Dynamic Calibration of the Cutting Temperature Sensor of NiCr/NiSi Thin-film Thermocouple

被引:10
|
作者
Cui Yunxian [1 ,2 ]
Yang Deshun [1 ]
Jia Ying [1 ]
Zeng Qiyong [3 ]
Sun Baoyuan [2 ]
机构
[1] Dalian Jiaotong Univ, Sch Mech Engn, Dalian 116028, Peoples R China
[2] Dalian Univ Technol, Minist Educ, Key Lab Precis & Nontradit Machining Technol, Dalian 116024, Peoples R China
[3] China Jiliang Univ, Coll Qual & Safety Engn, Hangzhou 310018, Peoples R China
基金
中国国家自然科学基金;
关键词
thin-film thermocouple; cutting temperature sensor; dynamic calibration; one-dimensional unsteady heat conduction; response time;
D O I
10.3901/CJME.2011.01.073
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
In high-speed cutting, natural thermocouple, artificial thermocouple and infrared radiation temperature measurement are usually adopted for measuring cutting temperature, but these methods have difficulty in measuring transient temperature accurately of cutting area on account of low response speed and limited cutting condition. In this paper, NiCr/NiSi thin-film thermocouples(TFTCs) are fabricated according to temperature characteristic of cutting area in high-speed cutting by means of advanced twinned microwave electro cyclotron resonance(MW-ECR) plasma source enhanced radio frequency(RF) reaction non-balance magnetron sputtering technique, and can be used for transient cutting temperature measurement. The time constants of the TFTCs with different thermo-junction film width are measured at four kinds of sampling frequency by using Ultra-CFR short pulsed laser system that established. One-dimensional unsteady heat conduction model is constnicted and the dynamic performance is analyzed theoretically. It can be seen from the analysis results that the NiCr/NiSi TFTCs are suitable for measuring transient temperature which varies quickly, the response speed of TFTCs can be obviously improved by reducing the thickness of thin-film, and the area of thermo-junction has little influence on dynamic response time. The dynamic calibration experiments are made on the constructed dynamic calibration system, and the experimental results confirm that sampling frequency should be larger than 50 kHz in dynamic measurement for stable response time, and the shortest response time is 0.042 ms. Measurement methods and devices of cutting heat and cutting temperature measurement are developed and improved by this research, which provide practical methods and instruments in monitoring cutting heat and cutting temperature for research and production in high-speed machining.
引用
收藏
页码:73 / 77
页数:5
相关论文
共 50 条
  • [21] THIN-FILM TEMPERATURE SENSOR
    MCLENNAN, WD
    OOTEN, RT
    BYNUM, DS
    PALMER, JL
    PENDLEY, KN
    BAXTER, CR
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1975, 12 (01): : 71 - 73
  • [22] Development of Multilayer Composition Thin Film Thermocouple Cutting Temperature Sensor Based on Magnetron Sputtering
    Cui, Y. X.
    Sun, B. Y.
    Ding, W. Y.
    Sun, F. D.
    ULTRA-PRECISION MACHINING TECHNOLOGIES, 2009, 69-70 : 515 - +
  • [23] The study of dynamic performance compensation of thin-film thermocouple for transient high temperature measurement
    Lei, M
    Wang, ZZ
    Ma, QD
    ISTM/99: 3RD INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, 1999, : 1014 - 1017
  • [24] TRANSPARENT THIN-FILM THERMOCOUPLE
    YUST, M
    KREIDER, KG
    THIN SOLID FILMS, 1989, 176 (01) : 73 - 78
  • [25] THIN-FILM THERMOCOUPLE GAUGE
    KUO, TC
    FLATTERY, J
    GHOSH, PK
    KORNREICH, PG
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1988, 6 (03): : 1150 - 1152
  • [26] THIN-FILM NICR RESISTOR
    BUCZEK, DM
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1978, 15 (02): : 370 - 372
  • [27] Fabrication and Characterization of NiCr/NiSi Functional Thin Films on Temperature Measurement of Cutter Sensor
    Cui, Yunxian
    Yang, Deshun
    Zeng, Qiyong
    Sun, Baoyuan
    MACHINING AND ADVANCED MANUFACTURING TECHNOLOGY X, 2010, 431-432 : 535 - 538
  • [28] Preliminary investigation of heat conduction time in SiO2 layer by bilayer NiSi/NiCr thin film thermocouple
    Bian, Yong
    Hao, Yuji
    Sun, Yi
    Liu, Jiabei
    Gao, Zhaoran
    Zhang, Chaoying
    Yin, Junwei
    Wang, Hualin
    Cui, Yunxian
    Liu, Zhihui
    Ding, Wanyu
    MATERIALS LETTERS, 2025, 390
  • [29] A STUDY OF THE TEMPERATURE-DEPENDENCE OF NICR THIN-FILM RESISTORS
    PELED, A
    ZLOOF, Y
    FARHADYAN, J
    PELED, AM
    SOLID-STATE ELECTRONICS, 1991, 34 (06) : 667 - 669
  • [30] Fabrication and Calibration of Pt-Rh10/Pt Thin-Film Thermocouple
    Wang, Fengxiang
    Lin, Zhenyu
    Zhang, Zhijie
    Li, Yanfeng
    Chen, Haoze
    Liu, Jiaqi
    Li, Chao
    MICROMACHINES, 2023, 14 (01)