Sparsity-enforced slice-selective MRI RF excitation pulse design

被引:49
|
作者
Zelinski, Adam C. [1 ]
Wald, Lawrence L. [2 ,3 ]
Setsompop, Kawin [1 ]
Goyal, Vivek K. [1 ]
Adalsteinsson, Elfar [1 ,4 ]
机构
[1] MIT, Dept Elect Engn & Comp Sci, Elect Res Lab, Cambridge, MA 02139 USA
[2] Harvard Univ, Sch Med, Dept Radiol, Boston, MA 02115 USA
[3] Harvard Univ, Sch Med, HST, Boston, MA 02115 USA
[4] Harvard MIT, Div Hlth Sci & Technol, Cambridge, MA 02139 USA
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
B-1 inhomogeneity mitigation; high field strength; magnetic resonance imaging (MRI) radio-frequency (RF) pulse sequence design; parallel transmission; sparse approximation; three-dimensional (3-D) RF excitation;
D O I
10.1109/TMI.2008.920605
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We introduce a novel algorithm for the design of fast slice-selective spatially-tailored magnetic resonance imaging (MRI) excitation pulses. This method, based on sparse approximation theory, uses a second-order cone optimization to place and modulate a small number of slice-selective sinc-like radio-frequency (RF) pulse segments ("spokes") in excitation k-space, enforcing sparsity on the number of spokes allowed while simultaneously encouraging those that remain to be placed and modulated in a way that best forms a user-defined in-plane target magnetization. Pulses are designed to mitigate B-1 inhomogeneity in a water phantom at 7 T and to produce highly-structured excitations in an oil phantom on an eight-channel parallel excitation system at 3 T. In each experiment, pulses generated by the sparsity-enforced method outperform those created via conventional Fourier-based techniques, e.g., when attempting to produce a uniform magnetization in the presence of severe B., inhomogeneity, a 5.7-ms 15-spoke pulse generated by the sparsity-enforced method produces an excitation with 1.28 times lower root mean square error than conventionally-designed 15-spoke pulses. To achieve this same level of uniformity, the conventional methods need to use 29-spoke pulses that are 7.8 ms long.
引用
收藏
页码:1213 / 1229
页数:17
相关论文
共 50 条
  • [31] ON THE USE OF A SLICE-SELECTIVE 270-DEGREES SELF-REFOCUSING GAUSSIAN PULSE FOR MAGNETIC-RESONANCE-IMAGING - COMMENTS
    EMSLEY, L
    BODENHAUSEN, G
    MAGNETIC RESONANCE IN MEDICINE, 1991, 19 (02) : 461 - 463
  • [32] Wedge-Shaped Slice-Selective Adiabatic Inversion Pulse for Controlling Temporal Width of Bolus in Pulsed Arterial Spin Labeling
    Guo, Jia
    Buxton, Richard B.
    Wong, Eric C.
    MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (03) : 838 - 847
  • [33] Flip-angle profile of slice-selective excitation and the measurement of the MR longitudinal relaxation time with steady-state magnetization
    Hsu, Jung-Jiin
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (15): : 5785 - 5801
  • [34] RF pulse design for MRI of samples including conductive implants
    Woo, Taeseong
    Kim, Dongmin
    Sekitani, Tsuyoshi
    Someya, Takao
    Sekino, Masaki
    2014 XXXITH URSI GENERAL ASSEMBLY AND SCIENTIFIC SYMPOSIUM (URSI GASS), 2014,
  • [35] Pulsed selective excitation theory and design in multiphoton MRI
    Han, Victor
    Chi, Jianshu
    Ipek, Tanya Deniz
    Chen, Jingjia
    Liu, Chunlei
    JOURNAL OF MAGNETIC RESONANCE, 2023, 348
  • [36] Joint Design of Excitation k-Space Trajectory and RF Pulse for Small-Tip 3D Tailored Excitation in MRI
    Hao, Sun
    Fessler, Jeffrey A.
    Noll, Douglas C.
    Nielsen, Jon-Fredrik
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (02) : 468 - 479
  • [37] A Novel Fast Algorithm for Parallel Excitation Pulse Design in MRI
    Feng, Shuo
    Ji, Jim
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 1102 - 1105
  • [38] Factors influencing flip angle mapping in MRI:: RF pulse shape, slice-select gradients, off-resonance excitation, and Bo inhomogeneities
    Wang, Jinghua
    Mao, Weihua
    Qiu, Maolin
    Smith, Michael B.
    Constable, R. Todd
    MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (02) : 463 - 468
  • [39] Homogeneous non-selective and slice-selective parallel-transmit excitations at 7 Tesla with universal pulses: A validation study on two commercial RF coils
    Gras, Vincent
    Boland, Markus
    Vignaud, Alexandre
    Ferrand, Guillaume
    Amadon, Alexis
    Mauconduit, Franck
    Le Bihan, Denis
    Stoecker, Tony
    Boulant, Nicolas
    PLOS ONE, 2017, 12 (08):
  • [40] Fast slice-selective radio-frequency excitation pulses for mitigating B1+ inhomogeneity in the human brain at 7 tesla
    Zelinski, Adam C.
    Wald, Lawrence L.
    Setsompop, Kawin
    Alagappan, Vijayanand
    Gagoski, Borjan A.
    Goyal, Vivek K.
    Adalsteinsson, Elfar
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (06) : 1355 - 1364