Traceable calibration of transfer standards for scanning probe microscopy

被引:65
|
作者
Haycocks, J [1 ]
Jackson, K [1 ]
机构
[1] Natl Phys Lab, Teddington TW11 0LW, Middx, England
关键词
atomic force microscopy; traceable calibrations; measurement uncertainty;
D O I
10.1016/j.precisioneng.2004.06.002
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A Metrological Atomic Force Microscope (MAFM) has been constructed for the traceable calibration of transfer standards for scanning probe microscopy. It uses optical interferometry to generate image scales with direct traceability to the national standard of length. Three interferometers monitor the relative displacements of the AFM tip and sample in the x, y, and z directions and the interferometer data is used directly to construct 3D images of sample surfaces. Traceable dimensional measurement of surface features may then be derived from the image data. This paper describes the MAFM instrument and presents a measurement uncertainty budget. Examples are given of measurements of pitch and step height on calibration transfer standards for scanning probe microscopy. Crown Copyright (c) 2004 Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:168 / 175
页数:8
相关论文
共 50 条
  • [1] New application for the calibration of scanning probe microscopy piezos
    Heyde, M
    Sturm, H
    Rademann, K
    SURFACE AND INTERFACE ANALYSIS, 1999, 27 (5-6) : 291 - 295
  • [2] New application for the calibration of scanning probe microscopy piezos
    Institute of Physical Chemistry, Humboldt University, Bunsenstr. 1, D-10117 Berlin, Germany
    不详
    Surf Interface Anal, 5 (291-295):
  • [3] Traceable High Impedance Calibration Standards
    Haase, Martin
    Hoffmann, Karel
    Hudlicka, Martin
    2016 87TH ARFTG MICROWAVE MEASUREMENT CONFERENCE (ARFTG), 2016,
  • [4] NEUTRON PROBE CALIBRATION BASED ON PLASTIC TRANSFER STANDARDS
    REGINATO, RJ
    NAKAYAMA, FS
    SOIL SCIENCE, 1988, 145 (05) : 381 - 384
  • [5] Transfer standard for traceable dynamic calibration of stroboscopic scanning white light interferometer
    Kassamakov, Ivan
    Tureanu, Anca
    Heikkinen, Ville
    Haeggstrom, Edward
    APPLIED OPTICS, 2017, 56 (09) : 2483 - 2488
  • [6] Modeling and calibration of micro/nano FBG temperature probe for scanning probe microscopy
    Liu, Zhenmin
    Chen, Na
    Li, Shaoying
    Liu, Yong
    Shang, Yana
    Chen, Zhenyi
    Pang, Fufei
    Wang, Tingyun
    OPTICS EXPRESS, 2023, 31 (12) : 19453 - 19462
  • [8] Wide range standard for scanning probe microscopy height calibration
    Brodowsky, HM
    Boehnke, UC
    Kremer, F
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (12): : 4198 - 4200
  • [9] Multicolor fluorescent microspheres as calibration standards for confocal laser scanning microscopy
    Zhang, YZ
    Carter, D
    APPLIED IMMUNOHISTOCHEMISTRY & MOLECULAR MORPHOLOGY, 1999, 7 (02) : 156 - 163
  • [10] Calibration of the oscillation amplitude of electrically excited scanning probe microscopy sensors
    Dagdeviren, Omur E.
    Miyahara, Yoichi
    Mascaro, Aaron
    Grutter, Peter
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (01):