摘要
Purpose: This study aims to propose a physics-based method of reducing beam-hardening artifacts induced by high-attenuation materials such as metal stents or other metallic implants. Methods: The proposed approach consists of deriving a sinogram inconsistency formula representing the energy dependence of the attenuation coefficient of high-attenuation materials. This inconsistency formula more accurately represents the inconsistencies of the sinogram than that of a previously reported formula(6) (called the MAC-BC method). This is achieved by considering the properties of the high-attenuation materials, which include the materials' shapes and locations and their effects on the incident X-ray spectrum, including their attenuation coefficients. Results: Numerical simulation and phantom experiment demonstrate that the modeling error of MAC-BC method are nearly completely removed by means of the proposed method. Conclusion: The proposed method reduces beam-hardening artifacts arising from high-attenuation materials by relaxing the assumptions of the MAC-BC method. In doing so, it outperforms the original MAC-BC method. Further research is required to address other potential sources of metal artifacts, such as photon starvation, scattering, and noise. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:E147 / E152
页数:6
相关论文
共 50 条
  • [1] Metal Artifact Reduction Based on Sinogram Correction in CT
    Jeong, Kye Young
    Ra, Jong Beom
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3480 - 3483
  • [2] Beam hardening artifact correction in CT based on re-projection
    Wu, Zhi-Hong
    Cong, Peng
    Liu, Xi-Ming
    [J]. Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2015, 49 (05): : 935 - 938
  • [3] Beam-hardening correction method based on original sinogram for X-CT
    Fu, J
    Lu, HN
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 556 (01): : 379 - 385
  • [4] A CT metal artifact reduction algorithm based on sinogram surgery
    Jeon, Soomin
    Lee, Chang-Ock
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2018, 26 (03) : 413 - 434
  • [5] Metal Artifact Reduction in CT-Based Attenuation Correction of PET using Sobolev Sinogram Restoration
    Mehranian, A.
    Ay, M. R.
    Rahmim, A.
    Zaidi, H.
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2936 - 2942
  • [6] Fully Convolutional Networks Based Sinogram Correction for Metal Artifact Reduction
    Zhu, Linlin
    Han, Yu
    Li, Ziheng
    Xi, Xiaoqi
    Li, Lei
    Yan, Bin
    Zhu, Mingwan
    [J]. AI IN OPTICS AND PHOTONICS (AOPC 2019), 2019, 11342
  • [7] CT sinogram-consistency learning for metal-induced beam hardening correction
    Park, Hyoung Suk
    Lee, Sung Min
    Kim, Hwa Pyung
    Seo, Jin Keun
    Chung, Yong Eun
    [J]. MEDICAL PHYSICS, 2018, 45 (12) : 5376 - 5384
  • [8] A Model-Based Image Reconstruction Algorithm With Simultaneous Beam Hardening Correction for X-Ray CT
    Jin, Pengchong
    Bouman, Charles A.
    Sauer, Ken D.
    [J]. IEEE TRANSACTIONS ON COMPUTATIONAL IMAGING, 2015, 1 (03): : 200 - 216
  • [9] Artifact reduction in truncated CT using Sinogram completion
    Chityala, R
    Hoffmann, KR
    Rudin, S
    Bednarek, DR
    [J]. Medical Imaging 2005: Image Processing, Pt 1-3, 2005, 5747 : 2110 - 2117
  • [10] Metal artifact reduction in CT-based attenuation correction of PET data using the virtual sinogram concept
    Abdoli, Mehrsima
    Ay, Mohammad Reza
    Ahmadian, Alireza
    Zaidi, Habib
    [J]. 3rd International Conference on Bioinformatics and Biomedical Engineering, iCBBE 2009, 2009,