Effect of Embedded Dexamethasone in Cochlear Implant Array on Insertion Forces in an Artificial Model of Scala Tympani

被引:11
|
作者
Yann Nguyen [1 ,2 ,3 ]
Bernardeschi, Daniele [1 ,2 ,3 ]
Kazmitcheff, Guillaume [1 ]
Miroir, Mathieu [1 ]
Vauchel, Thomas [1 ]
Ferrary, Evelyne [1 ,2 ,3 ]
Sterkers, Olivier [1 ,2 ,3 ]
机构
[1] INSERM, UMR, Paris, France
[2] Hosp Pitie Salpetriere, AP HP, Unit Otol Auditory Implants & Skull Base Surg, Dept Otolaryngol, Paris, France
[3] Univ Paris 06, Sorbonne Univ, Paris, France
关键词
Array insertion; Drug delivery; Silicone; Trauma; RESIDUAL HEARING; DRUG-RELEASE; PRESERVATION; CELLS;
D O I
10.1097/MAO.0000000000000521
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis Loading otoprotective drug into cochlear implant might change its mechanical properties, thus compromising atraumatic insertion. This study evaluated the effect of incorporation of dexamethasone (DXM) in the silicone of cochlear implant arrays on insertion forces. Background Local administration of DXM with embedded array can potentially reduce inflammation and fibrosis after cochlear implantation procedure to improve hearing preservation and reduce long-term impedances. Methods Four models of arrays have been tested: 0.5-mm distal diameter array (n = 5) used as a control, drug-free 0.4-mm distal diameter array (n = 5), 0.4-mm distal diameter array with 1% eluded DXM silicone (n = 5), and 0.4-mm distal diameter array with 10% eluded DXM silicone (n = 5). Via a motorized insertion bench, each array has been inserted into an artificial scala tympani model. The forces were recorded by a 6-axis force sensor. Each array was tested seven times for a total number of 140 insertions. Results During the first 10-mm insertion, no difference between the four models was observed. From 10- to 24-mm insertion, the 0.5-mm distal diameter array presented higher insertion forces than the drug-free 0.4-mm distal diameter arrays, with or without DXM. Friction forces for drug-free 0.4-mm distal diameter array and 0.4-mm distal diameter DXM eluded arrays were similar on all insertion lengths. Conclusion Incorporation of DXM in silicone for cochlear implant design does not change electrode array insertion forces. It does not raise the risk of trauma during array insertion, making it suitable for long-term in situ administration to the cochlea.
引用
收藏
页码:354 / 358
页数:5
相关论文
共 50 条
  • [31] Optimal path generation in scala tympani and path planning for robotic cochlear implant of perimodiolar electrode
    Wang, Zuo
    Li, Jianjun
    Wu, Yongzhen
    Zhu, Ruifeng
    Wang, Binrui
    Zhao, Kai
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2020, 234 (06) : 578 - 589
  • [32] Cochlear Size Assessment Predicts Scala Tympani Volume and Electrode Insertion Force- Implications in Robotic Assisted Cochlear Implant Surgery (vol 8, 723897, 2021)
    Dhanasingh, Anandhan
    Swords, Chloe
    Bance, Manohar
    Van Rompaey, Vincent
    van de Heyning, Paul
    FRONTIERS IN SURGERY, 2021, 8
  • [33] Evaluation of Insertion Forces and Cochlea Trauma Following Robotics-Assisted Cochlear Implant Electrode Array Insertion
    Kaufmann, Christopher R.
    Henslee, Allan M.
    Claussen, Alex
    Hansen, Marlan R.
    OTOLOGY & NEUROTOLOGY, 2020, 41 (05) : 631 - 638
  • [34] Measurement of Forces at the Tip of a Cochlear Implant During Insertion
    Wade, Scott A.
    Fallon, James B.
    Wise, Andrew K.
    Shepherd, Robert K.
    James, Natalie L.
    Stoddart, Paul R.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (04) : 1177 - 1186
  • [35] Beyond the phantom: Unroofing the scala vestibuli in a fresh temporal bone as a model for cochlear implant insertion experiments
    Smetak, Miriam R.
    Riojas, Katherine E.
    Sharma, Rahul K.
    Labadie, Robert F.
    JOURNAL OF NEUROSCIENCE METHODS, 2022, 382
  • [36] Durability of Hearing Preservation after Cochlear Implantation with Conventional-Length Electrodes and Scala Tympani Insertion
    Sweeney, Alex D.
    Hunter, Jacob B.
    Carlson, Matthew L.
    Rivas, Alejandro
    Bennett, Marc L.
    Gifford, Rene H.
    Noble, Jack H.
    Haynes, David S.
    Labadie, Robert F.
    Wanna, George B.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2016, 154 (05) : 907 - 913
  • [37] Experimental Validation of a Three-Dimensional Heat Transfer Model Within the Scala Tympani With Application to Magnetic Cochlear Implant Surgery
    Esmailie, Fateme
    Francoeur, Mathieu
    Ameel, Tim
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (09) : 2821 - 2832
  • [38] Simultaneous Bilateral Cochlear Implantation in an Adult with Bilateral Ossification of the Scala Tympani: Image-Guided Planning of Scala Vestibuli Insertion and Postoperative Outcomes
    De Bie, Charlotte
    Clement, Charis
    Mertens, Griet
    Brandt, Anouk Hofkens-Van den
    Lammers, Marc
    Vanderveken, Olivier M.
    Van Rompaey, Vincent
    B-ENT, 2024, 20 (04)
  • [39] Optical cochlear implant: evaluation of insertion forces of optical fibres in a cochlear model and of traumata in human temporal bones
    Balster, Sven
    Wenzel, Gentiana I.
    Warnecke, Athanasia
    Steffens, Melanie
    Rettenmaier, Alexander
    Zhang, Kaiyin
    Lenarz, Thomas
    Reuter, Guenter
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 (01): : 19 - 28
  • [40] Effect of deep insertion of the cochlear implant electrode array on pitch estimation and speech perception
    Hamzavi, Jafar
    Arnoldner, Christoph
    ACTA OTO-LARYNGOLOGICA, 2006, 126 (11) : 1182 - 1187