On the relative contributions of absorption and scattering to ultrasound attenuation in trabecular bone: A simulation study

被引:0
|
作者
Kaufman, JJ [1 ]
Luo, GM [1 ]
Siffert, RS [1 ]
机构
[1] Mt Sinai Sch Med, Dept Orthoped, New York, NY 10029 USA
来源
2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2 | 2003年
关键词
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound has been proposed as a means to non-invasively assess bone and particularly bone strength and fracture risk. Although there has been some success in this application, there is still much that is unknown regarding the propagation of ultrasound through bone. Because strength and fracture risk are a function of both bone mineral density as well as architectural structure and tissue quality, this study was carried out to further elucidate the mechanisms of interaction between ultrasound and bone. Frequency-dependent attenuation of an ultrasound wave in trabecular bone has been shown to be strongly dependent on bone mass and architecture and is currently used in several clinical devices for bone assessment. Since attenuation is due to both absorption by the biological tissues per se and scattering, it is of interest to understand the relative contributions of each. A sample of calcaneal trabecular bone was scanned with micro-CT and subjected to morphological image processing (erosions and dilations) to obtain a total of 11 three-dimensional (3D) data sets. Eleven two-dimensional (2D) slices obtained from the 3D data sets were then analyzed to evaluate their bone volume fractions (VF). Computer simulations of ultrasonic propagation through each of the 11 2D bone slices, which varied in VF from 0.088 - 0.181, were carried out in one of two modes. In the first instance, the component tissues (i.e., marrow and bone) were lossy, while in the second set of simulations the component tissues were lossless. In both cases the slope of the attenuation was computed over the frequency range 300 kHz - 900 kHz for the entire data set. Results obtained showed an average reduction in the attenuation slope of only 4.4% (SD=1.8%) in the lossles case as compared (pairwise) with the lossy case. This data indicates that scattering is the primary mechanism in trabecular bone with respect to overall attenuation measurements, and further suggests that models relating scattering to bone architecture and mass should be developed to further enhance the ability of ultrasound to non-invasively assess bone.
引用
收藏
页码:1519 / 1523
页数:5
相关论文
共 50 条
  • [41] Statistics of Envelope of High-Frequency Ultrasonic Backscatter from Trabecular Bone: Simulation Study
    Litniewski, Jerzy
    ARCHIVES OF ACOUSTICS, 2010, 35 (03) : 349 - 360
  • [42] Simulation study of axial ultrasound transmission in heterogeneous cortical bone model
    Takano, Koki
    Nagatani, Yoshiki
    Matsukawa, Mami
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (07)
  • [43] Improving broadband ultrasound attenuation assessment in cancellous bone by mitigating the influence of cortical bone: Phantom and in-vitro study
    Tasinkevych, Yuriy
    Falinska, Katarzyna
    Lewin, Peter A.
    Litniewski, Jerzy
    ULTRASONICS, 2019, 94 : 382 - 390
  • [44] Modeling ultrasound attenuation in porous structures with mono-disperse random pore distributions using the independent scattering approximation: a 2D simulation study
    Yousefian, Omid
    Karbalaeisadegh, Yasamin
    Muller, Marie
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (15):
  • [45] Familial resemblance of bone mineral density (BMD) and calcaneal ultrasound attenuation: The BMD in mothers and daughters study
    Danielson, ME
    Cauley, JA
    Baker, CE
    Newman, AB
    Dorman, JS
    Towers, JD
    Kuller, LH
    JOURNAL OF BONE AND MINERAL RESEARCH, 1999, 14 (01) : 102 - 110
  • [46] Simultaneous reconstruction of absorption and scattering maps with ultrasound localization: feasibility study using transmission geometry
    Huang, MM
    Xie, TQ
    Chen, NG
    Zhu, Q
    APPLIED OPTICS, 2003, 42 (19) : 4102 - 4114
  • [47] Simultaneous reconstruction of absorption and scattering maps with ultrasound localization: Feasibility study using transmission geometry
    Huang, Minming
    Xie, Tuqiang
    Chen, Nan Guang
    Zhu, Quing
    Applied Optics, 2003, 42 (19): : 4102 - 4114
  • [48] Genetic and environmental contributions to the association between quantitative ultrasound and bone mineral density measurements: A twin study
    Howard, GM
    Nguyen, TV
    Harris, M
    Kelly, PJ
    Eisman, JA
    JOURNAL OF BONE AND MINERAL RESEARCH, 1998, 13 (08) : 1318 - 1327
  • [49] Frequency-dependent analysis of ultrasound apparent absorption coefficient in multiple scattering porous media: application to cortical bone
    Yousefian, Omid
    Karbalaeisadegh, Yasamin
    Mueller, Marie
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03):
  • [50] Heel quantitative ultrasound (QUS) predicts incident fractures independently of trabecular bone score (TBS), bone mineral density (BMD), and FRAX: the OsteoLaus Study
    Antoine Métrailler
    Didier Hans
    Olivier Lamy
    Elena Gonzalez Rodriguez
    Enisa Shevroja