Estimation of lot-to-lot relative potency in the preparation of radioimmunoassay calibrators

被引:1
|
作者
Livesey, JH [1 ]
Yandle, TG
Evans, MJ
Ellis, MJ
机构
[1] Chirstchurch Hosp, Dept Endocrinol, Christchurch, New Zealand
[2] Christchurch Sch Med & Hlth Sci, Dept Med, Christchurch, New Zealand
来源
关键词
calibration; potency; immunoassay; standardisation; radioimmunoassay;
D O I
10.1081/IAS-120022935
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The aim of this study was to determine guidelines for estimating lotto-lot differences in the potency of calibrator materials or batches of standards for radioimmunoassays. Thirty one lots of standards for thirteen different analytes were compared to the previous lot for that analyte with the relative potency computed by nine different methods. Assays were performed manually. The nine different calculation methods included non-simultaneous fitting of pairs of standard curves, full or partial simultaneous fitting, and least squares or robust minimisation. The simultaneous methods were found superior to the non-simultaneous in minimising the variance of the relative potency estimates, while robust fitting procedures did not result in a lower variance than least-squares minimisation. The root mean square coefficient of variation for the simultaneous estimation of the relative potency by least squares was 6.1 %. On this basis, it is recommended that relative potency estimations in radioimmunoassay be based on at least eight independent pair-wise standard curve comparisons. Additional guidelines for preparing and comparing batches of standards are also given.
引用
收藏
页码:243 / 263
页数:21
相关论文
共 50 条
  • [31] Comparison of six regression-based lot-to-lot verification approaches
    Koh, Norman Wen Xuan
    Markus, Corey
    Loh, Tze Ping
    Lim, Chun Yee
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2022, 60 (08) : 1175 - 1185
  • [32] Effect of roller compaction on nisin raw material lot-to-lot variations
    Habib, WA
    Takka, S
    Sakr, A
    PHARMAZEUTISCHE INDUSTRIE, 2000, 62 (11): : 914 - 918
  • [33] LOT-TO-LOT VARIATIONS IN ELECTROMIGRATION PERFORMANCE FOR THIN-FILM MICROCIRCUITS
    SMITH, PM
    LLOYD, JR
    PROKOP, GS
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1984, 2 (02): : 220 - 224
  • [34] Sequential screening in semiconductor manufacturing .2. Exploiting lot-to-lot variability
    Ou, JH
    Wein, LM
    OPERATIONS RESEARCH, 1996, 44 (01) : 196 - 205
  • [35] Impact of lot-to-lot variability of cation exchange chromatography resin on process performance
    Wahome, Jane
    Zhou, Weichang
    Kundu, Amitava
    BIOPHARM INTERNATIONAL, 2008, 21 (05) : 48 - +
  • [36] Reproducibility of Elecsys® Anti-TSHR Test Results in a Lot-to-Lot Comparison
    Hermsen, D.
    Eckstein, A.
    Schinner, S.
    Willenberg, H. S.
    Thiel, A.
    Scherbaum, W. A.
    Schott, M.
    HORMONE AND METABOLIC RESEARCH, 2010, 42 (04) : 295 - 297
  • [37] Lot-to-Lot Reproducibility of Standardized Allergen Extracts in IgE In Vitro Diagnostic Tests
    Dinh, Lauren
    Sinson, Edsel
    Nguyen, Steven
    Liao, Cindy
    Rodems, Kelline
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 2020, 145 (02) : AB151 - AB151
  • [38] Coordinating model and simulation in lot-to-lot supply chain based on common replenishment epoch
    School of Management, Northeastern University, Shenyang 110004, China
    Jisuanji Jicheng Zhizao Xitong, 2008, 3 (482-486):
  • [39] Lot-to-lot variation in anti-hepatitis C signal-to-cutoff ratio
    Dufour, DR
    CLINICAL CHEMISTRY, 2004, 50 (05) : 958 - 960
  • [40] Lot-to-lot Consistency, Safety, Tolerability and Immunogenicity of an Investigational Hexavalent Vaccine in US Infants
    Block, Stanley L.
    Klein, Nicola P.
    Sarpong, Kwabena
    Russell, Stephen
    Fling, John
    Petrecz, Maria
    Flores, Sheryl
    Xu, Jin
    Liu, Guanghan
    Stek, Jon E.
    Foglia, Ginamarie
    Lee, Andrew W.
    PEDIATRIC INFECTIOUS DISEASE JOURNAL, 2017, 36 (02) : 202 - 208