摘要
A Metrological Atomic Force Microscope (MAFM) has been constructed for the traceable calibration of transfer standards for scanning probe microscopy. It uses optical interferometry to generate image scales with direct traceability to the national standard of length. Three interferometers monitor the relative displacements of the AFM tip and sample in the x, y, and z directions and the interferometer data is used directly to construct 3D images of sample surfaces. Traceable dimensional measurement of surface features may then be derived from the image data. This paper describes the MAFM instrument and presents a measurement uncertainty budget. Examples are given of measurements of pitch and step height on calibration transfer standards for scanning probe microscopy. Crown Copyright (c) 2004 Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:168 / 175
页数:8
相关论文
共 50 条
  • [31] Scanning probe microscopy
    Bian, Ke
    Gerber, Christoph
    Heinrich, Andreas J.
    Muller, Daniel J.
    Scheuring, Simon
    Jiang, Ying
    NATURE REVIEWS METHODS PRIMERS, 2021, 1 (01):
  • [32] SCANNING: Probe Microscopy
    Weeks, Brandon L.
    SCANNING, 2008, 30 (02) : 58 - 58
  • [33] SCANNING PROBE MICROSCOPY
    RAYCHAUDHURI, AK
    CURRENT SCIENCE, 1994, 67 (11): : 886 - 887
  • [34] Scanning probe microscopy
    Nature Reviews Methods Primers, 1 (1):
  • [35] New fluorescent microsphere standards for confocal laser scanning microscopy calibration.
    Zhang, YZ
    Brunner, P
    MOLECULAR BIOLOGY OF THE CELL, 1996, 7 : 912 - 912
  • [36] Scanning the Web for scanning probe microscopy
    Miller, S
    ANALYTICAL CHEMISTRY, 2003, 75 (19) : 433A - 434A
  • [37] New NIST-traceable standards for calibration and validation of DSC
    D. G. Archer
    Journal of Thermal Analysis and Calorimetry, 2006, 85 : 131 - 134
  • [38] Establishment of traceable measurement standards for acoustic emission energy calibration
    Yan, T.
    Jones, B.E.
    Key Engineering Materials, 2001, 204-205 : 331 - 340
  • [39] Establishment of traceable measurement standards for acoustic emission energy calibration
    Yan, T
    Jones, BE
    DAMAGE ASSESSMENT OF STRUCTURES, 2001, 204-2 : 331 - 339
  • [40] New NIST-traceable standards for calibration and validation of DSC
    Archer, D. G.
    JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2006, 85 (01) : 131 - 134