Evaluation of radiation dose to pediatric models from whole body PET/CT imaging

被引:6
|
作者
Mohammadi, Najmeh [1 ]
Akhlaghi, Parisa [2 ]
机构
[1] Sahand Univ Technol, Fac Sci, Phys Dept, Tabriz 513351996, Iran
[2] Tabriz Univ Med Sci, Fac Med, Dept Med Phys, Tabriz, Iran
来源
关键词
dosimetry; Monte Carlo; pediatrics; PET; CT; voxel phantoms; F-18-FDG PET; UF SERIES; CHILDREN; CT; GUIDELINES; DOSIMETRY; EXPOSURE; PHANTOMS;
D O I
10.1002/acm2.13545
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Positron emission tomography (PET)/computed tomography (CT) is a well-known modality for the diagnosis of various diseases in children and adult patients. On the other hand, younger patients are more radiosensitive than adults, so there are concerns about the level of ionizing radiation exposure in pediatric whole body PET/CT imaging. In this regard, comprehensive specific radiation dosimetry for whole body PET/CT imaging is highly desired for different ages, sizes, and shapes. Therefore, in this study, organ absorbed doses were evaluated for pediatric voxel models from 4 to 14 years old and compared with those of ICRP phantoms. Monte Carlo calculation was performed to evaluate S-value, absorbed dose, and effective dose from F-18-FDG radiotracers and whole body CT scan for different computational models, including 4- to 14-year-old phantoms. The results showed that the S-value and, therefore, absorbed dose of F-18-FDG strongly depended on the phantom anatomy. These variations were justified by the distance between source and target organs. Moreover, on average, the absorbed doses from whole body CT scans were 13.5 times lower than those from F-18-FDG for all organs. According to the results, various anatomies and ages should be considered for accurate dose evaluation. These data can be used for specific risk assessment of the pediatric population in clinical nuclear imaging.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Evaluation of radiation dose and image quality of CT scan for whole-body pediatric PET/CT: A phantom study
    Yang, Ching-Ching
    Liu, Shu-Hsin
    Mok, Greta S. P.
    Wu, Tung-Hsin
    MEDICAL PHYSICS, 2014, 41 (09)
  • [2] Dose optimization for pediatric FDG whole body PET/CT
    van Gent, M.
    Hamming, V. C.
    Schaar, J. E.
    Glaudemans, A. W. J. M.
    Willemsen, A. T. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S209 - S209
  • [3] CT Radiation Dose Optimization in Whole-Body PET/CT Examination
    Tonkopi, E.
    Ross, A.
    MacDonald, A.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [4] A retrospective evaluation of radiation dose associated with low dose FDG protocols in whole-body PET/CT
    Willowson, Kathy P.
    Bailey, Elizabeth A.
    Bailey, Dale L.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 49 - 53
  • [5] A RETROSPECTIVE EVALUATION OF RADIATION DOSE ASSOCIATED WITH LOW DOSE FDG PROTOCOLS IN WHOLE-BODY PET/CT
    Willowson, K.
    Bailey, E.
    Bailey, D.
    INTERNAL MEDICINE JOURNAL, 2012, 42 : 4 - 4
  • [6] A retrospective evaluation of radiation dose associated with low dose FDG protocols in whole-body PET/CT
    Kathy P. Willowson
    Elizabeth A. Bailey
    Dale L. Bailey
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 49 - 53
  • [7] Diagnostic Reference Levels of CT Radiation Dose in Whole-Body PET/CT
    Jallow, Ngoneh
    Christian, Paul
    Sunderland, John
    Graham, Michael
    Hoffman, John M.
    Nye, Jonathon A.
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (02) : 238 - 241
  • [8] Patient-specific radiation dose for Chinese pediatric patients undergoing whole-body PET/CT examinations
    Jia, Haoran
    Xue, Mengjia
    Li, Xianru
    Zhuang, Mingzan
    Xie, Tianwu
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (12):
  • [9] Whole-body FDG-PET/CT imaging with CT-dose adaptation
    Beyer, T
    Stattaus, J
    Rosenbaum, S
    Stergar, H
    Kuehl, H
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S407 - S408
  • [10] Clinical evaluation of CT radiation dose in whole-body 18F-FDG PET/CT in relation to scout imaging direction and arm position
    Inoue, Yusuke
    Nagahara, Kazunori
    Inoki, Yuri
    Hara, Toshimasa
    Miyatake, Hiroki
    ANNALS OF NUCLEAR MEDICINE, 2019, 33 (03) : 169 - 176