Microwave-assisted sample preparation of Hungarian raw propolis in quartz vessels and element analysis by ICP-OES and ICP-MS for geographical identification

被引:6
|
作者
Soos, Aron [1 ]
Bodi, Eva [1 ]
Varallyay, Szilvia [1 ]
Molnar, Szabolcs [2 ]
Kovacs, Bela [1 ]
机构
[1] Univ Debrecen, Inst Food Sci, Boszormenyi Str 138, H-4032 Debrecen, Hungary
[2] Eszterhazy Karoly Univ, Food & Wine Res Ctr, Leanyka Str 6, H-3300 Eger, Hungary
关键词
Microwave-assisted digestion; Sample preparation; Element analysis; Propolis; Geographical identification; SINGLE REACTION CHAMBER; DIFFERENT REGIONS; BOTANICAL ORIGIN; TOXIC ELEMENTS; WET DIGESTION; GREEN; CHEMISTRY; HONEYBEES; PRODUCTS;
D O I
10.1016/j.talanta.2021.122613
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
One of the aims of this study was to improve the sample throughput of a microwave-assisted closed vessel digestion system by using small quartz vials in polytetrafluomethylene (PTFE) vessels for the sample preparation of raw propolis samples in small amounts. The digested samples were measured by inductively coupled plasma optical emission spectrometry (ICP-OES) and inductively coupled plasma mass spectrometry (ICP-MS) analyzing 36 elements. Limit of detection was low enough to measure all the elements, with the exception of La, in all raw pmpolis samples. There were no cross-contamination between the quartz vials, therefore independent samples can be prepared in the same PTFE vessel. Accuracy of the method was checked by spike recoveries and by analyzing BCR 189 wholemeal flour and two plant samples from a collaborative study. The means of RSDs were 5.3%-14.4% in the case of measured elements. The sample throughput was increased by three times using quartz vials in PFTE vessels besides matching with the requirements of green chemistry. Another goal was the characterization of the element content and thereby geographical identification of Hungarian raw propolis. In total, 252 samples were analyzed and their statistical characteristics were described. We cannot find globally such results of pmpolis element content, which is representing one country and with such a number of elements and samples. All the elements have positive skew and positive kurtosis. Concentration range is above two orders of magnitude in the case of Ba, Zn, V, Cr, Ni, Cd and Eu elements. The decimal logarithm of element concentrations was used for geographical identification of raw pmpolis samples originating from seven regions of Hungary by linear discriminant analysis (LDA). Grouping of the samples of the Northern Great Plain was the most effective with 96.3% and 77.8% based on the original method and the cross validation, respectively. The same indicators for all the groups are 76.6% and 61.5%.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Microwave-assisted Digestion of Environmental Soil Sample and Measurement of Plutonium by ICP-MS
    Xu H.
    Ma T.
    Li D.
    Zhan J.
    Wang Y.
    Wang B.
    Wang Y.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2019, 53 (04): : 603 - 610
  • [22] Microwave-assisted digestion method using diluted nitric acid and hydrogen peroxide for the determination of major and minor elements in milk samples by ICP-OES and ICP-MS
    Lee, Jian
    Park, Ye-Seul
    Lee, Hwa-Jeong
    Koo, Yong Eui
    FOOD CHEMISTRY, 2022, 373
  • [23] Fast and green microwave-assisted digestion with diluted nitric acid and hydrogen peroxide and subsequent determination of elemental composition in brown and white rice by ICP-MS and ICP-OES
    Lee, Jian
    Park, Ye-Seul
    Lee, Do Yup
    LWT-FOOD SCIENCE AND TECHNOLOGY, 2023, 173
  • [24] Mercury speciation in whole blood by gas chromatography coupled to ICP-MS with a fast microwave-assisted sample preparation procedure
    Rodrigues, Jairo L.
    Alvarez, Carolina Rodriguez
    Farinas, Nuria Rodriguez
    Berzas Nevado, Juan J.
    Barbosa, Fernando, Jr.
    Rodriguez Martin-Doimeadios, Rosa C.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (02) : 436 - 442
  • [25] Increasing throughput and lowering detection levels in ICP-MS analysis sing a microwave digestion quartz rotor for sample preparation
    Milestone Inc., 25 Controls Dr., Shelton, CT 06484, United States
    Am Lab, 2007, 14 (30-31+10):
  • [26] Increasing throughput and lowering detection levels in ICP-MS analysis using a microwave digestion quartz rotor for sample preparation
    Michel, Tim
    Jacobs, Victoria
    Loechner, Merrill
    AMERICAN LABORATORY, 2007, 39 (14) : 30 - +
  • [27] Investigation of Changes in Element Contents of Different Citrus Fruit Peel and Flesh Parts by Microwave-Assisted Digestion and ICP-OES
    Ahmed, Isam A. Mohamed
    Al Juhaimi, Fahad
    Ozcan, Mehmet Musa
    Uslu, Nurhan
    Karrar, Emad
    APPLIED FRUIT SCIENCE, 2025, 67 (02)
  • [28] Determination of platinum and rhodium in dust and plant samples using microwave-assisted sample digestion and ICP-MS
    Niemelä, M
    Perämäki, P
    Piispanen, J
    Poikolainen, J
    ANALYTICA CHIMICA ACTA, 2004, 521 (02) : 137 - 142
  • [29] Green microwave-assisted wet digestion method of carbohydrate-rich foods with hydrogen peroxide using single reaction chamber and further elemental determination using ICP-OES and ICP-MS
    Muller, Edson I.
    Muller, Cristiano C.
    Souza, Juliana P.
    Muller, Aline L. H.
    Enders, Michele S. P.
    Doneda, Morgana
    Frohlich, Angelica C.
    Iop, Gabrielle D.
    Anschau, Kellen F.
    MICROCHEMICAL JOURNAL, 2017, 134 : 257 - 261
  • [30] Microwave-assisted wet digestion with H2O2 at high temperature and pressure using single reaction chamber for elemental determination in milk powder by ICP-OES and ICP-MS
    Muller, Edson I.
    Souza, Juliana P.
    Muller, Cristiano C.
    Muller, Aline L. H.
    Mello, Paola A.
    Bizzi, Cezar A.
    TALANTA, 2016, 156 : 232 - 238