A novel methodology for the optimization of transmission and dosimetric leaf gap parameters

被引:3
|
作者
DiCostanzo, Dominic J. [1 ]
Ayan, Ahmet S. [1 ]
机构
[1] Ohio State Univ, Columbus, OH 43210 USA
来源
关键词
commissioning; dosimetric leaf gap; multileaf collimators; quality assurance; treatment planning systems; IMRT; TOOL;
D O I
10.1002/acm2.13565
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Optimization of dosimetric leaf gap (DLG) and transmission is commonly performed through a manual trial and error process, which can lead to sub-optimal values. The purpose of this work is to create an alternative automated optimization process that provides the optimal DLG and transmission pair for use in a clinical setting. Methods Utilizing the treatment planning system application programming interface, a phase space of clinically viable DLG and transmission pairs was generated. The phase space contained 51,051 dose planes for DLGs between 0.0 and 2.5 mm and transmission values between 0.01% and 2.5%. Thirteen plans were measured for multiple multileaf collimator types and nominal beam energies. The optimization minimized the mean gamma-index and maximized the gamma-index pass rate. The optimized values were validated using five plans excluded from the optimization. Results Of the nominal beam energies and multileaf collimator system (MLC)-type combinations tested, 6/7 showed an increase in gamma-index pass rate and a decrease in mean gamma-index signifying better agreement between measurement and calculation. When comparing the optimized DLG and transmission values to the clinically implemented values identified via an iterative method, 5/7 energy, and MLC type combinations showed no statistically significant changes. In addition, the optimized values were benchmarked against three Task Group 119 plans with published gamma-index pass rates, which had been held out of the optimization. For those plans, the optimized DLG and transmission values provided the same or better gamma-index pass rates. Conclusion We presented a novel and viable automated alternative to current approaches of selecting the DLG and transmission parameters. This method will reduce the time required to determine the clinically acceptable DLG and transmission parameters and ensure optimality for the plans included in the optimization.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Leaf Gap Leakage and Dosimetric Accuracy for SBRT Treatment on the TrueBeam STx
    Kielar, K. N.
    Mok, E.
    Wang, L.
    Hsu, A.
    Maxim, P.
    Luxton, G.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S852 - S852
  • [22] Theory and Clinical Implications of the Constant Dosimetric Leaf Gap (DLG) Approximation
    Kumaraswamy, L.
    Bailey, D.
    Xu, Z.
    Schmitt, J.
    Podgorsak, M.
    MEDICAL PHYSICS, 2016, 43 (06) : 3630 - 3630
  • [23] Generalized Beam Modulation Based Dosimetric Leaf Gap Determination Scheme
    Chen, T.
    Yue, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 2917 - 2917
  • [24] Effect of Dose Rate and Leakage Correction for Dosimetric Leaf Gap Measurement
    Feng, W.
    Chu, A.
    Chi, Y.
    Hu, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 372 - 372
  • [25] Sensitivity of Patient Specific IMRT QA to Errors in the Dosimetric Leaf Gap
    Bollinger, D.
    Sun, B.
    MEDICAL PHYSICS, 2017, 44 (06) : 2891 - 2891
  • [26] Spatial variation of dosimetric leaf gap and its impact on dose delivery
    Kumaraswamy, Lalith K.
    Schmitt, Jonathan D.
    Bailey, Daniel W.
    Xu, Zheng Zheng
    Podgorsak, Matthew B.
    MEDICAL PHYSICS, 2014, 41 (11)
  • [27] Ideal Dosimetric Leaf Gap as a Function of Equivalent Square Field Size
    Seekamp, J.
    Pearson, D.
    MEDICAL PHYSICS, 2022, 49 (06) : E511 - E512
  • [28] Comparison of Monte Carlo Simulation and Measurement for Determination of Dosimetric Leaf Gap
    Popple, R.
    Sawkey, D.
    Yuan, Y.
    Sullivan, R.
    MEDICAL PHYSICS, 2019, 46 (06) : E262 - E262
  • [29] A Simple Method for Determining Dosimetric Leaf Gap for Rounded Leaf-End Multileaf Collimator Systems
    Shiau, A.
    Lin, C.
    Ji, J.
    Liang, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E302 - E303
  • [30] Impact of the dosimetric leaf gap on the AAA portal dose image prediction algorithm
    Calvo-Ortega, J.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1444 - S1446