Sparsity-enforced slice-selective MRI RF excitation pulse design

被引:49
|
作者
Zelinski, Adam C. [1 ]
Wald, Lawrence L. [2 ,3 ]
Setsompop, Kawin [1 ]
Goyal, Vivek K. [1 ]
Adalsteinsson, Elfar [1 ,4 ]
机构
[1] MIT, Dept Elect Engn & Comp Sci, Elect Res Lab, Cambridge, MA 02139 USA
[2] Harvard Univ, Sch Med, Dept Radiol, Boston, MA 02115 USA
[3] Harvard Univ, Sch Med, HST, Boston, MA 02115 USA
[4] Harvard MIT, Div Hlth Sci & Technol, Cambridge, MA 02139 USA
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
B-1 inhomogeneity mitigation; high field strength; magnetic resonance imaging (MRI) radio-frequency (RF) pulse sequence design; parallel transmission; sparse approximation; three-dimensional (3-D) RF excitation;
D O I
10.1109/TMI.2008.920605
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We introduce a novel algorithm for the design of fast slice-selective spatially-tailored magnetic resonance imaging (MRI) excitation pulses. This method, based on sparse approximation theory, uses a second-order cone optimization to place and modulate a small number of slice-selective sinc-like radio-frequency (RF) pulse segments ("spokes") in excitation k-space, enforcing sparsity on the number of spokes allowed while simultaneously encouraging those that remain to be placed and modulated in a way that best forms a user-defined in-plane target magnetization. Pulses are designed to mitigate B-1 inhomogeneity in a water phantom at 7 T and to produce highly-structured excitations in an oil phantom on an eight-channel parallel excitation system at 3 T. In each experiment, pulses generated by the sparsity-enforced method outperform those created via conventional Fourier-based techniques, e.g., when attempting to produce a uniform magnetization in the presence of severe B., inhomogeneity, a 5.7-ms 15-spoke pulse generated by the sparsity-enforced method produces an excitation with 1.28 times lower root mean square error than conventionally-designed 15-spoke pulses. To achieve this same level of uniformity, the conventional methods need to use 29-spoke pulses that are 7.8 ms long.
引用
收藏
页码:1213 / 1229
页数:17
相关论文
共 50 条
  • [1] Sparsity in MRI RF excitation pulse design
    Zelinski, Adam C.
    Goyal, Vivek K.
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    2008 42ND ANNUAL CONFERENCE ON INFORMATION SCIENCES AND SYSTEMS, VOLS 1-3, 2008, : 252 - +
  • [2] A CHEMICAL-SHIFT-INSENSITIVE SLICE-SELECTIVE RF PULSE (CHISS)
    WILLIAMS, DS
    LOWE, IJ
    JOURNAL OF MAGNETIC RESONANCE, 1991, 91 (01): : 57 - 64
  • [3] Improved Slice-Selective Adiabatic Excitation
    Balchandani, Priti
    Glover, Gary
    Pauly, John
    Spielman, Daniel
    MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (01) : 75 - 82
  • [4] Velocity sensitivity of slice-selective excitation
    Lewis, DP
    Tsui, BMW
    Moran, PR
    MAGNETIC RESONANCE IMAGING, 1998, 16 (08) : 907 - 916
  • [5] Adiabatic Slice-Selective Excitation for Surface Coils
    Magnetic Resonance Center, Dept. Molec. Biophys. and Biochem., Yale University School of Medicine, New Haven, CT 06510, United States
    不详
    J. Magn. Reson., 1 (72-79):
  • [6] Adiabatic slice-selective excitation for surface coils
    Shen, J
    Rothman, DL
    JOURNAL OF MAGNETIC RESONANCE, 1997, 124 (01) : 72 - 79
  • [7] SELECTIVE PROPERTIES OF TRAINS OF REAL AND COMPLEX RF MICROPULSES IN SLICE-SELECTIVE SPIN-ECHO MRI
    JUNG, B
    ERICSSON, A
    HEMMINGSSON, A
    UPSALA JOURNAL OF MEDICAL SCIENCES, 1986, 91 (01) : 37 - 43
  • [8] Composite slice-selective adiabatic excitation for prostate MRSI
    de Castro, C. S. Arteaga
    Luttje, M. P.
    van Vulpen, M.
    Luijten, P. R.
    van der Heide, U. A.
    Klomp, D. W. J.
    NMR IN BIOMEDICINE, 2013, 26 (04) : 436 - 442
  • [9] ADIABATIC SLICE-SELECTIVE RF PULSES AND A SINGLE-SHOT ADIABATIC LOCALIZATION PULSE SEQUENCE
    SLOTBOOM, J
    BOVEE, WMMJ
    CONCEPTS IN MAGNETIC RESONANCE, 1995, 7 (03): : 193 - 217
  • [10] Slice-selective tunable-flip adiabatic low peak-power excitation pulse
    Balchandani, Priti
    Pauly, John
    Spielman, Daniel
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) : 1072 - 1078