Impact of CT attenuation correction method on quantitative respiratory-correlated (4D) PET/CT imaging

被引:16
|
作者
Nyflot, Matthew J. [1 ]
Lee, Tzu-Cheng [2 ]
Alessio, Adam M. [3 ]
Wollenweber, Scott D. [4 ]
Stearns, Charles W. [4 ]
Bowen, Stephen R. [1 ,3 ]
Kinahan, Paul E. [3 ]
机构
[1] Univ Washington, Dept Radiat Oncol, Seattle, WA 98195 USA
[2] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[3] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
[4] GE Healthcare, Waukesha, WI 53188 USA
关键词
respiratory-correlated PET/CT; CT attenuation correction; 4DPET; 4DCT; quantification; CELL LUNG-CANCER; TOMOGRAPHY FDG-PET; MOTION; VOLUME; DELINEATION; IMAGES; PHASE; RECONSTRUCTION; REGISTRATION; RADIOTHERAPY;
D O I
10.1118/1.4903282
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Respiratory-correlated positron emission tomography (PET/CT) 4D PET/CT is used to mitigate errors from respiratory motion; however, the optimal CT attenuation correction (CTAC) method for 4D PET/CT is unknown. The authors performed a phantom study to evaluate the quantitative performance of CTAC methods for 4D PET/CT in the ground truth setting. Methods: A programmable respiratory motion phantom with a custom movable insert designed to emulate a lung lesion and lung tissue was used for this study. The insert was driven by one of five waveforms: two sinusoidal waveforms or three patient-specific respiratory waveforms. 3DPET and 4DPET images of the phantom under motion were acquired and reconstructed with six CTAC methods: helical breath-hold (3DHEL), helical free-breathing (3DMOT), 4D phase-averaged (4DAVG), 4D maximum intensity projection (4DMIP), 4D phase-matched (4DMATCH), and 4D end-exhale (4DEXH) CTAC. Recovery of SUVmax, SUVmean, SUVpeak, and segmented tumor volume was evaluated as RCmax, RCmean, RCpeak, and RCvol, representing percent difference relative to the static ground truth case. Paired Wilcoxon tests and Kruskal-Wallis ANOVA were used to test for significant differences. Results: For 4DPET imaging, the maximum intensity projection CTAC produced significantly more accurate recovery coefficients than all other CTAC methods (p < 0.0001 over all metrics). Over all motion waveforms, ratios of 4DMIP CTAC recovery were 0.2 +/- 5.4, -1.8 +/- 6.5, -3.2 +/- 5.0, and 3.0 +/- 5.9 for RCmax, RCpeak, RCmean, and RCvol. In comparison, recovery coefficients for phase-matched CTAC were -8.4 +/- 5.3, -10.5 +/- 6.2, -7.6 +/- 5.0, and -13.0 +/- 7.7 for RCmax, RCpeak, RCmean, and RCvol. When testing differences between phases over all CTAC methods and waveforms, end-exhale phases were significantly more accurate (p = 0.005). However, these differences were driven by the patient-specific respiratory waveforms; when testing patient and sinusoidal waveforms separately, patient waveforms were significantly different between phases (p < 0.0001) while the sinusoidal waveforms were not significantly different (p = 0.98). When considering only the subset of 4DMATCH images that corresponded to the end-exhale image phase, 4DEXH, mean and interquartile range were similar to 4DMATCH but variability was considerably reduced. Conclusions: Comparative advantages in accuracy and precision of SUV metrics and segmented volumes were demonstrated with the use of the maximum intensity projection and end-exhale CT attenuation correction. While respiratory phase-matched CTAC should in theory provide optimal corrections, image artifacts and differences in implementation of 4DCT and 4DPET sorting can degrade the benefit of this approach. These results may be useful to guide the implementation, analysis, and development of respiratory-correlated thoracic PET/CT in the radiation oncology and diagnostic settings. (c) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:110 / 120
页数:11
相关论文
共 50 条
  • [41] Optimization in 4D CT/PET acquisitions
    Mari, A.
    Valenti, M.
    Maggi, S.
    Cardinali, M.
    Ascoli, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S507 - S507
  • [42] Low Dose CT Attenuation Correction for Brain imaging with PET
    Bonab, A. A.
    Alpert, N. M.
    Liu, B.
    Fischman, A. J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S258 - S258
  • [43] Dose optimization: a review of CT imaging for PET attenuation correction
    Samuel L. Brady
    Barry L. Shulkin
    Clinical and Translational Imaging, 2017, 5 : 359 - 371
  • [44] Dose optimization: a review of CT imaging for PET attenuation correction
    Brady, Samuel L.
    Shulkin, Barry L.
    CLINICAL AND TRANSLATIONAL IMAGING, 2017, 5 (04) : 359 - 371
  • [45] Different CT protocols for attenuation correction in cardiac FDG imaging with PET/CT: a comparison with conventional PET
    Souvatzoglou, M
    Bengel, F
    Fernolendt, H
    Schwaiger, M
    Nekolla, S
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S317 - S318
  • [46] Challenges of Attenuation Correction in Conventional Respiratory Motion Corrected PET/CT
    Mawlawi, O.
    Meier, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E552 - E552
  • [47] Correspondence of respiratory-correlated CT data sets acquired using 4-D CT with data sets acquired using occlusion spirometry
    Starkschall, G
    Forster, K
    Fitzpatrick, M
    Stevens, C
    MEDICAL PHYSICS, 2004, 31 (06) : 1875 - 1875
  • [48] Automatic deformabie registration of the lung for 4D respiratory correlated CT (RCCT) datasets
    Söhn, M
    Birkner, M
    Yan, D
    Mageras, G
    Alber, M
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S59 - S59
  • [49] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT.
    Osman, MM
    Cohade, C
    Wahl, RL
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 305P - 305P
  • [50] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Medhat M. Osman
    Christian Cohade
    Yuji Nakamoto
    Richard L. Wahl
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 603 - 606