Runaway pacemaker caused by a stuck accelerometer

被引:0
|
作者
Nakazato, Y
Nakata, Y
Yasuda, M
Nakazato, K
Tanaka, M
Sumiyoshi, M
Sakurai, H
Yamaguchi, H
机构
[1] Juntendo Univ, Sch Med, Dept Internal Med, Div Cardiol, Tokyo 1138421, Japan
[2] Juntendo Univ, Izu Nagaoka Hosp, Shizuoka, Japan
来源
关键词
runaway pacemaker; accelerometer; malfunction;
D O I
10.1111/j.1540-8159.1999.tb00487.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a case of runaway pacemaker with a ventricular pacing rate of 190 beats/min. The runaway occurred when the accelerometer became stuck due to the magnet-application during VVIR pacing. Runaways in modern pacemakers ore particularly rare, but they do still occur. The best solution for this phenomenon is generator replacement.
引用
收藏
页码:538 / 539
页数:2