摘要
Peak distortion due to the injection was measured as a function of injection solvent strength, volume, mass, retention factor, and column selectivity. The concept of a method's sensitivity(s) to injection solvent strength was mathematically defined as a vector of theoretical plate counts compared to an ideal vector that does not change with injection solvent strength. Near ideal sensitivity (s > 0.90) was measured on all columns with all analytes in low volume injections of 1.25 mu L. Increasing the injection volume reduces the measured sensitivity from ideality to a greater extent than increasing the injection mass, with differing values for each column. Using column parameters measured from the hydrophobic-subtraction model and fitting parameters from the acetonitrile excess adsorption isotherm, differences among the columns studied are explained. Decreased ligand density and increased silanol activity provide a consistent peak shape with changes in injection volume or solvent strength. For method development, a quick test is suggested with the ratio of hydrophobic-subtraction column parameters, H/A, to predict the injection solvent sensitivity of a column. As H/A decreases, the sensitivity to injection solvent worsens. Sensitivity to organic modifiers other than acetonitrile are predicted with cited sorbed layer thickness, such that MeOH > EtOH > IPA approximate to THF approximate to MeCN, i.e., a strong MeOH diluent is more ideal (better) than a strong MeCN diluent. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:77 / 89
页数:13
相关论文
共 50 条
  • [1] INFLUENCE OF INJECTION SOLVENT ON THE REVERSED-PHASE CHROMATOGRAPHY OF TRIGLYCERIDES
    TSIMIDOU, M
    MACRAE, R
    JOURNAL OF CHROMATOGRAPHY, 1984, 285 (01): : 178 - 181
  • [2] INTERACTION INDEXES AND SOLVENT EFFECTS IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY
    COLIN, H
    GUIOCHON, G
    JANDERA, P
    ANALYTICAL CHEMISTRY, 1983, 55 (03) : 442 - 446
  • [3] Influence of Solvent Effects on Retention of Small Molecules in Reversed-Phase Liquid Chromatography
    Colin F. Poole
    Chromatographia, 2019, 82 : 49 - 64
  • [4] Organic solvent effects in reversed-phase liquid chromatography in relation to column testing
    K. Le Mapihan
    J. Vial
    A. Jardy
    Chromatographia, 2003, 57 : S163 - S170
  • [5] Organic solvent effects in reversed-phase liquid chromatography in relation to column testing
    Le Mapihan, K
    Vial, J
    Jardy, A
    CHROMATOGRAPHIA, 2003, 57 (Suppl 1) : S163 - S170
  • [6] Influence of Solvent Effects on Retention of Small Molecules in Reversed-Phase Liquid Chromatography
    Poole, Colin F.
    CHROMATOGRAPHIA, 2019, 82 (01) : 49 - 64
  • [7] STATIONARY PHASE EFFECTS IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY
    COLIN, H
    KRSTULOVIC, A
    GUIOCHON, G
    YUN, Z
    JOURNAL OF CHROMATOGRAPHY, 1983, 255 (JAN): : 295 - 309
  • [8] HYDROPHOBIC EFFECTS IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY
    KARGER, BL
    GANT, JR
    HARTKOPF, A
    WEINER, PH
    JOURNAL OF CHROMATOGRAPHY, 1976, 128 (01): : 65 - 78
  • [9] SILANOL EFFECTS IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY
    WELSCH, T
    FRANK, H
    VIGH, G
    JOURNAL OF CHROMATOGRAPHY, 1990, 506 : 97 - 108
  • [10] EFFECTS OF INJECTION SOLVENTS ON THE SEPARATION OF PORPHYRIN AND METALLOPORPHYRIN IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY
    HO, JW
    CANDY, LYF
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1994, 17 (03): : 549 - 558