Runaway pacemaker caused by a stuck accelerometer

被引:0
|
作者
Nakazato, Y
Nakata, Y
Yasuda, M
Nakazato, K
Tanaka, M
Sumiyoshi, M
Sakurai, H
Yamaguchi, H
机构
[1] Juntendo Univ, Sch Med, Dept Internal Med, Div Cardiol, Tokyo 1138421, Japan
[2] Juntendo Univ, Izu Nagaoka Hosp, Shizuoka, Japan
来源
关键词
runaway pacemaker; accelerometer; malfunction;
D O I
10.1111/j.1540-8159.1999.tb00487.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a case of runaway pacemaker with a ventricular pacing rate of 190 beats/min. The runaway occurred when the accelerometer became stuck due to the magnet-application during VVIR pacing. Runaways in modern pacemakers ore particularly rare, but they do still occur. The best solution for this phenomenon is generator replacement.
引用
收藏
页码:538 / 539
页数:2
相关论文
共 50 条
  • [31] RUNAWAY PACEMAKER IN 7 PATIENTS - PERSISTING PROBLEM
    NASRALLAH, A
    HALL, RJ
    GARCIA, E
    KYGER, ER
    HALLMAN, GL
    COOLEY, DA
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1975, 69 (03): : 365 - 368
  • [32] Runaway pacemaker syndrome and intermittent nonoutput as manifestations of end of life of a VVI pacemaker
    Böhm, A
    Hajdu, L
    Pintér, A
    Duray, G
    Préda, I
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (12): : 2143 - 2144
  • [33] RUNAWAY PACEMAKER - A STILL EXISTING COMPLICATION AND THERAPEUTIC GUIDELINES
    MICKLEY, H
    ANDERSEN, C
    NIELSEN, LH
    CLINICAL CARDIOLOGY, 1989, 12 (07) : 412 - 414
  • [34] VENTRICULAR ECTOPIC RHYTHMS DUE TO RAPID RUNAWAY PACEMAKER
    HASIN, Y
    LEWIS, BS
    APPLEBAUM, A
    GOTSMAN, MS
    CHEST, 1978, 73 (03) : 411 - 413
  • [35] Trans-innominate vein extraction of stuck pacemaker leads
    Gupta, Anish
    Devagourou, Velayoudam
    Choudhury, Minati
    Saurabh, Gaind
    INDIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2019, 35 (01) : 78 - 80
  • [36] Trans-innominate vein extraction of stuck pacemaker leads
    Anish Gupta
    Velayoudam Devagourou
    Minati Choudhury
    Gaind Saurabh
    Indian Journal of Thoracic and Cardiovascular Surgery, 2019, 35 : 78 - 80
  • [37] RUNAWAY ATRIOVENTRICULAR SEQUENTIAL PACEMAKER AFTER RADIATION-THERAPY
    LEE, RW
    HUANG, SK
    MECHLING, E
    BAZGAN, I
    AMERICAN JOURNAL OF MEDICINE, 1986, 81 (05): : 883 - 886
  • [38] Runaway pacemaker due to software-based programming error
    Levert, JV
    Hoorntje, JCA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (12): : 1689 - 1690
  • [39] RUNAWAY PACEMAKER RHYTHM AFTER 36 HOURS OF ELECTRICAL INACTIVITY
    HARPER, R
    PETER, T
    HUNT, D
    SMITH, D
    SLOMAN, G
    BRITISH HEART JOURNAL, 1974, 36 (06): : 610 - 612
  • [40] STEPWISE ANALYSIS OF THE CALIBRATION PROCEDURE OF AN ACCELEROMETER-BASED PACEMAKER
    MAHAUX, VA
    VERBOVEN, YJ
    WALEFFE, A
    KULBERTUS, H
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 1955 - 1959