Evaluation of ADC testing systems using ADC transfer standard

被引:3
|
作者
Haasz, V [1 ]
Roztocil, J [1 ]
Slepicka, D [1 ]
机构
[1] Czech Tech Univ, Fac Elect Engn, Dept Measurement, CZ-16627 Prague, Czech Republic
关键词
analog-to-digital converter (ADC) testing; analog-to-digital (AD) transfer standard; comparative measurements; ENOB; frequency spectrum averaging; generator instability; SINAD; THD;
D O I
10.1109/TIM.2005.847119
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A transportable high-stability reference AD device was designed and built to compare the systems for testing the dynamic quality of analog-to-digital converters (ADCs) or analog-to-digital (AD) modules. Three different input modules can be used in the frequency range up to 5 MHz. First, this paper refers to the results of an AD device application in the frequency range of the testing signal up to 100 kHz, where four ADC testing systems were compared in different laboratories, and a short time amplitude and/or frequency instability of testing signal generators was evaluated. Second, it describes the first experience of its application in the frequency range from 100-5 MHz.
引用
收藏
页码:1150 / 1155
页数:6
相关论文
共 50 条
  • [11] ADC modeling and testing
    Chiorboli, G
    Morandi, C
    IMTC/2001: PROCEEDINGS OF THE 18TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3: REDISCOVERING MEASUREMENT IN THE AGE OF INFORMATICS, 2001, : 1992 - 1999
  • [12] Using of AM and FM signal for ADC Testing
    Vedral, Josef
    Fexa, Pavel
    Svatos, Jakub
    2010 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE I2MTC 2010, PROCEEDINGS, 2010,
  • [13] Chirp based method for ADC testing:: Simulation and evaluation
    Holub, J
    Smíd, R
    Vedral, J
    THIRD INTERNATIONAL CONFERENCE ON ADVANCED A/D AND D/A CONVERSION TECHNIQUES AND THEIR APPLICATIONS, 1999, (466): : 94 - 96
  • [14] Comparative measurements of parameters of systems used for ADC testing
    Haasz, V
    Roztocil, J
    Stepicka, D
    COMPUTER STANDARDS & INTERFACES, 2004, 26 (01) : 21 - 30
  • [15] Bandpass Δσ ADC using pipelined SAR ADC
    Oh, S.
    Kim, K.
    Chae, H.
    ELECTRONICS LETTERS, 2020, 56 (10) : 480 - +
  • [16] ADC TESTING USING SINGLE CHANNEL REFERENCE SOURCE
    Kochan, Roman
    ELECTRICAL AND CONTROL TECHNOLOGIES, 2011, : 82 - 87
  • [17] Multi-Histogram ADC BIST System for ADC Linearity Testing
    Chan, Koay Soon
    Nordin, Nuzrul Fahmi
    Chan, Kim Chon
    Lok, Terk Zyou
    Yong, Chee Wai
    2013 22ND ASIAN TEST SYMPOSIUM (ATS), 2013, : 213 - 214
  • [18] ADC FUNCTIONAL TESTING USING ARTIFICIAL IMMUNE SYSTEM
    de Souza, Cleonilson Protasio
    Torres, Claudio Leao
    Freire, Raimundo C. S.
    de Assis, Francisco M.
    XIX IMEKO WORLD CONGRESS: FUNDAMENTAL AND APPLIED METROLOGY, PROCEEDINGS, 2009, : 698 - 701
  • [19] Static Range Testing of ADC
    Sindhu, B. A.
    Hariharan, K.
    2013 INTERNATIONAL CONFERENCE ON INFORMATION COMMUNICATION AND EMBEDDED SYSTEMS (ICICES), 2013, : 1094 - 1098
  • [20] Optimum DAC and ADC Testing
    Max, Solomon
    2008 IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-5, 2008, : 573 - 578