Investigating potential physicochemical errors in polymer gel dosimeters

被引:16
|
作者
Sedaghat, Mahbod [1 ]
Bujold, Rachel [2 ]
Lepage, Martin [1 ]
机构
[1] Univ Sherbrooke, Ctr Imagerie Mol Sherbrooke, Dept Med Nucl & Radiobiol, Sherbrooke, PQ J1K 2R1, Canada
[2] CHU Sherbrooke, Serv Radio Oncol, Sherbrooke, PQ J1H 5N4, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 18期
基金
加拿大自然科学与工程研究理事会;
关键词
DOSE DELIVERY ACCURACY; ACID-BASED GEL; POLYACRYLAMIDE-GEL; RADICAL POLYMERIZATION; N-ISOPROPYLACRYLAMIDE; PHOSPHONIUM CHLORIDE; POTASSIUM PERSULFATE; RADIATION-DOSIMETRY; TEMPERATURE; ACRYLAMIDE;
D O I
10.1088/0031-9155/56/18/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Measurement errors in polymer gel dosimetry can originate either during irradiation or scanning. One concern related to the exothermic nature of polymerization reaction was that the heat released in polymer gel dosimeters during irradiation modifies their dose response. In this paper, the effect of heat released from the exothermal polymerization reaction on the dose response of a number of dosimeters was studied. In addition, we investigated whether heat-generated geometric distortion existed in newly proposed gel dosimeters that contain highly thermoresponsive polymers. Our results suggest that despite a significant internal temperature increase in some gel compositions, their dose responses are not affected when oxygen is well expelled mechanically from the gel mixture. We also report on significant pre-irradiation instability in some recently developed polymer gel dosimeters but that geometric distortions were not observed. Data obtained by a set of small calibration vials are compared to those obtained from larger phantoms, and potential physicochemical causes of deviations between them are identified.
引用
收藏
页码:6083 / 6107
页数:25
相关论文
共 50 条
  • [1] Water equivalence of polymer gel dosimeters
    Sellakumar, P.
    Jebaseelan Samuel, E. James
    Supe, Sanjay S.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2007, 76 (07) : 1108 - 1115
  • [2] Ultrasonic absorption in polymer gel dosimeters
    Mather, ML
    Collings, AF
    Bajenov, N
    Whittaker, AK
    Baldock, C
    [J]. ULTRASONICS, 2003, 41 (07) : 551 - 559
  • [3] Ultrasound evaluation of polymer gel dosimeters
    Mather, ML
    Whittaker, AK
    Baldock, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (09): : 1449 - 1458
  • [4] Response modification in polymer gel dosimeters
    Maryanski, M
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1209 - 1209
  • [5] Radiological properties of normoxic polymer gel dosimeters
    Venning, AJ
    Nitschke, KN
    Keall, PJ
    Baldock, C
    [J]. MEDICAL PHYSICS, 2005, 32 (04) : 1047 - 1053
  • [6] Polymer gel dosimeters with PVA–GA matrix
    Seyed Mohammad Mahdi Abtahi
    Hamid Sadeghi Abandansari
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 651 - 658
  • [7] The potential of polymer gel dosimeters for 3D MR-IGRT quality assurance
    Roed, Y.
    Ding, Y.
    Wen, Z.
    Wang, J.
    Pinsky, L.
    Ibbott, G.
    [J]. 9TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, 2017, 847
  • [8] Polymer Gel Dosimeters for use in Targeted Radionuclide Therapy
    Gear, J. I.
    Flux, G. D.
    Partridge, M.
    Ott, R. J.
    Charles-Edwards, E.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S207 - S207
  • [9] Polymer gel dosimeters with PVA-GA matrix
    Abtahi, Seyed Mohammad Mahdi
    Abandansari, Hamid Sadeghi
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (03) : 651 - 658
  • [10] Measurement of ultrasonic attenuation coefficient in polymer gel dosimeters
    Mather, ML
    Charles, PH
    Baldock, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (20): : N269 - N275