Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc

被引:8
|
作者
Popple, Richard A. [1 ]
Sullivan, Rodney J. [1 ]
Yuan, Yuan [1 ]
Wu, Xingen [1 ]
Covington, Elizabeth L. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, 176F RM 2240H,619 19TH St S,1720 2nd Ave South, Birmingham, AL 35249 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2021年 / 22卷 / 12期
关键词
dosimetry; quality assurance; radiosurgery; ASTRO PRACTICE GUIDELINE; PLAN QUALITY; AMERICAN SOCIETY; SINGLE-ISOCENTER; FILM DOSIMETRY; RADIOLOGY ACR; RADIOSURGERY; DELIVERY; COLLEGE;
D O I
10.1002/acm2.13438
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate a two-dimensional diode array for patient-specific quality assurance of VMAT stereotactic radiosurgery (SRS) plans. Methods The diode array, an SRS MapCHECK (SRSMC), was composed of a 77 mm x77 mm face-centered array having a spacing of 2.47 mm. Sixty SRS plans were selected from our clinical database, 30 for treatment of a single target and 30 for multiple targets. The target sizes ranged from 2.4 mm to 44.7 mm equivalent diameter (median 8.7 mm). The plans were delivered to the diode array. For multiple target plans, two measurements were obtained at two locations, one corresponding to the largest target and the other to the smallest target. Gamma using a 3%/1 mm criteria and the dose to the center diode were compared with radiochromic film (RCF). Dose to selected regions of the detector electronics was calculated. Results The mean difference between the center diode and RCF was -1.2%. For a threshold of at least 95% of detectors/pixels having gamma < 1 for a 3%/1 mm criteria, SRSMC and RCF gave consistent results for 79 of the 90 measurements. For plans with an arc having a patient support angle of 90 degrees or 270 degrees, the median dose to the electronics was 0.65% of the prescription dose. Conclusions SRSMC is an efficient tool for accurate patient-specific quality assurance of VMAT single and multiple target radiosurgery, yielding similar clinical decisions as radiochromic film.
引用
收藏
页码:203 / 210
页数:8
相关论文
共 50 条
  • [31] Evaluation of a high resolution diode array for CyberKnife quality assurance
    Garcia-Fuentes, Juan D. D.
    Sevillano, David
    Colmenares, Rafael
    Capuz, Ana B. B.
    Moris, Rafael
    Camara, Miguel
    Galiano, Pablo
    Williamson, Sandra
    Bejar, Maria J.
    Prieto, Daniel
    Garcia-Vicente, Feliciano
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09):
  • [32] EVALUATION OF A HELICAL DIODE ARRAY DEVICE FOR IMRT QUALITY ASSURANCE
    Eenink, M.
    Van Egmond, J.
    Roijen, E.
    de Goede, M.
    Petoukhova, A.
    Rietveld, P.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S456 - S456
  • [33] Patient-specific quality assurance method for VMAT treatment delivery
    Schreibmann, Eduard
    Dhabaan, Anees
    Elder, Eric
    Fox, Tim
    MEDICAL PHYSICS, 2009, 36 (10) : 4530 - 4535
  • [34] Analysis of ion beam teletherapy patient-specific quality assurance
    Liu, Xiaoli
    Deng, Yu
    Schlegel, Nicki
    Huang, Zhijie
    Moyers, Michael F.
    MEDICAL DOSIMETRY, 2019, 44 (01) : 43 - 50
  • [35] Survey of patient-specific quality assurance practice for IMRT and VMAT
    Chan, Gordon H.
    Chin, Lee C. L.
    Abdellatif, Ady
    Bissonnette, Jean-Pierre
    Buckley, Lesley
    Comsa, Daria
    Granville, Dal
    King, Jenna
    Rapley, Patrick L.
    Vandermeer, Aaron
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (07): : 155 - 164
  • [36] Sensitivity evaluation of two commercial quality assurance systems to organ-dose variations of patient-specific VMAT plans
    Oderinde, Oluwaseyi M.
    Du Plessis, Freek
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) : 132 - 139
  • [37] Patient-specific quality assurance for IMRT delivery: A multicentre study
    Hizam, Diyana Afrina
    Jong, Wei Loong
    Zin, Hafiz Mohd
    Ng, Kwan Hoong
    Ung, Ngie Min
    RADIATION PHYSICS AND CHEMISTRY, 2023, 209
  • [38] Two-dimensional Er:YSGG microlaser array pumped with a monolithic two-dimensional laser diode array
    Waarts, Robert, 1738, Optical Soc of America, Washington, DC, United States (19):
  • [39] Assessing the suitability of a two-dimensional array for routine quality assurance checks of flatness and symmetry
    Ibrahim, Salma
    Seshadri, Venkatakrishnan
    Charles, Paul
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (04) : 1451 - 1460
  • [40] Assessing the suitability of a two-dimensional array for routine quality assurance checks of flatness and symmetry
    Salma Ibrahim
    Venkatakrishnan Seshadri
    Paul Charles
    Physical and Engineering Sciences in Medicine, 2020, 43 : 1451 - 1460