Electroanalytical Determination of Morpholine as a Corrosion Inhibitor at a Cathodically Pretreated Boron-Doped Diamond Electrode

被引:11
|
作者
Rocha, Luana Rianne [1 ]
D'Elia, Eliane [2 ]
Medeiros, Roberta Antigo [1 ]
Teixeira Tarley, Cesar Ricardo [1 ,3 ]
机构
[1] Univ Estadual Londrina, Ctr Ciencias Exatas, Dept Quim, Londrina, PR, Brazil
[2] Univ Fed Rio de Janeiro, Inst Quim, Rio De Janeiro, Brazil
[3] Univ Estadual Campinas UNICAMP, Inst Nacl Ciencia & Tecnol INCT Bioanalit, Inst Quim, Dept Quim Analit, Campinas, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
Boron-doped diamond electrode; cyclic voltammetry; differential pulse voltammetry; morpholine; square-wave voltammetry; CHROMATOGRAPHIC DETERMINATION; VOLTAMMETRIC DETERMINATION; MILD-STEEL; WATER; AIR; DERIVATIVES; PRODUCTS;
D O I
10.1080/00032719.2018.1515953
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An electroanalytical method for the determination of morpholine, a corrosion inhibitor, was developed at a cathodically pretreated boron-doped diamond electrode (BDDE). The voltammetric response of morpholine at the BDDE in 0.1 mol L-1 KCl (pH 10) shows an irreversible oxidation process at approximately 1.3 V vs. Ag/AgCl in 3.0 mol L-1 KCl. Using cyclic voltammetry, the number of electrons involved in the morpholine electroxidation mechanism was found to be 1. The application of chronoamperometry showed that the apparent diffusion coefficient (D-0) was 2.99 x 10(-6) cm(2) s(-1). Using square wave voltammetry under the optimized conditions (frequency of 30.0 Hz, pulse amplitude of 100 mV and step potential of 20 mV at pH 10.0), the developed method provided limits of detection and quantification of 2.1 and 6.9 mg L-1, respectively, with a linear range from 5.0 to 100.0 mg L-1 (r = 0.991). Intraday (n = 10) and interday (two consecutive day) precision values assessed as the relative standard deviation for solutions containing 30.0, 60.0, and 90.0 mg L-1 of morpholine were from 0.41 to 5.86% and 0.92 to 3.19%, respectively. The feasibility of the method for the interference-free determination of morpholine was verified by the analysis of synthetic boiler water samples containing CaCO3, Na2SO3, Na3PO4, FeCl3, and humic acid as organic matter. In addition, hydrazine was added as a possible interfering compound because of its widespread use in corrosion inhibition. Recovery values from 90.9 to 109.4% were obtained in the synthetic boiler water, thereby attesting to the accuracy of the method.
引用
收藏
页码:1083 / 1096
页数:14
相关论文
共 50 条
  • [1] Electroanalytical determination of salbutamol in pharmaceutical formulations using cathodically pretreated boron-doped diamond electrode
    Talay Pinar, Pinar
    Ali, Hoshyar Saadi
    Abdullah, Abdullah Abdulwahed
    Yardim, Yavuz
    Senturk, Zuhre
    JOURNAL OF RESEARCH IN PHARMACY, 2018, 22 (03): : 460 - 468
  • [2] Electroanalytical determination of the linuron herbicide using a cathodically pretreated boron-doped diamond electrode: comparison with a boron-doped diamond electrode modified with platinum nanoparticles
    Figueiredo-Filho, Luiz C. S.
    Sartori, Elen R.
    Fatibello-Filho, O.
    ANALYTICAL METHODS, 2015, 7 (02) : 643 - 649
  • [3] Simultaneous determination of hydrochlorothiazide and valsartan in combined dosage forms: Electroanalytical performance of cathodically pretreated boron-doped diamond electrode
    Pires Eisele, Ana Paula
    Mansano, Guilherme Roberto
    de Oliveira, Fernanda Midori
    Casarin, Juliana
    Teixeira Tarley, Cesar Ricardo
    Sartori, Elen Romao
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2014, 732 : 46 - 52
  • [4] DPV and SWV Determination of Estrone Using a Cathodically Pretreated Boron-Doped Diamond Electrode
    Brocenschi, Ricardo F.
    Rocha-Filho, Romeu C.
    Biaggio, Sonia R.
    Bocchi, Nerilso
    ELECTROANALYSIS, 2014, 26 (07) : 1588 - 1597
  • [5] Electroanalytical sensing of indigo carmine dye in water samples using a cathodically pretreated boron-doped diamond electrode
    Silva, Tiago Almeida
    Pereira, Gabriel Fernandes
    Fatibello-Filho, Orlando
    Barrios Eguiluz, Katlin Lyon
    Salazar-Banda, Giancarlo Richard
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2016, 769 : 28 - 34
  • [6] Simultaneous voltammetric determination of synthetic colorants in food using a cathodically pretreated boron-doped diamond electrode
    Medeiros, Roberta A.
    Lourencao, Bruna C.
    Rocha-Filho, Romeu C.
    Fatibello-Filho, Orlando
    TALANTA, 2012, 97 : 291 - 297
  • [7] Determination of Propylthiouracil in Pharmaceuticals by Differential Pulse Voltammetry Using a Cathodically Pretreated Boron-Doped Diamond Electrode
    Sartori, Elen Romao
    Trench, Aline Barrios
    Rocha-Filho, Romeu C.
    Fatibello-Filho, Orlando
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2013, 24 (09) : 1504 - 1511
  • [8] Voltammetric determination of ciprofloxacin in urine samples and its interaction with dsDNA on a cathodically pretreated boron-doped diamond electrode
    Garbellini, Gustavo Stoppa
    Rocha-Filho, Romeu C.
    Fatibello-Filho, Orlando
    ANALYTICAL METHODS, 2015, 7 (08) : 3411 - 3418
  • [9] Electrochemical oxidation and electroanalytical determination of xylitol at a boron-doped diamond electrode
    Lourenco, Anabel S.
    Sanches, Fatima A. C.
    Magalhaes, Renata R.
    Costa, Daniel J. E.
    Ribeiro, Williame F.
    Bichinho, Katia M.
    Salazar-Banda, Giancarlo R.
    Araujo, Mario C. U.
    TALANTA, 2014, 119 : 509 - 516
  • [10] Electroanalytical determination of estriol hormone using a boron-doped diamond electrode
    Santos, Keliana D.
    Braga, Otoniel C.
    Vieira, Iolanda C.
    Spinelli, Almir
    TALANTA, 2010, 80 (05) : 1999 - 2006