulat & Med Imaging, N-7491 Trondheim, Norway
[2] GE Vingmed, N-3183 Horten, Norway
[3] St Olavs Hosp, Dept Radiol & Nucl Med, N-7006 Trondheim, Norway
关键词
Blood flow imaging; intracardiac flow imaging; plane wave imaging; vector flow imaging; 4D ultrasound; VELOCITY ESTIMATION; MOTION ESTIMATION; TISSUE DOPPLER; B-MODE;
D O I
10.1109/TMI.2018.2844552
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In vivo characterization of intracardiac blood velocity vector fields may provide new clinical information but is currently not available for bedside evaluation. In this paper, 4-D vector flow imaging for intracardiac flow assessment is demonstrated using a clinical ultrasound (US) system and a matrix array transducer, without the use of contrast agent. Two acquisition schemes were developed, one for full volumetric coverage of the left ventricle (LA) at 50 vps and a 3-D thick-slice setup with continuous frame acquisition (4000 vps), both utilizing ECG-gating. The 3-D vector velocity estimates were obtained using a novel method combining phase and envelope information. In vitro validation in a rotating tissue-mimicking phantom revealed velocity estimates in compliance with the ground truth, with a linear regression slope of 0.80, 0.77, and 1.03 for the x, y, and z velocity components, and with standard deviations of 2.53, 3.19, and 0.95 cm/s, respectively. In vivo measurements in a healthy LV showed good agreement with PC-MRI. Quantitative analysis of energy loss (EL) and kinetic energy (KE) further showed similar trends, with peak KE at 1.5 and 2.4 mJ during systole and 3.6 and 3.1 mJ for diastole for US and PC-MRI. Similar for EL, 0.15-0.2 and 0.7 mW was found during systole and 0.6 and 0.7 mW during diastole, for US and PC-MRI, respectively. Overall, a potential for US as a future modality for 4D cardiac vector flow imaging was demonstrated, which will be further evaluated in clinical studies.
引用
收藏
页码:2619 / 2629
页数:11
相关论文
共 50 条
  • [1] Applications of phase-contrast flow and velocity imaging in cardiovascular MRI
    Peter D. Gatehouse
    Jennifer Keegan
    Lindsey A. Crowe
    Sharmeen Masood
    Raad H. Mohiaddin
    Karl-Friedrich Kreitner
    David N. Firmin
    European Radiology, 2005, 15 : 2172 - 2184
  • [2] Applications of phase-contrast flow and velocity imaging in cardiovascular MRI
    Gatehouse, PD
    Keegan, J
    Crowe, LA
    Masood, S
    Mohiaddin, RH
    Kreitner, KF
    Firmin, DN
    EUROPEAN RADIOLOGY, 2005, 15 (10) : 2172 - 2184
  • [3] Flow Imaging: Cardiac Applications of 3D Cine Phase-Contrast MRI
    Ebbers T.
    Current Cardiovascular Imaging Reports, 2011, 4 (2) : 127 - 133
  • [4] 4-D flow magnetic resonance imaging: blood flow quantification compared to 2-D phase-contrast magnetic resonance imaging and Doppler echocardiography
    Gabbour, Maya
    Schnell, Susanne
    Jarvis, Kelly
    Robinson, Joshua D.
    Markl, Michael
    Rigsby, Cynthia K.
    PEDIATRIC RADIOLOGY, 2015, 45 (06) : 804 - 813
  • [5] 4-D flow magnetic resonance imaging: blood flow quantification compared to 2-D phase-contrast magnetic resonance imaging and Doppler echocardiography
    Maya Gabbour
    Susanne Schnell
    Kelly Jarvis
    Joshua D. Robinson
    Michael Markl
    Cynthia K. Rigsby
    Pediatric Radiology, 2015, 45 : 804 - 813
  • [6] Intracardiac Vortex Dynamics by High-Frame-Rate Doppler Vortography-In Vivo Comparison With Vector Flow Mapping and 4-D Flow MRI
    Faurie, Julia
    Baudet, Mathilde
    Assi, Kondo Claude
    Auger, Dominique
    Gilbert, Guillaume
    Tournoux, Francois
    Garcia, Damien
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (02) : 424 - 432
  • [7] 4D REAL-TIME PHASE-CONTRAST FLOW MRI WITH SPARSE SAMPLING
    Sun, Aiqi
    Zhao, Bo
    Li, Rui
    Yuan, Chun
    2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2017, : 3252 - 3255
  • [8] Intracardiac Flow at 4D CT: Comparison with 4D Flow MRI
    Lantz, Jonas
    Gupta, Vikas
    Henriksson, Lilian
    Karlsson, Matts
    Persson, Anders
    Carlhall, Carl-Johan
    Ebbers, Tino
    RADIOLOGY, 2018, 289 (01) : 51 - 58
  • [9] The Feasibility of Using Volumetric Phase-Contrast MR Imaging (4D Flow) to Assess for Transjugular Intrahepatic Portosystemic Shunt Dysfunction
    Owen, Joseph W.
    Saad, Nael E.
    Foster, Glenn
    Fowler, Kathryn J.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 29 (12) : 1717 - 1724
  • [10] Radiomics-based aortic flow profile characterization with 4D phase-contrast MRI
    Huellebrand, Markus
    Jarmatz, Lina
    Manini, Chiara
    Laube, Ann
    Ivantsits, Matthias
    Schulz-Menger, Jeanette
    Nordmeyer, Sarah
    Harloff, Andreas
    Hansmann, Jochen
    Kelle, Sebastian
    Hennemuth, Anja
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10