A novel methodology for the optimization of transmission and dosimetric leaf gap parameters

被引:3
|
作者
DiCostanzo, Dominic J. [1 ]
Ayan, Ahmet S. [1 ]
机构
[1] Ohio State Univ, Columbus, OH 43210 USA
来源
关键词
commissioning; dosimetric leaf gap; multileaf collimators; quality assurance; treatment planning systems; IMRT; TOOL;
D O I
10.1002/acm2.13565
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Optimization of dosimetric leaf gap (DLG) and transmission is commonly performed through a manual trial and error process, which can lead to sub-optimal values. The purpose of this work is to create an alternative automated optimization process that provides the optimal DLG and transmission pair for use in a clinical setting. Methods Utilizing the treatment planning system application programming interface, a phase space of clinically viable DLG and transmission pairs was generated. The phase space contained 51,051 dose planes for DLGs between 0.0 and 2.5 mm and transmission values between 0.01% and 2.5%. Thirteen plans were measured for multiple multileaf collimator types and nominal beam energies. The optimization minimized the mean gamma-index and maximized the gamma-index pass rate. The optimized values were validated using five plans excluded from the optimization. Results Of the nominal beam energies and multileaf collimator system (MLC)-type combinations tested, 6/7 showed an increase in gamma-index pass rate and a decrease in mean gamma-index signifying better agreement between measurement and calculation. When comparing the optimized DLG and transmission values to the clinically implemented values identified via an iterative method, 5/7 energy, and MLC type combinations showed no statistically significant changes. In addition, the optimized values were benchmarked against three Task Group 119 plans with published gamma-index pass rates, which had been held out of the optimization. For those plans, the optimized DLG and transmission values provided the same or better gamma-index pass rates. Conclusion We presented a novel and viable automated alternative to current approaches of selecting the DLG and transmission parameters. This method will reduce the time required to determine the clinically acceptable DLG and transmission parameters and ensure optimality for the plans included in the optimization.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] A Novel Method to Automate the Optimization of Multileaf Collimator Transmission and Dosimetric Leaf Gap Parameters for Eclipse Treatment Planning System
    DiCostanzo, D.
    Ayan, A.
    MEDICAL PHYSICS, 2020, 47 (06) : E571 - E571
  • [2] Interpolation of Dose Planes for Use in the Optimization of Multi-Leaf Collimator Dosimetric Leaf Gap and Transmission
    DiCostanzo, Dominic J.
    Ayan, Ahmet S.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [3] Application of the Integral Quality Monitor (IQM) Transmission Detector for Optimization of Dosimetric Leaf Gap and Multi-Leaf Collimator Transmission Factor
    Martinez, S.
    Dean, D.
    Ghafarian, M.
    Price, M.
    Homann, K.
    MEDICAL PHYSICS, 2020, 47 (06) : E701 - E701
  • [4] Dosimetric Leaf Gap and Transmission for Millennium 120-Leaf MLC On Varian TrueBeam
    Darko, Johnson
    Bui, Patricia
    Gunasekara, Dinindu
    Fleck, Andre
    Osei, Ernest
    MEDICAL PHYSICS, 2018, 45 (06) : E335 - E335
  • [5] The effect of depth and control point number for MLC transmission and dosimetric leaf gap
    Kucukmorkoc, E.
    Doyuran, M.
    Caglar, M.
    Kucuk, N.
    Acar, H.
    Altinok, A.
    Caglar, H.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S763 - S764
  • [6] A Novel Approach for Measurement of Dosimetric Leaf Gap for a Rounded Leaf End Multi-Leaf Collimator
    Weinberg, R.
    Boyer, A.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [7] Experimental characterization of the dosimetric leaf gap
    Mullins, Joel
    DeBlois, Francois
    Syme, Alasdair
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2016, 2 (06):
  • [8] Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases
    Middlebrook, Nigel D.
    Sutherland, Bess
    Kairn, Tanya
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (04): : 133 - 139
  • [9] Dosimetric investigations of leaf sequencing parameters
    Agazaryan, N
    Solberg, T
    MEDICAL PHYSICS, 2002, 29 (06) : 1263 - 1263
  • [10] Study of Dosimetric Leaf Gap and Transmission Factor Variations Affecting Common Clinical QA Tools
    Monroe, J.
    Bull, C.
    MEDICAL PHYSICS, 2015, 42 (06) : 3500 - 3500