Artifact reduction in truncated CT using Sinogram completion

被引:26
|
作者
Chityala, R [1 ]
Hoffmann, KR [1 ]
Rudin, S [1 ]
Bednarek, DR [1 ]
机构
[1] SUNY Buffalo, Dept Phys, Buffalo, NY 14214 USA
关键词
Sinogram completion; truncated reconstruction; Sinogram; extrapolation; artifact reduction; dose reduction; reduced FOV;
D O I
10.1117/12.595450
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
Truncation of projection data in CT produces significant artifacts in the reconstruction process due to non-locality of the Radon transform. In this paper, we present a method for reducing these truncation artifacts by estimating features that lie outside the region of interest (ROI) and using these features to complete the truncated sinogram. Projection images of an object are obtained. A sinogram is obtained by stacking profile data from all projection angles. A simulated truncated sinogram is generated by setting pixel values outside an ROI to zero. The truncated sinogram is then transformed into a (radius, phase) image, with pixel values in what we term as the Polar representation (PR) image corresponding to the minimum value along sine curves given by x = r*cos(projection angle + phase). The PR image contains data for radii greater than the ROI radius. Pixel values outside the ROI in the completed sinogram are determined as follows. For each pixel in the PR image, a sine curve is generated in the completed sinogram image outside the ROI, having the same pixel value as that of the PR image for that radius and phase. Successive sine curves are laid and the values of each are summed. The intensity outside is then equalized to the intensity inside the ROL The completed sinogram is then reconstructed, to obtain completed reconstruction. The percentage error in the difference image between the full FOV reconstruction and the corresponding completed reconstruction and the extrapolated-average reconstruction are 1.1 % and 3.3 % respectively. This indicates that the completed reconstruction is closer to full FOV reconstruction. Thus, the sinogram completion can be used to improve reconstructions from truncated data.
引用
收藏
页码:2110 / 2117
页数:8
相关论文
共 50 条
  • [1] Physics-informed sinogram completion for metal artifact reduction in CT imaging
    Zhu, Manman
    Zhu, Qisen
    Song, Yuyan
    Guo, Yi
    Zeng, Dong
    Bian, Zhaoying
    Wang, Yongbo
    Ma, Jianhua
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (06):
  • [2] Deep Sinogram Completion With Image Prior for Metal Artifact Reduction in CT Images
    Yu, Lequan
    Zhang, Zhicheng
    Li, Xiaomeng
    Xing, Lei
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (01) : 228 - 238
  • [3] Metal Artifact Reduction in CT Using Unsupervised Sinogram Manifold Learning
    Peng, Junbo
    Chang, Chih-Wei
    Xie, Huiqiao
    Fan, Mingdong
    Wang, Tonghe
    Roper, Justin
    Qiu, Richard L. J.
    Tang, Xiangyang
    Yang, Xiaofeng
    [J]. MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [4] Metal Artifact Reduction Based on Sinogram Correction in CT
    Jeong, Kye Young
    Ra, Jong Beom
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3480 - 3483
  • [5] Metal Artifact Reduction in CT images by Sinogram TV inpainting
    Duan, Xinhui
    Zhang, Li
    Xiao, Yongshun
    Cheng, Jianping
    Chen, Zhiqiang
    Xing, Yuxiang
    [J]. 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3450 - +
  • [6] A CT metal artifact reduction algorithm based on sinogram surgery
    Jeon, Soomin
    Lee, Chang-Ock
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2018, 26 (03) : 413 - 434
  • [7] Metal Artifact Reduction in CT-Based Attenuation Correction of PET using Sobolev Sinogram Restoration
    Mehranian, A.
    Ay, M. R.
    Rahmim, A.
    Zaidi, H.
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2936 - 2942
  • [8] Metal artifact reduction in spiral fan-beam CT using a new sinogram segmentation scheme
    Yazdi, Mehran
    Mansourian, Zohre
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2017, 25 (05) : 737 - 749
  • [9] Metal artifact reduction in CT-based attenuation correction of PET data using the virtual sinogram concept
    Abdoli, Mehrsima
    Ay, Mohammad Reza
    Ahmadian, Alireza
    Zaidi, Habib
    [J]. 3rd International Conference on Bioinformatics and Biomedical Engineering, iCBBE 2009, 2009,
  • [10] Metal artifact reduction and image quality evaluation of lumbar spine CT images using metal sinogram segmentation
    Kaewlek, Titipong
    Koolpiruck, Diew
    Thongvigitmanee, Saowapak
    Mongkolsuk, Manus
    Thammakittiphan, Sastrawut
    Tritrakarn, Siri-on
    Chiewvit, Pipat
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2015, 23 (06) : 649 - 666