A portable micro-X-ray fluorescence spectrometer with polycapillary optics and vacuum chamber for archaeometric and other applications

被引:37
|
作者
Buzanich, G. [1 ]
Wobrauschek, P. [1 ]
Streli, C. [1 ]
Markowicz, A. [2 ,3 ]
Wegrzynek, D. [2 ,3 ]
Chinea-Cano, E. [2 ]
Bamford, S. [2 ]
机构
[1] TU Wien Atominst, A-1020 Vienna, Austria
[2] IAEA, Agcy Labs Seibersdorf, A-1400 Vienna, Austria
[3] AGH Univ Sci & Technol, Fac Phys & Appl Comp Sci, Dept Radiometry, PL-30059 Krakow, Poland
关键词
portable micro-X-ray spectrometer; vacuum chamber; polycapillary; in situ analysis;
D O I
10.1016/j.sab.2007.08.003
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A portable focused-beam XRF spectrometer was designed, constructed, and manufactured. The spectrometer allows to detect and perform analysis of chemical elements from Na upwards. The system is equipped with a compact vacuum chamber to reduce absorption of both the excitation and the fluorescence radiation in air. A low power Pd-anode tube operated up to 50 kV and 1 mA with a point focus of 400 mu m is used as excitation source. A polycapillary lens with a spot size of about 160 pm, or a collimator with a 1 min inner diameter can be used alternatively for either focusing or collimating the primary beam. The fluorescence radiation is collected by an Si drift detector with an active area of 10 min 2 and equipped with an 8 mu m Be entrance window. A compact vacuum chamber was designed to house the X-ray beam optics and the detector snout. The chamber is attached to the X-ray tube and can be pumped down to 0.1 mbar. A Kapton(TM) window of 7.5 mu m thickness allows to locate the investigated spot at about 1-2 mm distance outside of the chamber, thus minimizing absorption losses in the excitation and X-ray fluorescence radiation paths. Two lasers pointers are mounted inside the chamber. The laser beams cross at a point outside the chamber in front of the entrance window and coincide with the focal spot of the polycapillary. This paper reports some preliminary results obtained from an in situ analysis of bronze samples as well as a comparison of these data with those given by other laboratory spectrometers and the reference values provided by the Italian bronze foundry Venturi Arte Bologna, Italy. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:1252 / 1256
页数:5
相关论文
共 50 条
  • [1] A New Type of Portable Micro-X-Ray Fluorescence Spectrometer with Polycapillary Optics
    Duan Ze-ming
    Liu Jun
    Jiang Qi-li
    Pan Qiu-li
    Li Rong-wu
    Cheng Lin
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39 (01) : 303 - 309
  • [2] Development of Portable Micro-X-ray Fluorescence Spectrometer
    便携式微束X射线荧光谱仪的研发
    [J]. Cheng, Lin (chenglin@bnu.edu.cn), 2018, Atomic Energy Press (52):
  • [3] A new type of micro-X-ray diffractometer focused by polycapillary optics
    Jiang Qi-Li
    Duan Ze-Ming
    Shuai Qi-Lin
    Li Rong-Wu
    Pan Qiu-Li
    Cheng Lin
    [J]. ACTA PHYSICA SINICA, 2019, 68 (24)
  • [4] Development of a micro-X-ray fluorescence system based on polycapillary X-ray optics for non-destructive analysis of archaeological objects
    Cheng, Lin
    Ding, Xunliang
    Liu, Zhiguo
    Pan, Qiuli
    Chu, Xuelian
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2007, 62 (08) : 817 - 823
  • [5] Investigation of polycapillary half lenses for quantitative confocal micro-X-ray fluorescence analysis
    Iro, Michael
    Ingerle, Dieter
    Radtke, Martin
    Buzanich, Ana Guilherme
    Kregsamer, Peter
    Streli, Christina
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 2022, 29 : 1376 - 1384
  • [6] Note: Portable total reflection X-ray fluorescence spectrometer with small vacuum chamber
    Kunimura, Shinsuke
    Kudo, Shunpei
    Nagai, Hiroki
    Nakajima, Yoshihide
    Ohmori, Hitoshi
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (04):
  • [7] Confocal micro-x-ray fluorescence spectrometer for light element analysis
    Smolek, S.
    Pemmer, B.
    Foelser, M.
    Streli, C.
    Wobrauschek, P.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (08):
  • [8] A monochromatic confocal micro-x-ray fluorescence (μXRF) spectrometer for the lab
    Ingerle, D.
    Swies, J.
    Iro, M.
    Wobrauschek, P.
    Streli, C.
    Hradil, K.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (12):
  • [9] A newly micro-X-ray fluorescence spectrometer and applications for non-destructive analysis of archaeological objects
    Cheng Lin
    Ding Xun-Liang
    Liu Zhi-Guo
    Pan Qiu-Li
    Chu Xue-Lian
    Feng Song-Lin
    [J]. ACTA PHYSICA SINICA, 2007, 56 (12) : 6894 - 6898
  • [10] Monolithic polycapillary focusing optics and their applications in microbeam x-ray fluorescence
    Gao, N
    Ponomarev, IY
    Xiao, QF
    Gibson, WM
    Carpenter, DA
    [J]. APPLIED PHYSICS LETTERS, 1996, 69 (11) : 1529 - 1531