Comparison of Microwave Ablation and Multipolar Radiofrequency Ablation In Vivo Using Two Internally Cooled Probes

被引:77
|
作者
Fan, WeiJun [1 ,2 ]
Li, Xin [1 ,2 ]
Zhang, Liang [1 ,2 ]
Jiang, Hua [1 ,2 ]
Zhang, JianLei [1 ,2 ]
机构
[1] Sun Yat Sen Univ, Dept Med Imaging & Intervent Radiol, Ctr Canc, Guangzhou 510060, Guangdong, Peoples R China
[2] Sun Yat Sen Univ, State Key Lab Oncol S China, Guangzhou 510060, Guangdong, Peoples R China
关键词
ablation; microwave ablation; radiofrequency ablation; HEPATOCELLULAR-CARCINOMA; INITIAL-EXPERIENCE; BIPOLAR ELECTRODES; PROGNOSTIC-FACTORS; LIVER; PRINCIPLES; TUMORS; MODEL;
D O I
10.2214/AJR.11.6707
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The objective of our study was to compare the effectiveness of microwave ablation (MWA) and multipolar radiofrequency ablation (RFA) in vivo using two internally cooled probes. MATERIALS AND METHODS. MWA (n = 24) was performed by simultaneous application of double internally cooled-shaft antennae. Three power settings (60, 70, and 80 W) were used. Multipolar RFA (n = 16) was also performed by simultaneously using two internally cooled bipolar applicators (lengths: 3-cm T30 and 4-cm T40) at 60 and 80 W. Probe spacing was 2 cm. Each power setting was applied for eight ablations with a 10-minute duration for each. The cooled-shaft probes were inserted approximately 7 cm into the liver parenchyma of seven adult Wuzhishan pigs under ultrasound guidance, and ablations were performed in various segments of porcine liver. The long-axis diameter (Dl), short-axis diameter (Ds) and the ratio Ds/Dl for each ablation was measured. Temperature curves at 0, 2, and 3 cm from the middle of the two probes and the time to reach 60 degrees C at 0 cm from the parallel central line between the two probes were recorded. RESULTS. The long-axis diameter and short-axis diameter for all the power settings of MWA were significantly larger than that of both kinds of multipolar RFA (p < 0.05). The rates of temperature rise to 60 degrees C at 0 cm from the parallel central line between the two probes for all MWA power settings were significantly faster compared with RFA. CONCLUSION. MWA, by the simultaneous application of double antennae, can generate a larger ablation zone, in vivo, compared with multipolar RFA.
引用
下载
收藏
页码:W46 / W50
页数:5
相关论文
共 50 条
  • [1] Comparison of microwave ablation and multipolar radiofrequency ablation, both using a pair of internally cooled interstitial applicators: Results in ex vivo porcine livers
    Li, Xin
    Zhang, Liang
    Fan, Weijun
    Zhao, Ming
    Wang, Ligang
    Tang, Tian
    Jiang, Hua
    Zhang, Jianlei
    Liu, Yanfeng
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2011, 27 (03) : 240 - 248
  • [2] A comparison of microwave ablation and bipolar radiofrequency ablation both with an internally cooled probe: Results in ex vivo and in vivo porcine livers
    Yu, Jie
    Liang, Ping
    Yu, Xiaoling
    Liu, Fangyi
    Chen, Lei
    Wang, Yang
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (01) : 124 - 130
  • [3] Internally gas-cooled radiofrequency applicators as an alternative to conventional radiofrequency and microwave ablation devices: An in vivo comparison
    Rempp, Hansjoeg
    Voigtlaender, Matthias
    Schenk, Martin
    Enderle, Markus D.
    Scharpf, Marcus
    Greiner, Tim O.
    Neugebauer, Alexander
    Hoffmann, Ruediger
    Claussen, Claus D.
    Clasen, Stephan
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (08) : E350 - E355
  • [4] Radiofrequency renal ablation: In vivo comparison of internally cooled, multitined expandable and internally cooled perfusion electrodes
    Lee, JM
    Han, JK
    Chang, JM
    Chung, SY
    Son, KR
    Kim, SH
    Lee, JY
    Choi, BI
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (03) : 549 - 556
  • [5] Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling
    Clasen, S
    Schmidt, D
    Boss, A
    Dietz, K
    Kröber, SM
    Claussen, CD
    Pereira, PL
    RADIOLOGY, 2006, 238 (03) : 881 - 890
  • [6] Radiofrequency Ablation Zones in Ex Vivo Bovine and In Vivo Porcine Livers: Comparison of the Use of Internally Cooled Electrodes and Internally Cooled Wet Electrodes
    Cha, Jihoon
    Choi, Dongil
    Lee, Min Woo
    Rhim, Hyunchul
    Kim, Young-sun
    Lim, Hyo K.
    Yoon, Jung Hwan
    Park, Cheol Keun
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 32 (06) : 1235 - 1240
  • [7] Radiofrequency Ablation Zones in Ex Vivo Bovine and In Vivo Porcine Livers: Comparison of the Use of Internally Cooled Electrodes and Internally Cooled Wet Electrodes
    Jihoon Cha
    Dongil Choi
    Min Woo Lee
    Hyunchul Rhim
    Young-sun Kim
    Hyo K. Lim
    Jung Hwan Yoon
    Cheol Keun Park
    CardioVascular and Interventional Radiology, 2009, 32 : 1235 - 1240
  • [8] Radiofrequency ablation in pig lungs:: in vivo comparison of internally cooled, perfusion and multitined expandable electrodes
    Lee, J. M.
    Han, J. K.
    Chang, J. M.
    Chung, S. Y.
    Kim, S. H.
    Lee, J. Y.
    Choi, B. I.
    BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (943): : 562 - 571
  • [9] Multipolar radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver: Correlation between volume of coagulation and amount of applied energy
    Clasen, Stephan
    Rempp, Hansjoerg
    Schmidt, Diethard
    Schraml, Christina
    Hoffmann, Ruediger
    Claussen, Claus D.
    Pereira, Philippe L.
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (01) : 111 - 113
  • [10] Hepatic radiofrequency ablation with internally cooled probes: Effect of coolant temperature on lesion size
    Haemmerich, D
    Chachati, L
    Wright, AS
    Mahvi, DM
    Lee, FT
    Webster, JG
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2003, 50 (04) : 493 - 500