Variability of Retinal Thickness Measurements in Tilted or Stretched Optical Coherence Tomography Images

被引:8
|
作者
Uji, Akihito [1 ,2 ]
Abdelfattah, Nizar Saleh [1 ,2 ]
Boyer, David S. [3 ]
Balasubramanian, Siva [1 ,2 ]
Lei, Jianqin [1 ,2 ]
Sadda, SriniVas R. [1 ,2 ]
机构
[1] Doheny Eye Inst, Doheny Image Reading Ctr, 1355 San Pablo St, Los Angeles, CA 90033 USA
[2] Univ Calif Los Angeles, David Geffen Sch Med, Dept Ophthalmol, Los Angeles, CA 90095 USA
[3] Retina Vitreous Associates Med Grp, Beverly Hills, CA USA
来源
关键词
optical coherence tomography; retinal thickness; retina; MACULAR DEGENERATION; PATHOLOGICAL MYOPIA; DOMAIN; EYES; OCT; REPRODUCIBILITY; SEGMENTATION; VOLUME; EDEMA; LASER;
D O I
10.1167/tvst.6.2.1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate the level of inaccuracy of retinal thickness measurements in tilted and axially stretched optical coherence tomography (OCT) images. Methods: A consecutive series of 50 eyes of 50 patients with age-related macular degeneration were included in this study, and Cirrus HD-OCT images through the foveal center were used for the analysis. The foveal thickness was measured in three ways: (1) parallel to the orientation of the A-scan (Tx), (2) perpendicular to the retinal pigment epithelium (RPE) surface in the instrument-displayed aspect ratio image (Ty), and (3) thickness measured perpendicular to the RPE surface in a native aspect ratio image (Tz). Mathematical modeling was performed to estimate the measurement error. Results: The measurement error was larger in tilted images with a greater angle of tilt. In the simulation, with axial stretching by a factor of 2, Ty/Tz ratio was > 1.05 at a tilt angle between 13 degrees to 18 degrees and 72 degrees to 77 degrees, > 1.10 at a tilt angle between 19 degrees to 31 degrees and 59 degrees to 71 degrees, and > 1.20 at an angle ranging from 32 degrees to 58 degrees. Of note with even more axial stretching, the Ty/Tz ratio is even larger. Tx/Tz ratio was smaller than the Ty/Tz ratio at angles ranging from 0 degrees to 54 degrees. The actual patient data showed good agreement with the simulation. The Ty/Tz ratio was greater than 1.05 (5% error) at angles ranging from 13 degrees to 18 degrees and 72 degrees to 77 degrees, > greater than 1.10 (10% error) angles ranging from 19 degrees to 31 degrees and 59 degrees to 71 degrees, and greater than > 1.20 (20% error) angles ranging from 32 degrees to 58 degrees in the images axially stretched by a factor of 2 (b/a = 2), which is typical of most OCT instrument displays. Conclusions: Retinal thickness measurements obtained perpendicular to the RPE surface were overestimated when using tilted and axially stretched OCT images. Translational Relevance: If accurate measurements are to be obtained, images with a native aspect ratio similar to microscopy must be used.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Variability of Retinal Thickness Measurements in Tilted or Stretched Optical Coherence Tomography Images
    Uji, Akihito
    Abdelfattah, Nizar Saleh
    Boyer, David S.
    Balasubramanian, Siva
    Lei, Jianqin
    Sadda, Srinivas R.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [2] Retinal thickness measurements with optical coherence biometry and optical coherence tomography
    Resch, Miklos D.
    Takats, Judit
    Csakany, Bela
    Maneschg, Otto
    Marsovszky, Laszlo
    Szabo, Antal
    Papp, Andras
    Nemeth, Janos
    [J]. SPEKTRUM DER AUGENHEILKUNDE, 2014, 28 (03) : 121 - 125
  • [4] Retinal thickness measurements with optical coherence tomography and the scanning retinal thickness analyzer
    Konno, S
    Akiba, J
    Yoshida, A
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2001, 21 (01): : 57 - 61
  • [5] Reproducibility of retinal thickness measurements using Optical Coherence Tomography
    Wang, RC
    Koozekanani, D
    Herderick, EE
    Roberts, C
    Katz, SE
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S125 - S125
  • [6] Automated measurements of retinal thickness with optical coherence tomography - Reply
    Ray, R
    Stinnett, SS
    Jaffe, GJ
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2005, 140 (02) : 351 - 351
  • [7] Errors in retinal thickness measurements obtained by optical coherence tomography
    Sadda, SR
    Wu, ZQ
    Walsh, AC
    Richine, L
    Dougall, J
    Cortez, R
    LaBree, LD
    [J]. OPHTHALMOLOGY, 2006, 113 (02) : 285 - 293
  • [8] Factors associated with variability in retinal nerve fiber layer thickness measurements obtained by optical coherence tomography
    Wu, Ziqiang
    Vazeen, Matti
    Varma, Rohit
    Chopra, Vikas
    Walsh, Alexander C.
    LaBree, Laurie D.
    Sadda, Srinivas R.
    [J]. OPHTHALMOLOGY, 2007, 114 (08) : 1505 - 1512
  • [9] Reproducibility of retinal thickness measurements using optical coherence tomography (OCT)
    Liebmann, JM
    Baumann, M
    Gentile, RC
    DeLuca, R
    Ozols, A
    Chew, SJ
    Ritch, R
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 4464 - 4464
  • [10] Accuracy of retinal thickness measurements obtained with Cirrus optical coherence tomography
    Keane, P. A.
    Mand, P. S.
    Liakopoulos, S.
    Walsh, A. C.
    Sadda, S. R.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (11) : 1461 - 1467