Biomechanical analysis of plate osteosynthesis systems for proximal humerus fractures

被引:31
|
作者
Lever, James P. [2 ]
Aksenov, Sergei A. [2 ]
Zdero, Rad [1 ]
Ahn, Henry [2 ]
Mckee, Michael D. [2 ]
Schemitsch, Emil H. [1 ,2 ]
机构
[1] St Michaels Hosp, Martin Orthopaed Biomech Lab, Toronto, ON M5B 1W8, Canada
[2] Univ Toronto, Dept Surg, Toronto, ON, Canada
关键词
biomechanics; proximal humerus; fracture; plate fixation;
D O I
10.1097/BOT.0b013e31815c89ce
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To biomechanically assess five proximal, humeral, fracture-plate-fixation systems. Methods: Surgical neck fractures, with and without cortical contact, were created in 25 fresh-frozen cadaveric humeri. Five methods of plate fixation were used for repair: construct A [an eight-hole, low contact dynamic compression (LCDC) plate contoured into a blade shape, supported by one, 70-mm-long, 4.5-mm-diameter cortical screw acting as a truss], construct B (a 10-hole LCDC plate arrangement identical to construct A, but using one, 70-mm-long, 3.5-mm-diameter cortical screw as a truss), construct C [a five-hole dynamic compression (DC) blade plate with one, 6.5-mm-diameter cancellous screw], construct D (a five-hole T-plate supported by three, 6.5-mm-diameter cancellous screws), and construct E (a five-hole cloverleaf plate supported by five, 4-mm-diameter cancellous screws). Plates were posterior to the bicipital groove, 10 mm distal to the greater tuberosity tip, on the lateral aspect of the humeral shaft. Screw fixation was done using standard AO compression plating techniques. Stiffness of constructs was measured in bending and axial compression. Locked plates were not assessed. Results: For cortical contact [abduction of 20 degrees (P = 0.02), flexion of 20 degrees (P = 0.02), flexion of 90 degrees (P = 0.005)] and no cortical contact [flexion of 90 degrees (P = 0.0001)], construct A was significantly stiffer than other constructs. For no cortical contact in abduction of 90 degrees (P = 0.05), construct A was significantly stiffer than other constructs. Conclusions: Construct A was significantly stiffer than other constructs.
引用
收藏
页码:23 / 29
页数:7
相关论文
共 50 条
  • [31] Parallel versus orthogonal plate osteosynthesis of adult distal humerus fractures: a meta-analysis of biomechanical studies
    Shih, Chien-An
    Su, Wei-Ren
    Lin, Wei-Chin
    Tai, Tai-Wei
    INTERNATIONAL ORTHOPAEDICS, 2019, 43 (02) : 449 - 460
  • [32] Minimal invasive plate osteosynthesis (MIPO) with locking compression plate (LCP) in proximal diaphyseal humerus fractures
    Sommer, C
    Jenny, R
    Bircher, HP
    Wullschleger, M
    BRITISH JOURNAL OF SURGERY, 2005, 92 (07) : 908 - 908
  • [33] Locked plate fixation and intramedullary nailing for proximal humerus fractures: A biomechanical evaluation
    Hessmann, MH
    Hansen, WSM
    Krummenauer, F
    Pol, TF
    Rommens, M
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2005, 58 (06): : 1194 - 1201
  • [34] Biomechanical comparative study of three systems for osteosynthesis of proximal humoral fractures
    Gicquel, P
    Bonnomet, F
    Boutemy, P
    Schlemmer, B
    Kempf, JF
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1999, 85 (08): : 811 - 820
  • [35] Nagelosteosynthese proximaler HumerusfrakturenNail osteosynthesis of proximal humerus fractures
    J.C. Katthagen
    C. Voigt
    Gunnar Jensen
    Helmut Lill
    Obere Extremität, 2012, 7 (3) : 128 - 136
  • [36] Quality of Life After Osteosynthesis of Fractures of the Proximal Humerus
    Inauen, Cyril
    Platz, Andreas
    Meier, Christoph
    Zingg, Urs
    Rufibach, Kaspar
    Spross, Christian
    Dietrich, Michael
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (04) : E74 - E80
  • [37] Retrograde nailing versus locking plate osteosynthesis of proximal humeral fractures: a biomechanical study
    Dietz, Sven-Oliver
    Hartmann, Frank
    Schwarz, Thomas
    Nowak, Tobias E.
    Enders, Annalisa
    Kuhn, Sebastian
    Hofmann, Alexander
    Rommens, Pol Maria
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2012, 21 (05) : 618 - 624
  • [38] A biomechanical comparison of locking plate and locking nail implants used for fractures of the proximal humerus
    Kitson, Jeffrey
    Booth, Grant
    Day, Robert
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (03) : 362 - 366
  • [39] Minimally invasive plate osteosynthesis for humerus diaphyseal fractures
    Shetty, M. Shantharam
    Kumar, M. Ajith
    Sujay, K. T.
    Kini, Abhishek R.
    Kiran, Kanthi G.
    INDIAN JOURNAL OF ORTHOPAEDICS, 2011, 45 (06) : 520 - 526
  • [40] Minimally invasive plate osteosynthesis for humerus diaphyseal fractures
    M. Shantharam Shetty
    M. Ajith Kumar
    K. T. Sujay
    Abhishek R. Kini
    G. Kiran Kanthi
    Indian Journal of Orthopaedics, 2011, 45 : 520 - 526