Inter-observer agreement of vertebral fracture assessment with dual-energy x-ray absorptiometry equipment

被引:2
|
作者
Mostert, Jacob M. [1 ]
Romeijn, Stephan R. [1 ]
Dibbets-Schneider, Petra [1 ]
Rietbergen, Daphne D. D. [1 ]
Pereira Arias-Bouda, Lenka M. [1 ,2 ]
Goetz, Christoph [3 ]
DiFranco, Matthew D. [3 ]
Dimai, Hans Peter [4 ]
Grootjans, Willem [1 ]
机构
[1] Leiden Univ, Dept Radiol, Med Ctr, Leiden, Netherlands
[2] Alrijne Hosp, Dept Radiol, Leiderdorp, Netherlands
[3] Image Biopsy Lab, Vienna, Austria
[4] Med Univ Graz, Dept Internal Med, Div Endocrinol & Diabetol, Graz, Austria
关键词
Vertebral fracture assessment; Dual-energy x-ray absorptiometry; Vertebral morphometry; Inter-observer agreement; Osteoporosis; ASSESSMENT VFA; PREVALENT; OSTEOPOROSIS; DEFORMITIES; DIAGNOSIS; WOMEN; RISK; DXA;
D O I
10.1007/s11657-021-01046-w
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose To investigate the time and effort needed to perform vertebral morphometry, as well as inter-observer agreement for identification of vertebral fractures on vertebral fracture assessment (VFA) images. Methods Ninety-six images were retrospectively selected, and three radiographers independently performed semi-automatic 6-point morphometry. Fractures were identified and graded using the Genant classification. Time needed to annotate each image was recorded, and reader fatigue was assessed using a modified Simulator Sickness Questionnaire (SSQ). Inter-observer agreement was assessed per-patient and per-vertebra for detecting fractures of all grades (grades 1-3) and for grade 2 and 3 fractures using the kappa statistic. Variability in measured vertebral height was evaluated using the intraclass correlation coefficient (ICC). Results Per-patient agreement was 0.59 for grades 1-3 fracture detection, and 0.65 for grades 2-3 only. Agreement for per-vertebra fracture classification was 0.92. Vertebral height measurements had an ICC of 0.96. Time needed to annotate VFA images ranged between 91 and 540 s, with a mean annotation time of 259 s. Mean SSQ scores were significantly lower at the start of a reading session (1.29; 95% CI: 0.81-1.77) compared to the end of a session (3.25; 95% CI: 2.60-3.90; p < 0.001). Conclusion Agreement for detection of patients with vertebral fractures was only moderate, and vertebral morphometry requires substantial time investment. This indicates that there is a potential benefit for automating VFA, both in improving inter-observer agreement and in decreasing reading time and burden on readers.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Inter-observer agreement of vertebral fracture assessment with dual-energy x-ray absorptiometry equipment
    Jacob M. Mostert
    Stephan R. Romeijn
    Petra Dibbets-Schneider
    Daphne D. D. Rietbergen
    Lenka M. Pereira Arias-Bouda
    Christoph Götz
    Matthew D. DiFranco
    Hans Peter Dimai
    Willem Grootjans
    Archives of Osteoporosis, 2022, 17
  • [2] Vertebral fracture assessment by new dual-energy X-ray absorptiometry
    Bazzocchi, Alberto
    Spinnato, Paolo
    Fuzzi, Federica
    Diano, Danila
    Morselli-Labate, Antonio M.
    Sassi, Claudia
    Salizzoni, Eugenio
    Battista, Giuseppe
    Guglielmi, Giuseppe
    BONE, 2012, 50 (04) : 836 - 841
  • [3] RELIABILITY OF VERTEBRAL FRACTURE ASSESSMENT ON DUAL-ENERGY X-RAY ABSORPTIOMETRY
    Yasmine, M.
    Mariem, S.
    Miladi, S.
    Fazaa, A.
    Fguiri, E.
    Souebni, L.
    Ouenniche, K.
    Kassab, S.
    Chekili, S.
    Ben Abdelghani, K.
    Laatar, A.
    ANNALS OF THE RHEUMATIC DISEASES, 2021, 80 : 835 - 835
  • [4] Clinical review: Clinical applications of vertebral fracture assessment by dual-energy x-ray absorptiometry
    Lewiecki, E. Michael
    Laster, Andrew J.
    JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2006, 91 (11): : 4215 - 4222
  • [5] Vertebral fracture assessment by dual X-ray absorptiometry
    Fechtenbaum, J.
    Kolta, S.
    Briot, K.
    Roux, C.
    OSTEOPOROSIS INTERNATIONAL, 2017, 28 (03) : 1153 - 1154
  • [6] Vertebral fracture assessment by dual X-ray absorptiometry
    J. Fechtenbaum
    S. Kolta
    K. Briot
    C. Roux
    Osteoporosis International, 2017, 28 : 1153 - 1154
  • [7] DUAL-ENERGY X-RAY ABSORPTIOMETRY OF THE CALCANEUS - COMPARISON WITH VERTEBRAL DUAL-ENERGY X-RAY ABSORPTIOMETRY AND QUANTITATIVE COMPUTED-TOMOGRAPHY
    LAVALJEANTET, AM
    BERGOT, C
    WILLIAMS, M
    DAVIDSON, K
    LAVALJEANTET, M
    CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (01) : 14 - 18
  • [8] Calcaneus Ultrasonometry and Dual-Energy X-Ray Absorptiometry for the Evaluation of Vertebral Fracture Risk
    B. Frediani
    C. Acciai
    P. Falsetti
    F. Baldi
    G. Filippou
    C. Siagkri
    A. Spreafico
    M. Galeazzi
    R. Marcolongo
    Calcified Tissue International, 2006, 79 : 223 - 229
  • [9] Calcaneus ultrasonometry and dual-energy X-ray absorptiometry for the evaluation of vertebral fracture risk
    Frediani, B.
    Acciai, C.
    Falsetti, P.
    Baldi, F.
    Filippou, G.
    Siagkri, C.
    Spreafico, A.
    Galeazzi, M.
    Marcolongo, R.
    CALCIFIED TISSUE INTERNATIONAL, 2006, 79 (04) : 223 - 229
  • [10] New dual-energy X-ray absorptiometry equipment in the assessment of vertebral fractures: technical limits and software accuracy
    Alberto Bazzocchi
    Danila Diano
    Giuseppe Battista
    Ugo Albisinni
    Cristina Rossi
    Giuseppe Guglielmi
    Skeletal Radiology, 2012, 41 : 823 - 829