Impact of CT attenuation correction method on quantitative respiratory-correlated (4D) PET/CT imaging

被引:16
|
作者
Nyflot, Matthew J. [1 ]
Lee, Tzu-Cheng [2 ]
Alessio, Adam M. [3 ]
Wollenweber, Scott D. [4 ]
Stearns, Charles W. [4 ]
Bowen, Stephen R. [1 ,3 ]
Kinahan, Paul E. [3 ]
机构
[1] Univ Washington, Dept Radiat Oncol, Seattle, WA 98195 USA
[2] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[3] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
[4] GE Healthcare, Waukesha, WI 53188 USA
关键词
respiratory-correlated PET/CT; CT attenuation correction; 4DPET; 4DCT; quantification; CELL LUNG-CANCER; TOMOGRAPHY FDG-PET; MOTION; VOLUME; DELINEATION; IMAGES; PHASE; RECONSTRUCTION; REGISTRATION; RADIOTHERAPY;
D O I
10.1118/1.4903282
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Respiratory-correlated positron emission tomography (PET/CT) 4D PET/CT is used to mitigate errors from respiratory motion; however, the optimal CT attenuation correction (CTAC) method for 4D PET/CT is unknown. The authors performed a phantom study to evaluate the quantitative performance of CTAC methods for 4D PET/CT in the ground truth setting. Methods: A programmable respiratory motion phantom with a custom movable insert designed to emulate a lung lesion and lung tissue was used for this study. The insert was driven by one of five waveforms: two sinusoidal waveforms or three patient-specific respiratory waveforms. 3DPET and 4DPET images of the phantom under motion were acquired and reconstructed with six CTAC methods: helical breath-hold (3DHEL), helical free-breathing (3DMOT), 4D phase-averaged (4DAVG), 4D maximum intensity projection (4DMIP), 4D phase-matched (4DMATCH), and 4D end-exhale (4DEXH) CTAC. Recovery of SUVmax, SUVmean, SUVpeak, and segmented tumor volume was evaluated as RCmax, RCmean, RCpeak, and RCvol, representing percent difference relative to the static ground truth case. Paired Wilcoxon tests and Kruskal-Wallis ANOVA were used to test for significant differences. Results: For 4DPET imaging, the maximum intensity projection CTAC produced significantly more accurate recovery coefficients than all other CTAC methods (p < 0.0001 over all metrics). Over all motion waveforms, ratios of 4DMIP CTAC recovery were 0.2 +/- 5.4, -1.8 +/- 6.5, -3.2 +/- 5.0, and 3.0 +/- 5.9 for RCmax, RCpeak, RCmean, and RCvol. In comparison, recovery coefficients for phase-matched CTAC were -8.4 +/- 5.3, -10.5 +/- 6.2, -7.6 +/- 5.0, and -13.0 +/- 7.7 for RCmax, RCpeak, RCmean, and RCvol. When testing differences between phases over all CTAC methods and waveforms, end-exhale phases were significantly more accurate (p = 0.005). However, these differences were driven by the patient-specific respiratory waveforms; when testing patient and sinusoidal waveforms separately, patient waveforms were significantly different between phases (p < 0.0001) while the sinusoidal waveforms were not significantly different (p = 0.98). When considering only the subset of 4DMATCH images that corresponded to the end-exhale image phase, 4DEXH, mean and interquartile range were similar to 4DMATCH but variability was considerably reduced. Conclusions: Comparative advantages in accuracy and precision of SUV metrics and segmented volumes were demonstrated with the use of the maximum intensity projection and end-exhale CT attenuation correction. While respiratory phase-matched CTAC should in theory provide optimal corrections, image artifacts and differences in implementation of 4DCT and 4DPET sorting can degrade the benefit of this approach. These results may be useful to guide the implementation, analysis, and development of respiratory-correlated thoracic PET/CT in the radiation oncology and diagnostic settings. (c) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:110 / 120
页数:11
相关论文
共 50 条
  • [31] Compensating respiratory motion in PET image reconstruction using 4D PET/CT
    Qiao, Feng
    Pan, Tinsu
    Clark, John W., Jr.
    Mawlawi, Osama R.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2595 - 2598
  • [32] Accuracy of CT-Based attenuation correction in bone Imaging with PET/CT
    Abella, M.
    Mankof, D.
    Vaquero, J. J.
    Desco, M.
    Kinahan, P. E.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4485 - +
  • [33] Evaluation of respiratory-correlated 4D digital tomosynthesis imaging technique for image-guided radiation therapy
    Kim, Burnyoung
    Kang, Sooncheol
    Eom, Jisoo
    Lee, Seungwan
    OPTIK, 2018, 171 : 247 - 252
  • [34] Morphology supporting function: attenuation correction for SPECT/CT, PET/CT, and PET/MR imaging
    Lee, Tzu C.
    Alessio, Adam M.
    Miyaoka, Robert M.
    Kinahan, Paul E.
    QUARTERLY JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 60 (01): : 25 - 39
  • [35] 4D PET/CT gated: Technical acquisition, processing and imaging
    Palmieri, A.
    Cola, S.
    Bigi, G.
    Sangalli, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 : S214 - S214
  • [36] Four-dimensional (4D) PET/CT imaging of the thorax
    Nehmeh, SA
    Erdi, YE
    Pan, T
    Pevsner, A
    Rosenzweig, KE
    Yorke, E
    Mageras, GS
    Schoder, H
    Vernon, P
    Squire, O
    Mostafavi, H
    Larson, SM
    Humm, JL
    MEDICAL PHYSICS, 2004, 31 (12) : 3179 - 3186
  • [37] Development of respiratory-correlated 4D digital tomosynthesis imaging technique for image-guided radiation therapy
    Kim, Burnyoung
    Kang, Sooncheol
    Eom, Jisoo
    Lee, Seungwan
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [38] Qualitative and quantitative assessment of 4D PET/CT imaging based on Q.Freeze algorithm
    Minamimoto, Ryogo
    Miyata, Yoko
    Sunaoka, Fumio
    Mitsumoto, Takuya
    Morooka, Miyako
    Okasaki, Momoko
    Kubota, Kazuo
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [39] Qualitative and quantitative assessment of 4D PET/CT imaging based on Q.Freeze algorithm
    Minamimoto, Ryogo
    Miyata, Yoko
    Sunaoka, Fumio
    Mitsumoto, Takuya
    Morooka, Miyako
    Okasaki, Momoko
    Kubota, Kazuo
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [40] Respiratory synchronized CT image generation from 4D PET acquisitions
    Fayad, H.
    Lamare, F.
    Bettinardi, V.
    Roux, C.
    Visvikis, D.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4761 - +